Automated construction of an intraoperative high-dose-rate treatment plan library for the Varian brachytherapy treatment planning system

被引:5
作者
Deufel, Christopher L. [1 ]
Furutani, Keith M. [1 ]
Dahl, Robert A. [1 ]
Haddock, Michael G. [1 ]
机构
[1] Mayo Clin, Dept Radiat Oncol, 200 1st St SW, Rochester, MN 55905 USA
关键词
Intraoperative; IOHDR; Brachytherapy; Sarcoma; Atlas; Library; Optimization; THERAPY HDR-IORT; RADIATION-THERAPY; RECTAL-CANCER; PART; OPTIMIZATION; MANAGEMENT; RATIONALE; DELIVERY; SARCOMAS; AAPM;
D O I
10.1016/j.brachy.2016.04.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The ability to create treatment plans for intraoperative high-dose-rate (IOHDR) brachytherapy is limited by lack of imaging and time constraints. An automated method for creation of a library of high-dose-rate brachytherapy plans that can be used with standard planar applicators in the intraoperative setting is highly desirable. METHODS AND MATERIALS: Nonnegative least squares algebraic methods were used to identify dwell time values for flat, rectangular planar applicators. The planar applicators ranged in length and width from 2 cm to 25 cm. Plans were optimized to deliver an absorbed dose of 10 Gy to three different depths from the patient surface: 0 cm, 0.5 cm, and 1.0 cm. Software was written to calculate the optimized dwell times and insert dwell times and positions into a.XML plan template that can be imported into the Varian brachytherapy treatment planning system. The user may import the.XML template into the treatment planning system in the intraoperative setting to match the patient applicator size and prescribed treatment depth. RESULTS: A total of 1587 library plans were created for IOHDR brachytherapy. Median plan generation time was approximately 1 minute per plan. Plan dose was typically 100% +/- 1% (mean, standard deviation) of the prescribed dose over the entire length and width of the applicator. Plan uniformity was best for prescription depths of 0 cm and 0.5 cm from the patient surface. CONCLUSIONS: An IOHDR plan library may be created using automated methods. Thousands of plan templates may be optimized and prepared in a few hours to accommodate different applicator sizes and treatment depths and reduce treatment planning time. The automated method also enforces dwell time symmetry for symmetrical applicator geometries, which simplifies quality assurance. (C) 2016 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:531 / 536
页数:6
相关论文
共 50 条
  • [31] Comparison of high-dose-rate and low-dose-rate brachytherapy in the treatment of endometrial carcinoma
    Fayed, Alaa
    Mutch, David G.
    Rader, Janet S.
    Gibb, Randall K.
    Powell, Matthew A.
    Wright, Jason D.
    ElNaqa, Issam
    Zoberi, Imran
    Grigsby, Perry W.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (02): : 480 - 484
  • [32] High-Dose-Rate Brachytherapy as an Organ-Sparing Treatment for Early Penile Cancer
    Pohankova, Denisa
    Sirak, Igor
    Vosmik, Milan
    Kasaova, Linda
    Grepl, Jakub
    Paluska, Petr
    Holub, Lukas
    Spacek, Jiri
    Hodek, Miroslav
    Kopecek, Martin
    Petera, Jiri
    CANCERS, 2022, 14 (24)
  • [33] High-dose-rate (HDR) brachytherapy for the treatment of benign obstructive endobronchial granulation tissue
    Madu, Chika N.
    Machuzak, Michael S.
    Sterman, Daniel H.
    Musani, Ali
    Ahya, Vivek
    McDonough, James
    Metz, James M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (05): : 1450 - 1456
  • [34] Towards artificial intelligence-based automated treatment planning in clinical practice: A prospective study of the first clinical experiences in high-dose-rate prostate brachytherapy
    Barten, Danique L. J.
    Pieters, Bradley R.
    Bouter, Anton
    van der Meer, Marjolein C.
    Maree, Stef C.
    Hinnen, Karel A.
    Westerveld, Henrike
    Bosman, Peter A. N.
    Alderliesten, Tanja
    van Wieringen, Niek
    Bel, Arjan
    BRACHYTHERAPY, 2023, 22 (02) : 279 - 289
  • [35] Low-dose-rate or high-dose-rate brachytherapy in treatment of prostate cancer - between options
    Skowronek, Janusz
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2013, 5 (01) : 33 - 41
  • [36] High-dose-rate and pulsed-dose-rate brachytherapy in palliative treatment of head and neck cancers
    Bartochowska, Anna
    Wierzbicka, Malgorzata
    Skowronek, Janusz
    Leszczynska, Malgorzata
    Szyfter, Witold
    BRACHYTHERAPY, 2012, 11 (02) : 137 - 143
  • [37] High-dose-rate brachytherapy for vaginal cancer: Learning from treatment complications
    Tyree, WC
    Cardenes, H
    Randall, M
    Papiez, L
    INTERNATIONAL JOURNAL OF GYNECOLOGICAL CANCER, 2002, 12 (01) : 27 - 31
  • [38] Treatment of keloids by high-dose-rate brachytherapy:: A seven-year study
    Guix, B
    Henríquez, I
    Andrés, A
    Finestres, F
    Tello, JI
    Martínez, A
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 50 (01): : 167 - 172
  • [39] High-dose-rate brachytherapy boost for prostate cancer treatment: Different combinations of hypofractionated regimens and clinical outcomes
    Vigneault, Eric
    Mbodji, Khaly
    Magnan, Sindy
    Despres, Philippe
    Lavallee, Marie-Claude
    Aubin, Sylviane
    Beaulieu, Luc
    Foster, William
    Martin, Andre-Guy
    RADIOTHERAPY AND ONCOLOGY, 2017, 124 (01) : 49 - 55
  • [40] Custom-made micro applicators for high-dose-rate brachytherapy treatment of chronic psoriasis
    Buzurovic, Ivan M.
    O'Farrell, Desmond A.
    Bhagwat, Mandar S.
    Hansen, Jorgen L.
    Harris, Thomas C.
    Friesen, Scott
    Cormack, Robert A.
    Devlin, Philip M.
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2017, 9 (03) : 263 - 269