Image quality assessment of pediatric chest and abdomen CT by deep learning reconstruction

被引:30
作者
Yoon, Haesung
Kim, Jisoo
Lim, Hyun Ji
Lee, Mi-Jung [1 ]
机构
[1] Yonsei Univ, Severance Hosp, Coll Med, Dept Radiol, 50-1 Yonsei Ro, Seoul 03722, South Korea
关键词
Pediatric; CT; Image quality; Deep learning; Iterative reconstruction; STATISTICAL ITERATIVE RECONSTRUCTION; RADIATION-DOSE REDUCTION; FILTERED BACK-PROJECTION; ABDOMINAL CT; ASIR TECHNIQUE; NOISE;
D O I
10.1186/s12880-021-00677-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Efforts to reduce the radiation dose have continued steadily, with new reconstruction techniques. Recently, image denoising algorithms using artificial neural networks, termed deep learning reconstruction (DLR), have been applied to CT image reconstruction to overcome the drawbacks of iterative reconstruction (IR). The purpose of our study was to compare the objective and subjective image quality of DLR and IR on pediatric abdomen and chest CT images. Methods This retrospective study included pediatric body CT images from February 2020 to October 2020, performed on 51 patients (34 boys and 17 girls; age 1-18 years). Non-contrast chest CT (n = 16), contrast-enhanced chest CT (n = 12), and contrast-enhanced abdomen CT (n = 23) images were included. Standard 50% adaptive statistical iterative reconstruction V (ASIR-V) images were compared to images with 100% ASIR-V and DLR at medium and high strengths. Attenuation, noise, contrast to noise ratio (CNR), and signal to noise (SNR) measurements were performed. Overall image quality, artifacts, and noise were subjectively assessed by two radiologists using a four-point scale (superior, average, suboptimal, and unacceptable). A phantom scan was performed including the dose range of the clinical images used in our study, and the noise power spectrum (NPS) was calculated. Quantitative and qualitative parameters were compared using repeated-measures analysis of variance (ANOVA) with Bonferroni correction and Wilcoxon signed-rank tests. Results DLR had better CNR and SNR than 50% ASIR-V in both pediatric chest and abdomen CT images. When compared with 50% ASIR-V, high strength DLR was associated with noise reduction in non-contrast chest CT (33.0%), contrast-enhanced chest CT (39.6%), and contrast-enhanced abdomen CT (38.7%) with increases in CNR at 149.1%, 105.8%, and 53.1% respectively. The subjective assessment of overall image quality and the noise was also better on DLR images (p < 0.001). However, there was no significant difference in artifacts between reconstruction methods. From NPS analysis, DLR methods showed a pattern of reducing the magnitude of noise while maintaining the texture. Conclusion Compared with 50% ASIR-V, DLR improved pediatric body CT images with significant noise reduction. However, artifacts were not improved by DLR, regardless of strength.
引用
收藏
页数:11
相关论文
共 28 条
[1]   Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT [J].
Akagi, Motonori ;
Nakamura, Yuko ;
Higaki, Toru ;
Narita, Keigo ;
Honda, Yukiko ;
Zhou, Jian ;
Yu, Zhou ;
Akino, Naruomi ;
Awai, Kazuo .
EUROPEAN RADIOLOGY, 2019, 29 (11) :6163-6171
[2]  
Anton CG., 2020, RADIOLOGY, DOI [33937841, DOI 10.1148/RADIOL.2020202317:202317]
[3]   Effects of adaptive statistical iterative reconstruction on radiation dose reduction and diagnostic accuracy of pediatric abdominal CT [J].
Bae, Sohi ;
Kim, Myung-Joon ;
Yoon, Choon-Sik ;
Kim, Dong Wook ;
Hong, Jung Hwa ;
Lee, Mi-Jung .
PEDIATRIC RADIOLOGY, 2014, 44 (12) :1541-1547
[4]  
Boone J.M., 2011, SIZE SPECIFIC DOSE E
[5]   Artifact Suppressed Dictionary Learning for Low-Dose CT Image Processing [J].
Chen, Yang ;
Shi, Luyao ;
Feng, Qianjing ;
Yang, Jian ;
Shu, Huazhong ;
Luo, Limin ;
Coatrieux, Jean-Louis ;
Chen, Wufan .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (12) :2271-2292
[6]   Multisection CT Protocols: Sex- and Age-specific Conversion Factors Used to Determine Effective Dose from Dose-Length Product [J].
Deak, Paul D. ;
Smal, Yulia ;
Kalender, Willi A. .
RADIOLOGY, 2010, 257 (01) :158-166
[7]   A Third-Generation Adaptive Statistical Iterative Reconstruction Technique: Phantom Study of Image Noise, Spatial Resolution, Lesion Detectability, and Dose Reduction Potential [J].
Euler, Andre ;
Solomon, Justin ;
Marin, Daniele ;
Nelson, Rendon C. ;
Samei, Ehsan .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2018, 210 (06) :1301-1308
[8]   A simple approach to measure computed tomography (CT) modulation transfer function (MTF) and noise-power spectrum (NPS) using the American College of Radiology (ACR) accreditation phantom [J].
Friedman, Saul N. ;
Fung, George S. K. ;
Siewerdsen, Jeffrey H. ;
Tsui, Benjamin M. W. .
MEDICAL PHYSICS, 2013, 40 (05)
[9]   Computed Tomography Image Quality Evaluation of a New Iterative Reconstruction Algorithm in the Abdomen (Adaptive Statistical Iterative Reconstruction-V) a Comparison With Model-Based Iterative Reconstruction, Adaptive Statistical Iterative Reconstruction, and Filtered Back Projection Reconstructions [J].
Goodenberger, Martin H. ;
Wagner-Bartak, Nicolaus A. ;
Gupta, Shiva ;
Liu, Xinming ;
Yap, Ramon Q. ;
Sun, Jia ;
Tamm, Eric P. ;
Jensen, Corey T. .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (02) :184-190
[10]   Image quality and dose reduction opportunity of deep learning image reconstruction algorithm for CT: a phantom study [J].
Greffier, Joel ;
Hamard, Aymeric ;
Pereira, Fabricio ;
Barrau, Corinne ;
Pasquier, Hugo ;
Beregi, Jean Paul ;
Frandon, Julien .
EUROPEAN RADIOLOGY, 2020, 30 (07) :3951-3959