An evaluation of image quality and accuracy of eye bank measurement of donor cornea endothelial cell density in the Specular Microscopy Ancillary Study

被引:59
|
作者
Lass, JH
Gal, RL
Ruedy, KJ
Benetz, BA
Beck, RW
Baratz, KH
Holland, EJ
Kalajian, A
Kollman, C
Manning, FJ
Mannis, MJ
McCoy, K
Montoya, M
Stulting, RD
Xing, DY
Densley, D
Moke, PS
Powe, A
Edwards, S
Karpinecz, L
Rice, C
Redford, M
Cotch, MF
机构
[1] Jaeb Ctr Hlth Res, Cornea Donor Study Coordinating Ctr, Tampa, FL 33647 USA
[2] Case Western Reserve Univ, Cleveland, OH 44106 USA
[3] Univ Hosp Cleveland, Cleveland, OH 44106 USA
[4] NEI, Bethesda, MD 20892 USA
关键词
D O I
10.1016/j.ophtha.2004.10.045
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The Specular Microscopy Ancillary Study was designed to examine donor corneal endothelial specular image quality, compare the central endothelial cell density determined by eye banks with the endothelial cell density determined by a central specular microscopy reading center, and evaluate donor factors that may have an impact on specular image quality and endothelial cell density accuracy. Design: Nonrandomized comparative trial. Participants: Endothelial specular images of donor corneas assigned in the Cornea Donor Study. Methods: Certified readers assessed donor image quality (analyzable from fair to excellent vs. unanalyzable) and determined the central endothelial cell density. Independent adjudication was performed if there was a difference in the quality of grading or if the endothelial cell density varied by greater than or equal to5.0% between readers. Average reading center-determined endothelial cell density was compared with the endothelial cell density determined by each eye bank. Main Outcome Measures: Evaluation of image quality and accuracy of endothelial cell density. Results: Of 688 donor endothelial images submitted by 23 eye banks, 663 (96%) were analyzable (excellent, 40 [6%]; good, 302 [44%]; fair, 321 [47%]), and 25 (4%) were unanalyzable by reading center standards. In situ retrieval and greater epithelial exposure correlated with a higher image quality grading. The eye bank-determined endothelial cell density of 434 of the 663 (65%) analyzable images were within 10% of the endothelial cell density determined by the reading center, whereas 185 (28%) were more than 10% higher and 44 (7%) were more than 10% lower. Greater variation in endothelial cell density between the eye banks and the reading center was observed with shorter time of death to preservation, presence of an epithelial defect, folds in Descemet's membrane, lower image quality, and the use of fixed-frame or center method endothelial cell density analysis. Conclusions: Overall, donor endothelial specular image quality and accuracy of endothelial cell density determination were good. However, the data suggest that factors that may affect image quality and contribute to variation in interpretation of the endothelial cell density should be addressed, because the donor endothelial cell density is an important parameter for assessing long-term corneal graft survival. (C) 2005 by the American Academy of Ophthalmology.
引用
收藏
页码:431 / 440
页数:10
相关论文
共 22 条
  • [1] An evaluation of image quality and accuracy of eye bank measurement of donor cornea endothelial cell density in the Specular Microscopy Ancillary Study (vol 112, pg 431, 2005)
    Lass, JH
    Gal, RL
    Ruedy, KJ
    Benetz, BA
    Beck, RW
    Baratz, KH
    Holland, EJ
    Kalajian, A
    Kollman, C
    Manning, FJ
    Mannis, MJ
    McCoy, K
    Montoya, M
    Stulting, RD
    Xing, DY
    Densley, D
    Moke, PS
    Powe, A
    Edwards, S
    Karpinecz, L
    Rice, C
    Redford, M
    Cotch, MF
    OPHTHALMOLOGY, 2005, 112 (08) : 1394 - 1394
  • [2] Image quality and accuracy of eye bank measurement of donor cornea endothelial cell density in The Cornea Donor Study
    Lass, JH
    Beck, RW
    Edwards, SR
    Gal, RL
    Ruedy, KJ
    Mannis, MJ
    Holland, EJ
    Rice, CL
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U635 - U635
  • [3] Specular microscopy ancillary study methods for donor endothelial cell density determination of cornea donor study images
    Benetz, BA
    Gal, RL
    Ruedy, KJ
    Rice, C
    Beck, RW
    Kalajian, AD
    Lass, JH
    CURRENT EYE RESEARCH, 2006, 31 (04) : 319 - 327
  • [4] Systematic and individual differences in donor cornea Endothelial Cell Density (ECD) measurements with specular microscopy vs. sucrose light microscopy
    Van Dooren, Bart
    Claerhout, Ilse
    Mulder, Paul
    Pels, Elisabeth
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [5] Donor Age and Factors Related to Endothelial Cell Loss 10 Years after Penetrating Keratoplasty Specular Microscopy Ancillary Study
    Lass, Jonathan H.
    Benetz, Beth Ann
    Gal, Robin L.
    Kollman, Craig
    Raghinaru, Dan
    Dontchev, Mariya
    Mannis, Mark J.
    Holland, Edward J.
    Chow, Christopher
    McCoy, Kristen
    Price, Francis W., Jr.
    Sugar, Alan
    Verdier, David D.
    Beck, Roy W.
    OPHTHALMOLOGY, 2013, 120 (12) : 2428 - 2435
  • [6] Comparison of Donor Cornea Endothelial Cell Density Determined by Eye Banks and by a Central Reading Center in the Cornea Preservation Time Study
    Benetz, Beth Ann
    Stoeger, Christopher G.
    Patel, Sanjay V.
    O'Brien, Robert C.
    Szczotka-Flynn, Loretta B.
    Ayala, Allison R.
    Maguire, Maureen G.
    Menegay, Harry J.
    Bedard, Peter
    Clover, Jameson M.
    Gupta, Pankaj C.
    McCoy, Kristen E.
    Song, Jonathan C.
    Lass, Jonathan H.
    CORNEA, 2019, 38 (04) : 426 - 432
  • [7] Analyses of Factors Affecting Endothelial Cell Density in an Eye Bank Corneal Donor Database
    Kwon, Ji Won
    Cho, Kyong Jin
    Kim, Hong Kyu
    Lee, Jimmy K.
    Gore, Patrick K.
    McCartney, Mitchell D.
    Chuck, Roy S.
    CORNEA, 2016, 35 (09) : 1206 - 1210
  • [8] Comparison of Endothelial Cell Density of Organ Cultured Corneas With Cornea Donor Study
    Campolmi, Nelly
    He, Zhiguo
    Acquart, Sophie
    Trone, Marie-Caroline
    Bernard, Aurelien
    Gauthier, Anne-Sophie
    Garraud, Olivier
    Forest, Fabien
    Peoc'h, Michel
    Gain, Philippe
    Thuret, Gilles
    CORNEA, 2014, 33 (06) : 597 - 603
  • [9] Significant differences between specular microscopy and corneal bank endothelial cell counts - a pilot study
    Rickmann, Annekatrin
    Boden, Katrin E.
    Wahl, Silke
    Jung, Sascha
    Boden, Karl T.
    Szurman, Peter
    Januschowski, Kai
    ACTA OPHTHALMOLOGICA, 2019, 97 (08) : E1077 - E1081
  • [10] A STUDY IN THE MEASUREMENT OF CORNEAL ENDOTHELIAL CELL-DENSITY USING THE SPECULAR MICROSCOPE