Peak clustering in two-dimensional gas chromatography with mass spectrometric detection based on theoretical calculation of two-dimensional peak shapes: The 2DAid approach

被引:6
作者
van Stee, Leo L. P. [1 ]
Brinkman, Udo A. Th. [1 ]
机构
[1] Free Univ Amsterdam, Dept Analyt Chem, NL-1081 HV Amsterdam, Netherlands
关键词
Comprehensive two-dimensional gas chromatography; Mass spectrometry; Data processing; Peak detection; CHEMOMETRIC ANALYSIS; METABOLITES; SEPARATIONS; PREDICTION; TIMES;
D O I
10.1016/j.chroma.2011.08.081
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A method is presented to facilitate the non-target analysis of data obtained in temperature-programmed comprehensive two-dimensional (2D) gas chromatography coupled to time-of-flight mass spectrometry (GC x GC-ToF-MS). One main difficulty of GC x GC data analysis is that each peak is usually modulated several times and therefore appears as a series of peaks (or peaklets) in the one-dimensionally recorded data. The proposed method, 2DAid, uses basic chromatographic laws to calculate the theoretical shape of a 2D peak (a cluster of peaklets originating from the same analyte) in order to define the area in which the peaklets of each individual compound can be expected to show up. Based on analyte-identity information obtained by means of mass spectral library searching, the individual peaklets are then combined into a single 2D peak. The method is applied, amongst others, to a complex mixture containing 362 analytes. It is demonstrated that the 2D peak shapes can be accurately predicted and that clustering and further processing can reduce the final peak list to a manageable size. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:7878 / 7885
页数:8
相关论文
共 22 条
[1]   Flow regime at ambient outlet pressure and its influence in comprehensive two-dimensional gas chromatography [J].
Beens, J ;
Janssen, HG ;
Adahchour, M ;
Brinkman, UAT .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1086 (1-2) :141-150
[2]   Simple, non-moving modulation interface for comprehensive two-dimensional gas chromatography [J].
Beens, J ;
Adahchour, M ;
Vreuls, RJJ ;
van Altena, K ;
Brinkman, UAT .
JOURNAL OF CHROMATOGRAPHY A, 2001, 919 (01) :127-132
[3]  
Blumberg L.M., 1997, J HIGH RES CHROMATOG, V20, P597, DOI DOI 10.1002/JHRC.1240201106
[4]   Comparison of the behaviour of gas-liquid and gas-liquid-solid capillary columns through the determination of thermodynamic characteristics [J].
Castello, G ;
Vezzani, S ;
Moretti, P .
JOURNAL OF CHROMATOGRAPHY A, 1996, 742 (1-2) :151-160
[5]   CONCEPTS AND COMPARISONS IN MULTIDIMENSIONAL SEPARATION [J].
GIDDINGS, JC .
JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1987, 10 (05) :319-323
[6]  
GUIOCHON G, 1966, CHROMATOGR REV, V8, P1
[7]   Automated resolution of nontarget analyte signals in GC x GC-TOFMS data using parallel factor analysis [J].
Hoggard, Jamin C. ;
Synovec, Robert E. .
ANALYTICAL CHEMISTRY, 2008, 80 (17) :6677-6688
[8]   Comprehensive two-dimensional gas chromatography with time-of-flight mass spectrometry detection: analysis of amino acid and organic acid trimethylsilyl derivatives, with application to the analysis of metabolites in rye grass samples [J].
Hope, JL ;
Prazen, BJ ;
Nilsson, EJ ;
Lidstrom, ME ;
Synovec, RE .
TALANTA, 2005, 65 (02) :380-388
[9]  
Johnson KJ, 2002, J SEP SCI, V25, P297, DOI 10.1002/1615-9314(20020401)25:5/6<297::AID-JSSC297>3.0.CO
[10]  
2-I