Effect of cutting flute design on cortical bone screw insertion torque and pullout strength

被引:44
作者
Yerby, S
Scott, CC
Evans, NJ
Messing, KL
Carter, DR
机构
[1] Vet Affairs Palo Alto Hlth Care Syst, Rehabil Res & Dev Ctr, Palo Alto, CA USA
[2] Stanford Univ, Dept Funct Restorat, Div Orthopaed Surg, Stanford, CA 94305 USA
[3] Stanford Univ, Dept Engn Mech, Biomech Engn Div, Stanford, CA 94305 USA
关键词
bone screw; cutting flute; femur; insertion; torque; pullout strength;
D O I
10.1097/00005131-200103000-00012
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: To determine the effect of the number and length of cutting flutes on the insertion torque and pullout strength for self-tapping 4.5-millimeter cortical bone screws. Design: Screws were self-tapped in the diaphysis of human cadaver femurs. Each of the six screw types studied had different designs with varying cutting flute lengths and numbers. Bone mineral density, insertion torque, and pullout strength were measured. Setting: The study was conducted at an experimental biomechanics laboratory associated with a university medical center. Outcome Measurements: Insertion torque and pullout strength were normalized by the local bone mineral density. Results: The mean normalized insertion torque of the design with four full-length cutting flutes was less than the design with three full-length flutes and the two designs with one-third length flutes (p < 0.05). The mean normalized pullout strength of the screw with four full-length flutes was significantly greater than that of all screws with fewer than three flutes (p < 0.05). Conclusions: Priorities for a cutting flute design should ideally include ease of screw insertion, minimal soft tissue irritation, and maximal screw holding power. Screws with more than two flutes were easier to insert and did not cause cortical damage during insertion. The screw with four full-length flutes showed a trend toward being the easiest to insert and having the greatest holding strength.
引用
收藏
页码:216 / 221
页数:6
相关论文
共 18 条
[1]  
[Anonymous], 1959, METALS ENG BONE JOIN
[2]  
Ansell R H, 1968, J Biomech, V1, P279, DOI 10.1016/0021-9290(68)90023-7
[3]  
*ASTM, 1996, ANN BOOK ASTM STAND, P1085
[4]  
BAUMGART FW, 1993, INJURY S1, V24, P1
[5]   Self-tapping versus standard tapped titanium screw fixation in the upper extremity [J].
Bickley, MBT ;
Hanel, DP .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 1998, 23A (02) :308-311
[6]   TORQUE AND PULLOUT ANALYSIS OF 6 CURRENTLY AVAILABLE SELF-TAPPING AND EMERGENCY SCREWS [J].
BOYLE, JM ;
FROST, DE ;
FOLEY, WL ;
GRADY, JJ .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1993, 51 (01) :45-50
[7]  
CARTER DR, 1992, J BONE MINER RES, V7, P137
[8]   Factors affecting the pullout strength of cancellous bone screws [J].
Chapman, JR ;
Harrington, RM ;
Lee, KM ;
Anderson, PA ;
Tencer, AF ;
Kowalski, D .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (03) :391-398
[9]   DESIGN AND TESTING OF EXTERNAL FIXATOR BONE SCREWS [J].
EVANS, M ;
SPENCER, M ;
WANG, Q ;
WHITE, SH ;
CUNNINGHAM, JL .
JOURNAL OF BIOMEDICAL ENGINEERING, 1990, 12 (06) :457-462
[10]   UNIAXIAL PULLOUT EVALUATION OF INTERNAL SCREW FIXATION [J].
FOLEY, WL ;
FROST, DE ;
PAULIN, WB ;
TUCKER, MR .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1989, 47 (03) :277-280