Miniaturized Endoscopic Probe for Optical Coherence Tomography with a Tiny Magnet Driving Device

被引:0
作者
Peng, Ziwei [1 ]
Wu, Jigang [1 ]
机构
[1] Shanghai Jiao Tong Univ, Joint Inst, Univ Michigan, Biophoton Lab, Shanghai 200240, Peoples R China
来源
OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS VI | 2014年 / 9268卷
关键词
optical coherence tomography; endoscopic probe;
D O I
10.1117/12.2071433
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
This paper proposes an endoscopic probe for optical coherence tomography (OCT) applying on side-imaging of internal organs. The probe consists of single-mode fiber, a gradient-index (GRIN) lens, and a mirror of cylindrical wedge shape attaching to a magnetized metal piece by epoxy with a short steel wire. For OCT scanning, we use magnetic field generated by a larger magnet externally to drive the rotation of the magnetized metal. Compare with other probes, our probe design has two distinct advantages: 1) The exit beam will be unobstructed during 360 degree circumference scanning because there are no connecting wires in the scanning part. 2) In principle, the probe can be made very tiny because of the simple structure consisting only the single-mode fiber, GRIN lens, reflection mirror and the magnetized metal piece. The OCT system has axial resolution of 14 mu m and SNR of 98.6 dB. The probe prototype we made has an outer diameter of 1.4 mm.
引用
收藏
页数:6
相关论文
共 9 条
[1]  
[Anonymous], 2008, OPTICAL COHERENCE TO
[2]   Stationary-fiber rotary probe with unobstructed 360° view for optical coherence tomography [J].
Chang, Shoude ;
Murdock, Erroll ;
Mao, Youxin ;
Flueraru, Costel ;
Disano, John .
OPTICS LETTERS, 2011, 36 (22) :4392-4394
[3]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[4]   Ultrathin side-viewing needle probe for optical coherence tomography [J].
Lorenser, D. ;
Yang, X. ;
Kirk, R. W. ;
Quirk, B. C. ;
McLaughlin, R. A. ;
Sampson, D. D. .
OPTICS LETTERS, 2011, 36 (19) :3894-3896
[5]  
Piyawattanametha Wibool, 2013, 2013 International Conference on Optical MEMS and Nanophotonics (OMN), P53, DOI 10.1109/OMN.2013.6659055
[6]   In vivo three-dimensional microelectromechanical endoscopic swept source optical coherence tomography [J].
Su, Jianping ;
Zhang, Jun ;
Yu, Linfeng ;
Chen, Zhongping .
OPTICS EXPRESS, 2007, 15 (16) :10390-10396
[7]   Three-Dimensional Coronary Artery Microscopy by Intracoronary Optical Frequency Domain Imaging [J].
Tearney, Guillermo J. ;
Waxman, Sergio ;
Shishkov, Milen ;
Vakoc, Benjamin J. ;
Suter, Melissa J. ;
Freilich, Mark I. ;
Desjardins, Adrien E. ;
Oh, Wang-Yul ;
Bartlett, Lisa A. ;
Rosenberg, Mireille ;
Bouma, Brett E. .
JACC-CARDIOVASCULAR IMAGING, 2008, 1 (06) :752-761
[8]   Intravascular optical coherence tomography imaging at 3200 frames per second [J].
Wang, Tianshi ;
Wieser, Wolfgang ;
Springeling, Geert ;
Beurskens, Robert ;
Lancee, Charles T. ;
Pfeiffer, Tom ;
van der Steen, Antonius F. W. ;
Huber, Robert ;
van Soest, Gijs .
OPTICS LETTERS, 2013, 38 (10) :1715-1717
[9]   Molecular contrast optical coherence tomography: a pump-probe scheme using indocyanine green as a contrast agent [J].
Yaqoob, Zahid ;
McDowell, Emily ;
Wu, Jigang ;
Heng, Xin ;
Fingler, Jeff ;
Yang, Changhuei .
JOURNAL OF BIOMEDICAL OPTICS, 2006, 11 (05)