An evaluation of techniques for dose calculation on cone beam computed tomography

被引:55
作者
Giacometti, Valentina [1 ]
King, Raymond B. [1 ,2 ]
Agnew, Christina E. [2 ]
Irvine, Denise M. [2 ]
Jain, Suneil [1 ,2 ]
Hounsell, Alan R. [1 ,2 ]
McGarry, Conor K. [1 ,2 ]
机构
[1] Queens Univ Belfast, Ctr Canc Res & Cell Biol, Belfast, Antrim, North Ireland
[2] Northern Ireland Canc Ctr, Radiotherapy Phys, Belfast, Antrim, North Ireland
关键词
ADAPTIVE RADIATION-THERAPY; IMAGE REGISTRATION; CT IMAGES; RADIOTHERAPY; FEASIBILITY; RECONSTRUCTION; DOSIMETRY; ACCURACY; SYSTEM; IMRT;
D O I
10.1259/bjr.20180383
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To assess the accuracy and efficiency of four different techniques, thus determining the optimum method for recalculating dose on cone beam CT (CBCT) images acquired during radiotherapy treatments. Methods: Four established techniques were investigated and their accuracy assessed via dose calculations: (1) applying a standard planning CT (pCT) calibration curve, (2) applying a CBCT site-specific calibration curve, (3) performing a density override and (4) using deformable registration. Each technique was applied to 15 patients receiving volumetric modulated arc therapy to one of three treatment sites, head and neck, lung and prostate. Differences between pCT and CBCT recalculations were determined with dose volume histogram metrics and 2.0%/0.1 mm gamma analysis using the pCT dose distribution as a reference. Results: Dose volume histogram analysis indicated that all techniques yielded differences from expected results between 0.0 and 2.3% for both target volumes and organs at risk. With volumetric gamma analysis, the dose recalculation on deformed images yielded the highest pass-rates. The median pass-rate ranges at 50% threshold were 99.6-99.9%, 94.6-96.0%, and 94.8.0-96.0% for prostate, head and neck and lung patients, respectively. Conclusion: Deformable registration, HU override and site-specific calibration curves were all identified as dosimetrically accurate and efficient methods for dose calculation on CBCT images. Advances in knowledge: With the increasing adoption of CBCT, this study provides clinical radiotherapy departments with invaluable information regarding the comparison of dose reconstruction methods, enabling a more accurate representation of a patient's treatment. It can also integrate studies in which CBCT is used in image-guided radiation therapy and for adaptive radiotherapy planning processes.
引用
收藏
页数:11
相关论文
共 40 条
[1]   Automated algorithm for CBCT-based dose calculations of prostate radiotherapy with bilateral hip prostheses [J].
Almatani, Turki ;
Hugtenburg, Richard P. ;
Lewis, Ryan D. ;
Barley, Susan E. ;
Edwards, Mark A. .
BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1066)
[2]   Simplified material assignment for cone beam computed tomography-based dose calculations of prostate radiotherapy with hip prostheses [J].
Almatani, Turki ;
Hugtenburg, Richard P. ;
Lewis, Ryan ;
Barley, Susan ;
Edwards, Mark .
JOURNAL OF RADIOTHERAPY IN PRACTICE, 2016, 15 (02) :170-180
[3]   RESULTS OF A MULTI-INSTITUTION DEFORMABLE REGISTRATION ACCURACY STUDY (MIDRAS) [J].
Brock, Kristy K. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (02) :583-596
[4]   Comparison of 12 deformable registration strategies in adaptive radiation therapy for the treatment of head and neck tumors [J].
Castadot, Pierre ;
Lee, John Aldo ;
Parraga, Adriane ;
Geets, Xavier ;
Macq, Benoit ;
Gregoire, Vincent .
RADIOTHERAPY AND ONCOLOGY, 2008, 89 (01) :1-12
[5]   Feasibility of CBCT-based dose with a patient-specific stepwise HU-to-density curve to determine time of replanning [J].
Chen, Shifeng ;
Le, Quynh ;
Mutaf, Yildirim ;
Lu, Wei ;
Nichols, Elizabeth M. ;
Yi, Byong Yong ;
Leven, Tish ;
Prado, Karl L. ;
D'Souza, Warren D. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05) :64-69
[6]   Motion Management for Radical Radiotherapy in Non-small Cell Lung Cancer [J].
Cole, A. J. ;
Hanna, G. G. ;
Jain, S. ;
O'Sullivan, J. M. .
CLINICAL ONCOLOGY, 2014, 26 (02) :67-80
[7]   Accuracy of dose calculations on kV cone beam CT images of lung cancer patients [J].
de Smet, Mariska ;
Schuring, Danny ;
Nijsten, Sebastiaan ;
Verhaegen, Frank .
MEDICAL PHYSICS, 2016, 43 (11) :5934-5941
[8]   A study on adaptive IMRT treatment planning using kV cone-beam CT [J].
Ding, George X. ;
Duggan, Dennis M. ;
Coffey, Charles W. ;
Deeley, Matthew ;
Hallahan, Dennis E. ;
Cmelak, Anthony ;
Malcolm, Arnold .
RADIOTHERAPY AND ONCOLOGY, 2007, 85 (01) :116-125
[9]   Comparison of CT number calibration techniques for CBCT-based dose calculation [J].
Dunlop, Alex ;
McQuaid, Dualta ;
Nill, Simeon ;
Murray, Julia ;
Poludniowski, Gavin ;
Hansen, Vibeke N. ;
Bhide, Shreerang ;
Nutting, Christopher ;
Harrington, Kevin ;
Newbold, Kate ;
Oelfke, Uwe .
STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 (12) :970-978
[10]   Quasi real time in vivo dosimetry for VMAT [J].
Fidanzio, A. ;
Porcelli, A. ;
Azario, L. ;
Greco, F. ;
Cilla, S. ;
Grusio, M. ;
Balducci, M. ;
Valentini, V. ;
Piermattei, A. .