Automated construction of an intraoperative high-dose-rate treatment plan library for the Varian brachytherapy treatment planning system

被引:5
作者
Deufel, Christopher L. [1 ]
Furutani, Keith M. [1 ]
Dahl, Robert A. [1 ]
Haddock, Michael G. [1 ]
机构
[1] Mayo Clin, Dept Radiat Oncol, 200 1st St SW, Rochester, MN 55905 USA
关键词
Intraoperative; IOHDR; Brachytherapy; Sarcoma; Atlas; Library; Optimization; THERAPY HDR-IORT; RADIATION-THERAPY; RECTAL-CANCER; PART; OPTIMIZATION; MANAGEMENT; RATIONALE; DELIVERY; SARCOMAS; AAPM;
D O I
10.1016/j.brachy.2016.04.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: The ability to create treatment plans for intraoperative high-dose-rate (IOHDR) brachytherapy is limited by lack of imaging and time constraints. An automated method for creation of a library of high-dose-rate brachytherapy plans that can be used with standard planar applicators in the intraoperative setting is highly desirable. METHODS AND MATERIALS: Nonnegative least squares algebraic methods were used to identify dwell time values for flat, rectangular planar applicators. The planar applicators ranged in length and width from 2 cm to 25 cm. Plans were optimized to deliver an absorbed dose of 10 Gy to three different depths from the patient surface: 0 cm, 0.5 cm, and 1.0 cm. Software was written to calculate the optimized dwell times and insert dwell times and positions into a.XML plan template that can be imported into the Varian brachytherapy treatment planning system. The user may import the.XML template into the treatment planning system in the intraoperative setting to match the patient applicator size and prescribed treatment depth. RESULTS: A total of 1587 library plans were created for IOHDR brachytherapy. Median plan generation time was approximately 1 minute per plan. Plan dose was typically 100% +/- 1% (mean, standard deviation) of the prescribed dose over the entire length and width of the applicator. Plan uniformity was best for prescription depths of 0 cm and 0.5 cm from the patient surface. CONCLUSIONS: An IOHDR plan library may be created using automated methods. Thousands of plan templates may be optimized and prepared in a few hours to accommodate different applicator sizes and treatment depths and reduce treatment planning time. The automated method also enforces dwell time symmetry for symmetrical applicator geometries, which simplifies quality assurance. (C) 2016 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:531 / 536
页数:6
相关论文
共 50 条
  • [21] Intraoperative high-dose-rate brachytherapy for paranasal sinus tumors
    Nag, S
    Tippin, D
    Grecula, J
    Schuller, D
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 58 (01): : 155 - 160
  • [22] Benefit of adaptive CT-based treatment planning in high-dose-rate endorectal brachytherapy for rectal cancer
    van den Ende, Roy P. J.
    Rijkmans, Eva C.
    Kerkhof, Ellen M.
    Nout, Remi A.
    Ketelaars, Martijn
    Laman, Mirjam S.
    Marijnen, Corrie A. M.
    van der Heide, Uulke A.
    BRACHYTHERAPY, 2018, 17 (01) : 78 - 85
  • [23] Verification of high-dose-rate brachytherapy treatment planning dose distribution using liquid-filled ionization chamber array
    Yoosuf, A. B. Mohamed
    Jeevanandam, Prakash
    Whitten, Glenn
    Workman, Geraldine
    McGarry, Conor K.
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2018, 10 (02) : 142 - 154
  • [24] Predicting treatment plan approval probability for high-dose-rate brachytherapy of cervical cancer using adversarial deep learning
    Gao, Yin
    Gonzalez, Yesenia
    Nwachukwu, Chika
    Albuquerque, Kevin
    Jia, Xun
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (09)
  • [25] Interstitial high-dose-rate brachytherapy in the treatment of base of tongue carcinoma
    Takácsi-Nagy, Z
    Polgár, C
    Oberna, F
    Somogyi, A
    Major, T
    Remenár, É
    Fodor, J
    Kásler, M
    Németh, G
    STRAHLENTHERAPIE UND ONKOLOGIE, 2004, 180 (12) : 768 - 775
  • [26] Three different intraoperative radiation modalities (electron beam, high-dose-rate brachytherapy, and iodine-125 brachytherapy) in the adjuvant treatment of patients with recurrent colorectal adenocarcinoma
    Martínez-Monge, R
    Nag, S
    Martin, EW
    CANCER, 1999, 86 (02) : 236 - 247
  • [27] Sensitivity of dose-volume indices to computation settings in high-dose-rate prostate brachytherapy treatment plan evaluation
    van der Meer, Marjolein C.
    Bosman, Peter A. N.
    Pieters, Bradley R.
    Niatsetski, Yury
    van Wieringen, Niek
    Alderliesten, Tanja
    Bel, Arjan
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (04): : 66 - 74
  • [28] Treatment of Vaginal Recurrences in Endometrial Carcinoma by High-dose-rate Brachytherapy
    Sorbe, Bengt
    Soderstrom, Karin
    ANTICANCER RESEARCH, 2013, 33 (01) : 241 - 247
  • [29] Perioperative High-Dose-Rate Brachytherapy in the Treatment of Recurrent Malignant Gliomas
    Fabrini, Maria Grazia
    Perrone, Franco
    De Franco, Lucia
    Pasqualetti, Francesco
    Grespi, Silvia
    Vannozzi, Riccardo
    Cionini, Luca
    STRAHLENTHERAPIE UND ONKOLOGIE, 2009, 185 (08) : 524 - 529
  • [30] High-Dose-Rate Brachytherapy for the Treatment of Recalcitrant Keloids: A Unique, Effective Treatment Protocol
    van Leeuwen, Michiel C. E.
    Stokmans, Suzanne C.
    Bulstra, Anne-Eva J.
    Meijer, Otto W. M.
    van Leeuwen, Paul A. M.
    Niessen, Frank B.
    PLASTIC AND RECONSTRUCTIVE SURGERY, 2014, 134 (03) : 527 - 534