Quantitative analysis of common elements in steel using a handheld μ-LIBS instrument

被引:71
作者
Afgan, Muhammad Sher [1 ]
Hou, Zongyu [1 ]
Wang, Zhe [1 ]
机构
[1] Tsinghua Univ, Dept Thermal Engn, Tsinghua BP Clean Energy Ctr, Beijing 100084, Peoples R China
基金
中国国家自然科学基金;
关键词
INDUCED BREAKDOWN SPECTROSCOPY; SPECTRUM STANDARDIZATION METHOD; PARTIAL LEAST-SQUARES; MICROCHIP LASER; CARBON ANALYSIS; MN; SPECTROMETRY; IMPROVEMENT; CHROMIUM; NICKEL;
D O I
10.1039/c7ja00219j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Laser-induced breakdown spectroscopy (LIBS) was named a future superstar. Nevertheless, in the last couple of years, several commercial instruments of LIBS and its different variants have become available on a limited scale due to the confinement of quantitative analysis. In this paper, quantitative analysis of various steel samples using a specially designed handheld mu-LIBS instrument and dominant factor based partial-least squares (PLS) regression with spectral standardization is presented. Average absolute measurement errors of 0.019, 0.039, 0.013 and 0.001% for the elements of Si, Cr, Mn and Ni, respectively, were achieved. The average relative standard deviation (RSD) for these elements was less than 5%. A linear calibration response was obtained for C which is difficult to achieve using traditional PLS and for Cu whose concentration in the samples is very low. The results are better than the results of previous LIBS studies using mu J lasers or portable handheld devices, comparable to those obtained using conventional benchtop LIBS systems and better than those of commercial handheld XRF instruments. The present work presents the best results for steel analysis using portable LIBS systems so far and shows the feasibility of the technology for real industrial applications.
引用
收藏
页码:1905 / 1915
页数:11
相关论文
共 36 条
[1]  
[Anonymous], LASER INDUCED BREAKD
[2]   Atomic spectrometry update. Review of advances in the analysis of metals, chemicals and functional materials [J].
Carter, Simon ;
Fisher, Andy ;
Garcia, Raquel ;
Gibson, Bridget ;
Lancaster, Steve ;
Marshallf, John ;
Whiteside, Ian .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2015, 30 (11) :2249-2294
[3]   Application of Handheld Laser-Induced Breakdown Spectroscopy (LIBS) to Geochemical Analysis [J].
Connors, Brendan ;
Somers, Andrew ;
Day, David .
APPLIED SPECTROSCOPY, 2016, 70 (05) :810-815
[4]  
Cremers DA, 2013, HANDBOOK OF LASER-INDUCED BREAKDOWN SPECTROSCOPY, 2ND EDITION, P1, DOI 10.1002/9781118567371
[5]   Recent progress on the application of LIBS for metallurgical online analysis in China [J].
Dong, Feng-Zhong ;
Chen, Xing-Long ;
Wang, Qi ;
Sun, Lan-Xiang ;
Yu, Hai-Bin ;
Liang, Yun-Xian ;
Wang, Jing-Ge ;
Ni, Zhi-Bo ;
Du, Zhen-Hui ;
Ma, Yi-Wen ;
Lu, Ji-Dong .
FRONTIERS OF PHYSICS, 2012, 7 (06) :679-689
[6]   Influence of sample temperature on the expansion dynamics and the optical emission of laser-induced plasma [J].
Eschlboeck-Fuchs, S. ;
Haslinger, M. J. ;
Hinterreiter, A. ;
Kolmhofer, P. ;
Huber, N. ;
Roessler, R. ;
Heitz, J. ;
Pedarnig, J. D. .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2013, 87 :36-42
[7]   A PLS model based on dominant factor for coal analysis using laser-induced breakdown spectroscopy [J].
Feng, Jie ;
Wang, Zhe ;
West, Logan ;
Li, Zheng ;
Ni, Weidou .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (10) :3261-3271
[8]   A compact and low cost laser induced breakdown spectroscopic system: Application for simultaneous determination of chromium and nickel in steel using multivariate calibration [J].
Gonzaga, Fabiano Barbieri ;
Pasquini, Celio .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2012, 69 :20-24
[9]   Microchip laser-induced breakdown spectroscopy: A preliminary feasibility investigation [J].
Gornushkin, IB ;
Amponsah-Manager, K ;
Smith, BW ;
Omenetto, N ;
Winefordner, JD .
APPLIED SPECTROSCOPY, 2004, 58 (07) :762-769
[10]   Accuracy improvement of quantitative analysis by spatial confinement in laser-induced breakdown spectroscopy [J].
Guo, L. B. ;
Hao, Z. Q. ;
Shen, M. ;
Xiong, W. ;
He, X. N. ;
Xie, Z. Q. ;
Gao, M. ;
Li, X. Y. ;
Zeng, X. Y. ;
Lu, Y. F. .
OPTICS EXPRESS, 2013, 21 (15) :18188-18195