Highly Sensitive TiO2/Au/Graphene Layer-Based Surface Plasmon Resonance Biosensor for Cancer Detection

被引:98
作者
Mostufa, Shahriar [1 ]
Akib, Tarik Bin Abdul [1 ]
Rana, Md Masud [1 ]
Islam, Md Rabiul [2 ]
机构
[1] Rajshahi Univ Engn & Technol, Dept Elect & Elect Engn, Rajshahi 6204, Bangladesh
[2] Univ Wollongong, Sch Elect Comp & Telecommfiunicat Engn, Wollongong, NSW 2522, Australia
来源
BIOSENSORS-BASEL | 2022年 / 12卷 / 08期
关键词
cancer detection; surface plasmon resonance biosensor; FEM; angular interrogation; biosensor; numerical approach; REFRACTIVE-INDEX SENSOR; LABEL-FREE; NUMERICAL APPROACH; SPR BIOSENSOR; BREAST-CANCER; GRAPHENE; CELLS; ENHANCEMENT; PERFORMANCE; DESIGN;
D O I
10.3390/bios12080603
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this article, a hybrid TiO2/Au/graphene layer-based surface plasmon resonance (SPR) sensor with improved sensitivity and capability for cancer detection is presented. The finite element method (FEM) was used for numerical analysis. The proposed SPR biosensor was structured based on the angular analysis of the attenuated total reflection (ATR) method for the detection of various types of cancer using the refractive index component. The resonance angle shifted owing to the increment of normal and cancerous cells' refractive index, which varied between 1.36 and 1.401 for six different types of normal and cancerous cells. According to numerical results, the obtained sensitivities for skin (basal), cervical (HeLa), adrenal gland (PC12), blood (Jurkat), and breast (MCF-7 and MDA-MB-231) cancer cells were 210 deg/RIU, 245.83 deg/RIU, 264.285 deg/RIU, 285.71 deg/RIU, 292.86 deg/RIU, and 278.57 deg/RIU, respectively. Furthermore, the detection accuracy (DA), figure of merits (EOM), and signal-to-noise ratio (SNR) were also obtained, with values of 0.263 deg(-1), 48.02 RIU-1, and 3.84, respectively. Additionally, the distribution of the electric field and the propagation of the magnetic field for resonant and non-resonant conditions of the proposed structure were illustrated. It was found that an enhanced field was exhibited on the surface of the plasmonic material for resonant conditions. We also measured the penetration depth of 180 nm using decayed electric field intensity. Furthermore, the impact of using a TiO2/Au/graphene layer was demonstrated. We further conducted analyses of the effects of the thickness of the gold layer and the effects of additional graphene layers on overall sensitivities for six different types of cancer. The proposed TiO2/Au/graphene layered structure exhibited the highest overall sensitivity in terms of detecting cancerous cells from healthy cells. Moreover, the proposed sensor was numerically analyzed for a wide range of biological solutions (refractive index 1.33-1.41), and the sensor linearity was calculated with a linear regression coefficient (R-2) of 0.9858. Finally, numerical results obtained in this manuscript exhibited high sensitivity in comparison with previously reported studies.
引用
收藏
页数:22
相关论文
共 97 条
[1]   Design and Numerical Analysis of a Graphene-Coated SPR Biosensor for Rapid Detection of the Novel Coronavirus [J].
Akib, Tarik Bin Abdul ;
Mou, Samia Ferdous ;
Rahman, Md Motiur ;
Rana, Md Masud ;
Islam, Md Rabiul ;
Mehedi, Ibrahim M. ;
Mahmud, M. A. Parvez ;
Kouzani, Abbas Z. .
SENSORS, 2021, 21 (10)
[2]   3D-printed microfluidic devices [J].
Amin, Reza ;
Knowlton, Stephanie ;
Hart, Alexander ;
Yenilmez, Bekir ;
Ghaderinezhad, Fariba ;
Katebifar, Sara ;
Messina, Michael ;
Khademhosseini, Ali ;
Tasoglu, Savas .
BIOFABRICATION, 2016, 8 (02)
[3]   Crucial Role of the Adhesion Layer on the Plasmonic Fluorescence Enhancement [J].
Aouani, Heykel ;
Wenger, Jerome ;
Gerard, Davy ;
Rigneault, Herve ;
Devaux, Eloise ;
Ebbesen, Thomas W. ;
Mahdavi, Farhad ;
Xu, Tingjun ;
Blair, Steve .
ACS NANO, 2009, 3 (07) :2043-2048
[4]   Graphene-based hybrid materials and devices for biosensing [J].
Artiles, Mayra S. ;
Rout, Chandra Sekhar ;
Fisher, Timothy S. .
ADVANCED DRUG DELIVERY REVIEWS, 2011, 63 (14-15) :1352-1360
[5]   Photonic Crystal Fiber-Based Refractive Index Sensor for Early Detection of Cancer [J].
Ayyanar, N. ;
Raja, G. Thavasi ;
Sharma, Mohit ;
Kumar, D. Sriram .
IEEE SENSORS JOURNAL, 2018, 18 (17) :7093-7099
[6]   Highly Sensitive Metamaterial Biosensor for Cancer Early Detection [J].
Azab, Mohammad Y. ;
Hameed, Mohamed Farhat O. ;
Nasr, Abed M. ;
Obayya, S. S. A. .
IEEE SENSORS JOURNAL, 2021, 21 (06) :7748-7755
[7]   Graphene and its derivatives as biomedical materials: future prospects and challenges [J].
Banerjee, Arghya Narayan .
INTERFACE FOCUS, 2018, 8 (03)
[8]   Design and Simulation of a High Sensitive Surface Plasmon Resonance Biosensor for Detection of Biomolecules [J].
Basak, Chaity ;
Hosain, Md Kamal ;
Sazzad, Ahsan Ahmed .
SENSING AND IMAGING, 2019, 21 (01)
[9]  
Baxter J., 2012, NAT PHOTONICS, V6, P268, DOI [10.1038/nphoton.2012.96, DOI 10.1038/NPHOTON.2012.96]
[10]   Real-time RT-PCR correlates with immunocytochemistry for the detection of disseminated epithelial cells in bone marrow aspirates of patients with breast cancer [J].
Benoy, IH ;
Elst, H ;
Van der Auwera, I ;
Van Laere, S ;
van Dam, P ;
Van Marck, E ;
Scharpé, S ;
Vermeulen, PB ;
Dirix, LY .
BRITISH JOURNAL OF CANCER, 2004, 91 (10) :1813-1820