Design, manufacture, and evaluation of an anthropomorphic pelvic phantom purpose-built for radiotherapy dosimetric intercomparison

被引:24
作者
Harrison, K. M. [1 ,2 ]
Ebert, M. A. [3 ,4 ]
Kron, T. [5 ]
Howlett, S. J. [6 ]
Cornes, D. [7 ]
Hamilton, C. S. [8 ]
Denham, J. W. [1 ,9 ]
机构
[1] Calvary Mater Newcastle, Dept Radiat Oncol, Waratah, NSW 2298, Australia
[2] Univ Newcastle, Sch Phys, Callaghan, NSW 2308, Australia
[3] Sir Charles Gairdner Hosp, Dept Radiat Oncol, Nedlands, WA 6009, Australia
[4] Univ Western Australia, Sch Phys, Nedlands, WA 6009, Australia
[5] Peter MacCallum Canc Ctr, Dept Phys Sci, Melbourne, Vic 8006, Australia
[6] Australiasian Coll Phys Scientists & Engineers Me, Sydney, NSW 2020, Australia
[7] Calvary Mater Newcastle, Trans Tasman Radiat Oncol Grp, Waratah, NSW 2298, Australia
[8] Heidelberg Repatriat Hosp, Heidelberg, Vic 3081, Australia
[9] Univ Newcastle, Sch Med & Populat Hlth, Callaghan, NSW 2308, Australia
关键词
Anthropomorphic phantoms; Quality Assurance; QUALITY-ASSURANCE; IMPLEMENTATION; INSTITUTIONS; PHYSICS; AUDIT;
D O I
10.1118/1.3626573
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: An anthropomorphic pelvic phantom was designed and constructed to meet specific criteria for multicenter radiotherapy dosimetric intercomparison. Methods: Three dimensional external and organ outlines were generated from a computed tomography image set of a male pelvis, forming the basis of design for an anatomically realistic phantom. Clinically relevant points of interest were selected throughout the dataset where point-dose values could be measured with thermoluminescence dosimeters and a small-volume ionization chamber. Following testing, three materials were selected and the phantom was manufactured using modern prototyping techniques into five separate coronal slices. Time lines and resource requirements for the phantom design and manufacture were recorded. The ability of the phantom to mimic the entire treatment chain was tested. Results: The phantom CT images indicated that organ densities and geometries were comparable to those of the original patient. The phantom proved simple to load for dosimetry and rapid to assemble. Due to heat release during manufacture, small air gaps and density heterogeneities were present throughout the phantom. The overall cost for production of the prototype phantom was comparable to other commercial anthropomorphic phantoms. The phantom was shown to be suitable for use as a "patient" to mimic the entire treatment chain for typical external beam radiotherapy for prostate and rectal cancer. Conclusions: The phantom constructed for the present study incorporates all characteristics necessary for accurate Level III intercomparison studies. Following use in an extensive Level III dosimetric comparison over a large time scale and geographic area, the phantom retained mechanical stability and did not show signs of radiation-induced degradation. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3626573]
引用
收藏
页码:5330 / 5337
页数:8
相关论文
共 28 条
[1]  
[Anonymous], 5 INT C INT RA UNPUB
[2]  
[Anonymous], CONS PRIC IND G2
[3]  
[Anonymous], MED PH IN PRESS 0727
[4]  
[Anonymous], TROG 01 04 PROTOCOL
[5]  
[Anonymous], AAPM REP
[6]  
[Anonymous], THESIS U NEWCASTLE
[7]  
[Anonymous], TROG 03 04
[8]  
[Anonymous], AUSTRALAS PHYS ENG S
[9]   A SURVEY OF TISSUE SUBSTITUTES AND PHANTOMS IN 98 UNITED-STATES MEDICAL-CENTERS [J].
BAIRD, LC ;
WHITE, DR .
MEDICAL PHYSICS, 1982, 9 (03) :434-435
[10]  
Bernier J, 2002, EUR J CANCER, V38, pS155