Technological capabilities of surface layers formation on implant made of Ti-6Al-4V ELI alloy

被引:25
作者
Kiel-Jamrozik, Marta [1 ]
Szewczenko, Janusz [1 ]
Basiaga, Marcin [1 ]
Nowinska, Katarzyna [2 ]
机构
[1] Silesian Tech Univ, Fac Biomed Engn, Dept Biomat & Med Devices Engn, PL-41800 Zabrze, Poland
[2] Silesian Tech Univ, Fac Min & Geol, Inst Appl Geol, Gliwice, Poland
关键词
corrosion resistance; ion infiltration; metallic biomaterials; surface modification; titanium alloys; TITANIUM-ALLOYS; TI6AL4V;
D O I
10.5277/ABB-00065-2014-03
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Purpose: The aim of the presented research was to find a combination of surface modification methods of implants made of the Ti-6Al-4V ELI alloy, that lead to formation of effective barrier for metallic ions that may infiltrate into solution. Methods: To this end, the following tests were carried out: roughness measurement, the voltamperometric tests (potentiodynamic and potentiostatic), and the ion infiltration test. Results: The electropolishing process resulted in the lowering of surface roughness in comparison with mechanical treatment of the surface layer. The anodization process and steam sterilization increased corrosion resistance regardless of the mechanical treatment or electropolishing. The crevice corrosion tests revealed that independent of the modification method applied, the Ti-6Al-4V ELI alloy has excellent crevice corrosion resistance. The smallest quantity of ions infiltrated to the solution was observed for surface modification consisting in the mechanical treatment and anodization with the potential of 97 V. Conclusions: Electric parameters determined during studies were the basis for effectiveness estimation of particular surface treatment methods. The research has shown that the anodization process significantly influences the pitting corrosion resistance of the Ti-6Al-4V ELI alloy independent of the previous surface treatment methods (mechanical and electrochemical). The surface layer after such modification is a protective barrier for metallic ions infiltrated to solution and protects titanium alloy against corrosive environment influence.
引用
收藏
页码:31 / 37
页数:7
相关论文
共 22 条
[1]  
Al-Mobarak NA, 2014, INT J ELECTROCHEM SC, V9, P32
[2]  
[Anonymous], 2009, 4287 PNEN ISO
[3]   Effect of different surface finishing and of hydroxyapatite coatings on passive and corrosion current of Ti6Al4V alloy in simulated physiological solution [J].
Cabrini, M ;
Cigada, A ;
Rondelli, G ;
Vicentini, B .
BIOMATERIALS, 1997, 18 (11) :783-787
[4]  
Fadl-allah S.A., 2013, J BIOMATERIALS NANOB, V4, P74, DOI DOI 10.4236/JBNB.2013.41010
[5]   Corrosion behavior of electrochemically assembled nanoporous titania for biomedical applications [J].
Indira, K. ;
Mudali, U. Kamachi ;
Rajendran, N. .
CERAMICS INTERNATIONAL, 2013, 39 (02) :959-967
[6]  
Kiel M, 2012, LECT N BIOINFORMAT, V7339, P369
[7]  
Kiel-Jamrozik M, 2012, PRZ ELEKTROTECHNICZN, V88, P232
[8]  
Krasicka-Cydzik E., 2012, ACHIEVING ENHANCED P, P175
[9]   Surface modification of titanium, titanium alloys, and related materials for biomedical applications [J].
Liu, XY ;
Chu, PK ;
Ding, CX .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2004, 47 (3-4) :49-121
[10]  
Marciniak J., 2008, BIOMATERIALY STOMATO