Intramedullary femoral nails: one or two lag screws? A preliminary study

被引:44
作者
Wang, CJ
Brown, CJ [1 ]
Yettram, AL
Procter, P
机构
[1] Brunel Univ, Dept Mech Engn, Uxbridge UB8 3PH, Middx, England
[2] Stryker Trauma SA, CH-1228 Geneva, Switzerland
关键词
fractured femur; proximal femoral nails; gamma nail; finite element method; lag screws; cut-out;
D O I
10.1016/S1350-4533(00)00081-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Failures of proximal femoral nails that treat unstable femoral fractures have been reported. In this communication, a finite element model to include a proximal femoral nail within a fractured femur was used to carry out preliminary investigations into configurations of single or double lag screws. The effects of the different types of fracture were investigated. The results show that in order to share the load evenly between two lag screws, a good configuration seems to be to have a slightly larger screw above the lower screw. This also ameliorates stresses in the nail at the lag screw insertion holes. However, using two screws in this way can lead to large stresses in the cancellous bone in the femoral head, and these stresses may be significant in the initiation of cut-our. (C) 2001 IPEM. Published by Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:613 / 624
页数:12
相关论文
共 36 条
[1]   Complications and technical problems with the gamma nail [J].
Albareda, J ;
Laderiga, A ;
Palanca, D ;
Paniagua, L ;
Seral, F .
INTERNATIONAL ORTHOPAEDICS, 1996, 20 (01) :47-50
[2]   THE VALUE OF THE TIP-APEX DISTANCE IN PREDICTING FAILURE OF FIXATION OF PERITROCHANTERIC FRACTURES OF THE HIP [J].
BAUMGAERTNER, MR ;
CURTIN, SL ;
LINDSKOG, DM ;
KEGGI, JM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (07) :1058-1064
[3]  
BRUNNER P, 1996, BIOMED TECH, V41, P30
[4]   TELEMETRIC FORCE MEASUREMENTS ACROSS THE HIP AFTER TOTAL ARTHROPLASTY [J].
DAVY, DT ;
KOTZAR, GM ;
BROWN, RH ;
HEIPLE, KG ;
GOLDBERG, VM ;
HEIPLE, KG ;
BERILLA, J ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (01) :45-50
[5]  
DEGLIGIANNI DD, 1991, J MAT SCI MAT MED, V2, P168
[6]  
EBERLE C, 1993, HELV CHIR ACTA, V59, P527
[7]   ROLE OF THE COLLAR ON THE FEMORAL STEM OF CEMENTED TOTAL HIP REPLACEMENTS [J].
FAGAN, MJ ;
LEE, AJC .
JOURNAL OF BIOMEDICAL ENGINEERING, 1986, 8 (04) :295-304
[8]  
Ford CM, 1996, J BONE MINER RES, V11, P377
[9]   A 3-DIMENSIONAL NONLINEAR FINITE-ELEMENT STUDY OF THE EFFECT OF CEMENT-PROSTHESIS DEBONDING IN CEMENTED FEMORAL TOTAL HIP COMPONENTS [J].
HARRIGAN, TP ;
HARRIS, WH .
JOURNAL OF BIOMECHANICS, 1991, 24 (11) :1047-1058
[10]   Comparative dynamic evaluation of the sliding/characteristics of the Gamma nail: a biomechanical analysis [J].
Haynes, RC ;
Miles, AW .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (05) :411-417