Subcutaneous implantable cardioverter-defibrillator and defibrillation testing: A propensity-matched pilot study

被引:17
作者
Forleo, Giovanni B. [1 ]
Gasperetti, Alessio [1 ,2 ,3 ]
Breitenstein, Alexander [4 ]
Laredo, Mikael [5 ]
Schiavone, Marco [1 ]
Ziacchi, Matteo [6 ]
Vogler, Julia [7 ]
Ricciardi, Danilo [8 ]
Palmisano, Pietro [9 ]
Piro, Agostino [10 ]
Compagnucci, Paolo [2 ]
Waintraub, Xavier [5 ]
Mitacchione, Gianfranco [11 ]
Carrassa, Gianmarco [12 ]
Russo, Giulia [13 ]
De Bonis, Silvana [14 ]
Angeletti, Andrea [6 ]
Bisignani, Antonio [14 ]
Picarelli, Francesco [8 ]
Casella, Michela [2 ]
Bressi, Edoardo [15 ]
Rovaris, Giovanni [16 ]
Calo, Leonardo [15 ]
Santini, Luca [17 ]
Pignalberi, Carlo [18 ]
Lavalle, Carlo [10 ]
Viecca, Maurizio [1 ]
Pisano, Ennio [13 ]
Olivotto, Iacopo [12 ]
Curnis, Antonio [11 ]
Dello Russo, Antonio [2 ]
Tondo, Claudio [19 ,20 ]
Love, Charles J. [3 ]
Di Biase, Luigi [21 ]
Steffel, Jan [4 ]
Tilz, Roland [7 ]
Badenco, Nicolas [5 ]
Biffi, Mauro [6 ]
机构
[1] Luigi Sacco Univ Hosp, Cardiol Unit, Milan, Italy
[2] Univ Hosp Umberto I Salesi Lancisi, Cardiol & Arrhythmol Clin, Ancona, Italy
[3] Johns Hopkins Univ, Div Cardiol, Baltimore, MD USA
[4] Zurich Univ Hosp, Cardiol Dept, Zurich, Switzerland
[5] Hop La Pitie Salpetriere, AP HP, Paris, France
[6] Univ Bologna, St Orsola Hosp, Cardiol Unit, Bologna, Italy
[7] Univ Hosp Lubeck, Cardiol Dept, Lubeck, Germany
[8] Campus Biomed, Cardiol Dept, Rome, Italy
[9] Tricase Hosp, Cardiol Dept, Tricase, Italy
[10] Policlin Umberto 1, Cardiol Dept, Rome, Italy
[11] Spedali Civili Brescia, Cardiol Dept, Brescia, Italy
[12] Careggi Univ Hosp, Cardiomyopathy Unit, Florence, Italy
[13] Vito Fazzi Hosp, Cardiol Dept, Lecce, Italy
[14] Ferrari Hosp, Cardiol Dept, Cosenza, Italy
[15] Policlin Casilino, Cardiol Dept, Rome, Italy
[16] San Gerardo Hosp, Cardiol Dept, Monza, Italy
[17] Osped GB Grassi, Cardiol Dept, Ostia, Italy
[18] Osped San Filippo Neri, Cardiol Dept, Rome, Italy
[19] Univ Milan, Dept Biomed Surg & Dent Sci, Milan, Italy
[20] Ctr Cardiol Monzino IRCCS, Dept Clin Electrophysiol & Cardiac Pacing, Milan, Italy
[21] Montefiore Med Ctr, Albert Einstein Coll Med, Bronx, NY 10467 USA
关键词
Defibrillation testing; PRAETORIAN score; Propensity matching; Subcutaneous implantable cardioverter-defibrillator; Sudden cardiac death; VENTRICULAR DEFIBRILLATION; EFFICACY;
D O I
10.1016/j.hrthm.2021.06.1201
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND To date, only a few comparisons between subcutaneous implantable cardioverter-defibrillator (S-ICD) patients undergoing and those not undergoing defibrillation testing (DT) at implantation (DT+ vs DT-) have been reported. OBJECTIVE The purpose of this study was to compare long-term clinical outcomes of 2 propensity-matched cohorts of DT+ and DT- patients. METHODS Among consecutive S-ICD patients implanted across 17 centers from January 2015 to October 2020, DT-patients were 1:1 propensity-matched for baseline characteristics with DT+ patients. The primary outcome was a composite of ineffective shocks and cardiovascular mortality. Appropriate and inappropriate shock rates were deemed secondary outcomes. RESULTS Among 1290 patients, a total of 566 propensity-matched patients (283 DT+; 283 DT-) served as study population. Over median follow-up of 25.3 months, no significant differences in primary outcome event rates were found (10 DT+ vs 14 DT-; P =.404) as well as for ineffective shocks (5 DT- vs 3 DT+; P =.725). At multivariable Cox regression analysis, DT performance was associated with a reduction of neither the primary combined outcome nor ineffective shocks at follow-up. A high PRAETORIAN score was positively associated with both the primary outcome (hazard ratio 3.976; confidence interval 1.339-11.802; P =.013) and ineffective shocks alone at follow-up (hazard ratio 19.030; confidence interval 4.752-76.203; P =.003). CONCLUSION In 2 cohorts of strictly propensity-matched patients, DT performance was not associated with significant differences in cardiovascular mortality and ineffective shocks. The PRAETORIAN score is capable of correctly identifying a large percentage of patients at risk for ineffective shock conversion in both cohorts.
引用
收藏
页码:2072 / 2079
页数:8
相关论文
共 18 条
[1]  
Al-Ghamdi B, 2017, CARDIOL RES, V8, P319, DOI 10.14740/cr638w
[2]   Intra-operative defibrillation testing and clinical shock efficacy in patients with implantable cardioverter-defibrillators: the NORDIC ICD randomized clinical trial [J].
Baensch, Dietmar ;
Bonnemeier, Hendrik ;
Brandt, Johan ;
Bode, Frank ;
Svendsen, Jesper Hastrup ;
Taborsky, Milos ;
Kuster, Stefan ;
Blomstrom-Lundqvist, Carina ;
Felk, Angelika ;
Hauser, Tino ;
Suling, Anna ;
Wegscheider, Karl .
EUROPEAN HEART JOURNAL, 2015, 36 (37) :2500-2507
[3]   Is 40 Joules Enough to Successfully Defibrillate With Subcutaneous Implantable Cardioverter-Defibrillators? [J].
Biffi, Mauro ;
Bongiorni, Maria Grazia ;
D'Onofrio, Antonio ;
Manzo, Michele ;
Pieragnoli, Paolo ;
Palmisano, Pietro ;
Ottaviano, Luca ;
Perego, Giovanni Battista ;
Pangallo, Antonio ;
Lavalle, Carlo ;
Bonfantino, Vincenzo ;
Nigro, Gerardo ;
Landolina, Maurizio Eugenio ;
Katsouras, Grigorios ;
Diemberger, Igor ;
Viani, Stefano ;
Bianchi, Valter ;
Lovecchio, Mariolina ;
Valsecchi, Sergio ;
Ziacchi, Matteo .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2021, 7 (06) :767-776
[4]   Defibrifladon testing at the time of tmp antatlon of cardoverter defibrillator in the dinica practice: a nadon-wide survey [J].
Brignole, Michele ;
Raciti, Giovanni ;
Bongiorni, Maria Grazia ;
De Martino, Giuseppe ;
Favale, Stefano ;
Gasparini, Maurizio ;
Luise, Raffaele ;
Occhetta, Eraldo ;
Proclemer, A. .
EUROPACE, 2007, 9 (07) :540-543
[5]   Defibrillation Testing Should the Paradigm Shift? [J].
Estes, N. A. Mark, III .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (11) :988-989
[6]   Implantation technique and optimal subcutaneous defibrillator chest position: a PRAETORIAN score-based study [J].
Francia, Pietro ;
Biffi, Mauro ;
Adduci, Carmen ;
Ottaviano, Luca ;
Migliore, Federico ;
De Bonis, Silvana ;
Dello Russo, Antonio ;
De Filippo, Paolo ;
Viani, Stefano ;
Bongiorni, Maria Grazia ;
Caravati, Fabrizio ;
Lavalle, Carlo ;
Landolina, Maurizio Eugenio ;
Pisano, Ennio ;
Giorgi, Davide ;
Lovecchio, Mariolina ;
Valsecchi, Sergio ;
Diemberger, Igor .
EUROPACE, 2020, 22 (12) :1822-1829
[7]   Trends and In-Hospital Outcomes Associated With Adoption of the Subcutaneous Implantable Cardioverter Defibrillator in the United States [J].
Friedman, Daniel J. ;
Parzynski, Craig S. ;
Varosy, Paul D. ;
Prutkin, Jordan M. ;
Patton, Kristen K. ;
Mithani, Ali ;
Russo, Andrea M. ;
Curtis, Jeptha P. ;
Al-Khatib, Sana M. .
JAMA CARDIOLOGY, 2016, 1 (08) :900-911
[8]   Efficacy and temporal stability of reduced safety margins for ventricular defibrillation - Primary results from the Low Energy Safety Study (LESS) [J].
Gold, MR ;
Higgins, S ;
Klein, R ;
Gilliam, FR ;
Kopelman, H ;
Hessen, S ;
Payne, J ;
Strickberger, SA ;
Breiter, D ;
Hahn, S .
CIRCULATION, 2002, 105 (17) :2043-2048
[9]   Cardioverter defibrillator implantation without induction of ventricular fibrillation: a single-blind, non-inferiority, randomised controlled trial (SIMPLE) [J].
Healey, Jeff S. ;
Hohnloser, Stefan H. ;
Glikson, Michael ;
Neuzner, Jorg ;
Mabo, Phillipe ;
Vinolas, Xavier ;
Kautzner, Josef ;
O'Hara, Gilles ;
VanErven, Lieselot ;
Gadler, Fredrik ;
Pogue, Janice ;
Appl, Ursula ;
Gilkerson, Jim ;
Pochet, Thierry ;
Stein, Kenneth M. ;
Merkely, Bela ;
Chrolavicius, Susan ;
Meeks, Brandi ;
Foldesi, Csaba ;
Thibault, Bernard ;
Connolly, Stuart J. .
LANCET, 2015, 385 (9970) :785-791
[10]   Defibrillation threshold testing fails to show clinical benefit during long-term follow-up of patients undergoing cardiac resynchronization therapy defibrillator implantation [J].
Michowitz, Yoav ;
Lellouche, Nicolas ;
Contractor, Tahmeed ;
Bourke, Tara ;
Wiener, Isaac ;
Buch, Eric ;
Boyle, Noel ;
Bersohn, Malcolm ;
Shivkumar, Kalyanam .
EUROPACE, 2011, 13 (05) :683-688