High-precision analysis of calcium isotopes using a Nu Sapphire collision cell (CC)-MC-ICP-MS

被引:13
|
作者
Gao, Bing-Yu [1 ]
Su, Ben-Xun [1 ,2 ]
Li, Wen-Jun [1 ]
Yuan, Meng [1 ,2 ]
Sun, Jian [3 ]
Zhao, Ye [4 ]
Liu, Xia [1 ,2 ]
机构
[1] Chinese Acad Sci, Inst Geol & Geophys, Key Lab Mineral Resources, Beijing 100029, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
[3] Chinese Acad Geol Sci, Inst Geol, MNR Key Lab Isotope Geol, Key Lab Deep Earth Dynam,Minist Nat Resources, Beijing 100037, Peoples R China
[4] Nu Instruments, Unit 74 Clywedog Rd South,Wrexham Ind Estate, Wrexham LL13 9XS, Wales
关键词
MC-ICP-MS; CA ISOTOPES; IGNEOUS ROCKS; FRACTIONATION; MANTLE; DELTA-CA-44/40; CONSTRAINTS; SYSTEMATICS; CARBONATES; INSIGHTS;
D O I
10.1039/d2ja00150k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study presents high-precision analyses of calcium (Ca) isotope ratios using a collision-cell multi-collector inductively coupled plasma mass spectrometer (CC-MC-ICP-MS, Nu Sapphire). This instrument allows Ca isotope measurement in the low resolution mode, requiring much less sample compared with TIMS and conventional MC-ICP-MS, and has additional measurement capability of both radiogenic (epsilon Ca-40) and stable isotope ratios (delta Ca-44/40 and delta Ca-44/42). The effects of several factors on the precision and accuracy of Ca isotopic data were evaluated, including the total Ca concentration, Ca intensity mismatch between the sample and standard, HNO3 molarity mismatch between the sample and standard, and the presence of matrix elements. High precision was obtained on 100 ng g(-1) Ca samples (i.e., for 6 replicates similar to 600 ng of Ca). Accurate measurements require the following conditions: (1) Ca intensities matched within 3% between the sample and standard, (2) low contents of Sr, Mg and Cr (Sr/Ca, Mg/Ca and Cr/Ca ratios need to be lower than 0.01%, and K/Ca, Al/Ca, Ti/Ca and Cr/Ca ratios <1%). Eleven standard reference materials analyzed using the established method yield highly consistent Ca isotope results with literature values. The long-term external precisions are higher than 0.07 parts per thousand (2 SD) for delta Ca-44/40 and higher than 0.05 parts per thousand (2 SD) for delta Ca-44/42. The slope of the fractionation line between delta Ca-44/40 and delta Ca-44/42 is 1.9616 +/- 0.0794, within the error of the slope value from the exponential law.
引用
收藏
页码:2111 / 2121
页数:11
相关论文
共 50 条
  • [41] High precision analysis of potassium stable isotopes using the collision/reaction cell Neoma MC-ICPMS/MS
    Nie, Nicole X.
    Grigoryan, Rosa
    Tissot, Francois L. H.
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (08) : 2038 - 2048
  • [42] Stable silicon isotope analysis on nanomole quantities using MC-ICP-MS with a hexapole gas-collision cell
    Sun, Xiaole
    Andersson, Per
    Land, Magnus
    Humborg, Christoph
    Morth, Carl-Magnus
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2010, 25 (02) : 156 - 162
  • [43] High-Precision Mass-Dependent Molybdenum Isotope Variations in Magmatic Rocks Determined by Double-Spike MC-ICP-MS
    Willbold, Matthias
    Hibbert, Kate
    Lai, Yi-Jen
    Freymuth, Heye
    Hin, Remco C.
    Coath, Christopher
    Vils, Flurin
    Elliott, Tim
    GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2016, 40 (03) : 389 - 403
  • [44] High-precision zinc isotopic characterization of twenty soil reference materials from China determined by MC-ICP-MS
    Zhai, Hua-Ye
    Wang, Xuan-Ce
    Li, Chao-Feng
    Wilde, Simon A.
    Li, Xiang-Zhong
    Xu, Bei
    Zhang, Xian-Liang
    Zhang, Ping
    RSC ADVANCES, 2023, 13 (28) : 19030 - 19038
  • [45] Precise measurement of stable neodymium isotopes of geological materials by using MC-ICP-MS
    Ma, Jinlong
    Wei, Gangjian
    Liu, Ying
    Ren, Zhongyuan
    Xu, Yigang
    Yang, Yonghong
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2013, 28 (12) : 1926 - 1931
  • [46] Precise determination of triple Sr isotopes (δ87Sr and δ88Sr) using MC-ICP-MS
    Liu, Hou-Chun
    You, Chen-Feng
    Huang, Kuo-Fang
    Chung, Chuan-Hsiung
    TALANTA, 2012, 88 : 338 - 344
  • [47] High intermediate precision Sm isotope measurements in geological samples by MC-ICP-MS
    Bai, Jiang-Hao
    Lin, Mang
    Zhong, Song-Xiong
    Deng, Yi-Nan
    Zhang, Le
    Luo, Kai
    Wu, Hao
    Ma, Jin-Long
    Wei, Gang-Jian
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2023, 38 (03) : 629 - 637
  • [48] High precision measurements of gallium isotopic compositions in geological materials by MC-ICP-MS
    Zhang, Ting
    Zhou, Lian
    Yang, Lu
    Wang, Qian
    Feng, Lan-ping
    Liu, Yong-Sheng
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2016, 31 (08) : 1673 - 1679
  • [49] High precision determination of silver isotope ratios in commercial products by MC-ICP-MS
    Yang, Lu
    Dabek-Zlotorzynska, Ewa
    Celo, Valbona
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (11) : 1564 - 1569
  • [50] A two-stage column protocol for the separation of Cu from geological materials and high-precision Cu isotopic analysis on a MC-ICP-MS
    Li, Jie
    Zhong, Qiao-Hui
    Zhang, Le
    Yu, Hong-Xia
    Wang, Zhao-Yang
    Bao, Hou-Yin
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2022, 37 (04) : 849 - 860