Functionally graded Co-Cr-Mo coating on Ti-6Al-4V alloy structures

被引:154
作者
Krishna, B. Vamsi [1 ]
Xue, Weichang [1 ]
Bose, Susmita [1 ]
Bandyopadhyay, Amit [1 ]
机构
[1] Washington State Univ, Sch Mech & Mat Engn, WM Keck Biomed Mat Res Lab, Pullman, WA 99164 USA
关键词
functionally graded materials; laser processing; laser engineered net shaping (LENS); biocompatibility; osteoblast;
D O I
10.1016/j.actbio.2007.10.005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Functionally graded, hard and wear-resistant Co-Cr-Mo alloy was coated on Ti-6Al-4V alloy with a metallurgically sound interface using Laser Engineering Net Shaping (LENS (TM)). The addition of the Co-Cr-Mo alloy onto the surface of Ti-6Al-4V alloy significantly increased the surface hardness without any intermetallic phases in the transition region. A 100% Co-Cr-Mo transition from Ti-6Al-4V was difficult to produce due to cracking. However, using optimized LENS (TM) processing parameters, crack-free coatings containing up to 86% Co-Cr-Mo, were deposited on Ti-6Al-4V alloy with excellent reproducibility. Human osteoblast cells were cultured to test in vitro biocompatibility of the coatings. Based on in vitro biocompatibility, increasing the Co-Cr-Mo concentration in the coating reduced the live cell numbers after 14 days of culture on the coating compared with base Ti-6Al-4V alloy. However, coated samples always showed better bone cell proliferation than 100% Co-Cr-Mo alloy. Producing near net shape components with graded compositions using LENS (TM) could potentially be a viable route for manufacturing unitized structures for metal-on-metal prosthetic devices to minimize the wear-induced osteolysis and aseptic loosening that are significant problems in current implant design. (c) 2007 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:697 / 706
页数:10
相关论文
共 40 条
  • [1] AKIHIKO C, 2007, ACTA MAT, V55, P1309
  • [2] Amstutz H. C., 1996, CLIN ORTHOP S, V329, P11
  • [3] [Anonymous], 1992, METALS HDB, V3
  • [4] CHAN FW, 1999, CLIN ORTHOP RELAT R, V369, P10
  • [5] The influence of the enthalpy of mixing during the laser deposition of complex titanium alloys using elemental blends
    Collins, PC
    Banerjee, R
    Fraser, HL
    [J]. SCRIPTA MATERIALIA, 2003, 48 (10) : 1445 - 1450
  • [6] Cooper R A, 1992, J Arthroplasty, V7, P285, DOI 10.1016/0883-5403(92)90050-Z
  • [7] Dougherty D., 2004, Strategic Organization, V2, P35, DOI 10.1177/1476127004040914
  • [8] Mechanical behavior, wear surface morphology, and clinical performance of UHMWPE acetabular components after 10 years of implantation
    Edidin, AA
    Rimnac, CM
    Goldberg, VM
    Kurtz, SM
    [J]. WEAR, 2001, 250 (PART 1) : 152 - 158
  • [9] Frank L, 1999, J Assoc Nurses AIDS Care, V10, P24, DOI 10.1016/S1055-3290(06)60320-8
  • [10] The electrochemical and mechanical behavior of passivated and TiN/AlN-coated CoCrMo and Ti6Al4V alloys
    Goldberg, JR
    Gilbert, JL
    [J]. BIOMATERIALS, 2004, 25 (05) : 851 - 864