Time-resolved dosimetric verification of respiratory-gated radiotherapy exposures using a high-resolution 2D ionisation chamber array

被引:11
作者
King, R. B. [1 ]
Agnew, C. E. [2 ]
O'Connell, B. F. [2 ]
Prise, K. M. [1 ]
Hounsell, A. R. [1 ,2 ]
McGarry, C. K. [1 ,2 ]
机构
[1] Queens Univ Belfast, Ctr Canc Res & Cell Biol, Belfast BT9 7AE, Antrim, North Ireland
[2] Belfast Hlth & Social Care Trust, Northern Ireland Canc Ctr, Radiotherapy Phys, Belfast BT9 7AB, Antrim, North Ireland
关键词
time-resolved dosimetry; respiratory gating; treatment verification; QUALITY-ASSURANCE; MOTION; IMPLEMENTATION; FILES;
D O I
10.1088/0031-9155/61/15/5529
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of this work was to track and verify the delivery of respiratory-gated irradiations, performed with three versions of TrueBeam linac, using a novel phantom arrangement that combined the OCTAVIUS (R) SRS 1000 array with a moving platform. The platform was programmed to generate sinusoidal motion of the array. This motion was tracked using the real-time position management (RPM) system and four amplitude gating options were employed to interrupt MV beam delivery when the platform was not located within set limits. Time-resolved spatial information extracted from analysis of x-ray fluences measured by the array was compared to the programmed motion of the platform and to the trace recorded by the RPM system during the delivery of the x-ray field. Temporal data recorded by the phantom and the RPM system were validated against trajectory log files, recorded by the linac during the irradiation, as well as oscilloscope waveforms recorded from the linac target signal. Gamma analysis was employed to compare time-integrated 2D x-ray dose fluences with theoretical fluences derived from the probability density function for each of the gating settings applied, where gamma criteria of 2%/2 mm, 1%/ 1 mm and 0.5%/0.5 mm were used to evaluate the limitations of the RPM system. Excellent agreement was observed in the analysis of spatial information extracted from the SRS 1000 array measurements. Comparisons of the average platform position with the expected position indicated absolute deviations of <0.5 mm for all four gating settings. Differences were observed when comparing time-resolved beam-on data stored in the RPM files and trajectory logs to the true target signal waveforms. Trajectory log files underestimated the cycle time between consecutive beam-on windows by 10.0 +/- 0.8 ms. All measured fluences achieved 100% pass-rates using gamma criteria of 2%/2 mm and 50% of the fluences achieved pass-rates >90% when criteria of 0.5%/0.5 mm were used. Results using this novel phantom arrangement indicate that the RPM system is capable of accurately gating x-ray exposure during the delivery of a fixed-field treatment beam.
引用
收藏
页码:5529 / 5546
页数:18
相关论文
共 28 条
[1]   Improvement in clinical step and shoot intensity modulated radiation therapy delivery accuracy on an integrated linear accelerator control system [J].
Agnew, C. E. ;
Irvine, D. M. ;
Hounsell, A. R. ;
McGarry, C. K. .
PRACTICAL RADIATION ONCOLOGY, 2014, 4 (01) :43-49
[2]   Implementation of phantom-less IMRT delivery verification using Varian DynaLog files and R/V output [J].
Agnew, C. E. ;
King, R. B. ;
Hounsell, A. R. ;
McGarry, C. K. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (21) :6761-6777
[3]   A technique for respiratory-gated radiotherapy treatment verification with an EPID in cine mode [J].
Berbeco, RI ;
Neicu, T ;
Rietzel, E ;
Chen, GTY ;
Jiang, SB .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (16) :3669-3679
[4]   Evaluation of integrated respiratory gating systems on a Novalis Tx system [J].
Chang, Zheng ;
Liu, TongHai ;
Cai, Jing ;
Chen, Qing ;
Wang, Zhiheng ;
Yin, Fang-Fang .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (03) :71-79
[5]   The potential for undertaking slow CT using a modern CT scanner [J].
Chinneck, C. D. ;
McJury, M. ;
Hounsell, A. R. .
BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (992) :687-693
[6]   Development of a novel experimental model to investigate radiobiological implications of respiratory motion in advanced radiotherapy [J].
Cole, Aidan J. ;
McGarry, Conor K. ;
Butterworth, Karl T. ;
Prise, Kevin M. ;
O'Sullivan, Joe M. ;
Hounsell, Alan R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (22) :N411-N420
[7]   Compensation for respiratory motion by gated radiotherapy:: an experimental study [J].
Dietrich, L ;
Tücking, T ;
Nill, S ;
Oelfke, U .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (10) :2405-2414
[8]   Dosimetric effect of respiration-gated beam on IMRT delivery [J].
Duan, J ;
Shen, S ;
Fiveash, JB ;
Brezovich, IA ;
Popple, RA ;
Pareek, PN .
MEDICAL PHYSICS, 2003, 30 (08) :2241-2252
[9]   Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imaging [J].
Ford, EC ;
Mageras, GS ;
Yorke, E ;
Rosenzweig, KE ;
Wagman, R ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (02) :522-531
[10]   Margin selection to compensate for loss of target dose coverage due to target motion during external-beam radiation therapy of the lung [J].
Foster, W. Kyle ;
Osei, Ernest ;
Barnett, Rob .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (01) :139-158