Evaluation of an Ion-Exchange HPLC Device for HbA1c Measurement

被引:6
作者
Altawallbeh, Ghaith [1 ]
Makky, Vicky F. [1 ]
Saenger, Amy K. [1 ]
Peters, Jennifer M. [1 ]
Killeen, Anthony A. [1 ]
机构
[1] Univ Minnesota, Dept Lab Med & Pathol, 420 Delaware St SE,MMC 609, Minneapolis, MN 55455 USA
关键词
HEMOGLOBIN A(1C); S TRAITS; HBA(1C);
D O I
10.1093/jalm/jfaa042
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The ADAMS (TM) HA-8180V is the 8th generation of a fully automated ion-exchange HPLC system from ARKRAY, and the first to be released onto the US market. We evaluated the HA-8180V, for routine hemoglobin A1c measurement in comparison with the Roche Cobas c501, the Tosoh G8 analyzer for normal hemoglobin, and with the Trinity analyzer for hemoglobin variants. Methods: The analytical performance (linearity, precision, carryover, and sample stability) was assessed based on the Clinical and Laboratory Standards Institute (CLSI) and manufacturer guidelines. A comparison of the HA-8180V against two major analytical methods was performed for 100 whole blood samples. HA-8180V variant mode was also compared against the Trinity ultra2 A1c analyzer for 50 samples containing hemoglobin variants (HbC 14, HbS 14, HbD 12, and HbE 10). Results: The within-run and total CVs were <0.01 and 0.75% at low HbA1c concentration and 0.46 and 0.63% at high HbA1c concentration, respectively. Linearity was shown in the concentration range 3.4-18.1% HbA1c, carryover was 0.00%, and stability values were excellent. Method comparison demonstrated a high concordance between methods. Conclusion: The eighth generation ADAMS HA-8180V A1c analyzer demonstrated high analytical performance adequate for routine clinical use.
引用
收藏
页码:695 / 703
页数:9
相关论文
共 16 条
[3]  
[Anonymous], 2016, ISBN, V978, P88
[4]   Comparison of Arkray/ELITech ADAMS HA-8180V® with Bio-Rad Variant,™II Turbo2.0® and Tosoh Bioscience HLC®-723G8 for HbA1c Determination [J].
Badiou, Stephanie ;
Guillot, Julie ;
Kuster, Nils ;
Bargnoux, Anne-Sophie ;
Aguilar-Martinez, Patricia ;
Boissier, Elodie ;
Cristol, Jean-Paul ;
Dupuy, Anne-Marie .
JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2014, 28 (06) :428-434
[5]   Type 2 diabetes [J].
Chatterjee, Sudesna ;
Khunti, Kamlesh ;
Davies, Melanie J. .
LANCET, 2017, 389 (10085) :2239-2251
[6]  
CLSI, 2014, EP 15 A3 USER VERIFI
[7]  
CLSI, 2014, EP10A3AMD CLSI, V3rd ed
[8]  
CLSI, 2003, CLSI document EP06
[9]   Evaluation of enzymatic BM Test HbA1c on the JCA-BM6010/C and comparison with Bio-Rad Variant II Turbo, Tosoh HLC 723 G8, and AutoLab immunoturbidimetry assay [J].
Lee, Jehoon ;
Kim, Myungshin ;
Chae, Hyojin ;
Kim, Yeongsic ;
Park, Hae-il ;
Kim, Yonggoo ;
Chi, Hyunyoung ;
Kwon, Hi Jeong .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (11) :2201-2208
[10]   Effects of hemoglobin C, D, E, and S traits on measurements of HbA1c by six methods [J].
Lin, Chia-Ni ;
Emery, Todd J. ;
Little, Randie R. ;
Hanson, Steve E. ;
Rohlfing, Curt L. ;
Jaisson, Stephane ;
Gillery, Philippe ;
Roberts, William L. .
CLINICA CHIMICA ACTA, 2012, 413 (7-8) :819-821