Primary stability of threaded cups in THR - An experimental study

被引:8
作者
Schwarz, MLR
Scheller, G
Effenberger, H
机构
[1] Univ Mannheim, Orthopad Klin, Lab Biomech & Expt Orthopad, D-68167 Mannheim, Germany
[2] Goethe Univ Frankfurt, Orthopad Klin, D-6000 Frankfurt, Germany
[3] Poliklin Friedrichsteim, Frankfurt, Germany
来源
BIOMEDIZINISCHE TECHNIK | 2003年 / 48卷 / 12期
关键词
threaded cups; primary stability; lever-out moment; torque-in moment; constructive features;
D O I
10.1515/bmte.2003.48.12.334
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Four threaded cups were tested up to their lever-out moments, torque-in moments and their resistance to failure. A was a parabolic-shaped, B was a spherical, C was a spherical-shaped too, and D was a conical shaped cup. Cup A and D represent cups which have proven themselves in clinical applications, but not cup B. The threads were determined and showed different constructive features. The cups were torqued into precise cavities in PVC foam cubes, after that they were levered out in a testing machine. The lever-out moments of all the cups showed significant differences; the results were: A: 78.4 Nm, B: 88.7 Nm, C: 117.5 Nm, D: 136.6 Nm. In the case of the torque-in moments there were no significant differences between A and B, neither between C and D. The differences in stiffness between B and C were not significant, but they were between the others. The primary stability against lever-out and the torque-in moment of threaded cups for artificial hip replacement can be basically influenced by different constructive features. Hence lever-out moment and torque-in moment should be understood and tested as independent variables.
引用
收藏
页码:334 / 338
页数:5
相关论文
共 18 条
[1]  
Adler E, 1992, J Arthroplasty, V7, P295, DOI 10.1016/0883-5403(92)90052-R
[2]  
ALTMANN D, 1996, PRACTICAL STAT MED R
[3]  
Benjamin J, 1997, ORTHOPEDICS, V20, P307
[4]  
BLASCHKE H, 1998, PRIM U DOZ DR ULT DO
[5]  
BLOMER W, 1992, ZEMENTLOSE HUFTPROTH
[6]   Survivorship of rough-surfaced threaded acetabular cups - 382 consecutive primary Zweymuller cups followed for 0.2-12 years [J].
Delaunay, CP ;
Kapandji, AI .
ACTA ORTHOPAEDICA SCANDINAVICA, 1998, 69 (04) :379-383
[7]   Experimental study of bone-implant contact with a parabolic acetabular component (Hofer-Imhof) [J].
Effenberger, H ;
Böhm, G ;
Huber, M ;
Lintner, F ;
Hofer, H .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2000, 120 (3-4) :160-165
[8]  
EFFENBERGER H, 1991, NONCEMENTED TOTAL HI
[9]  
EFFENBERGER H, 1997, THREADED CUP DESIGN, V21, P232
[10]  
Fahrmeir L., 1997, STAT WEG DATENANALYS