The pressure drop across the endotracheal tube in mechanically ventilated pediatric patients

被引:14
作者
Spaeth, Johannes [1 ]
Steinmann, Daniel [2 ]
Kaltofen, Heike [1 ]
Guttmann, Josef [1 ]
Schumann, Stefan [1 ]
机构
[1] Univ Med Ctr Freiburg, Dept Anesthesiol & Intens Care Med, D-79106 Freiburg, Germany
[2] Univ Med Ctr Freiburg, Dept Occupat Hlth, D-79106 Freiburg, Germany
关键词
equipment - tubes tracheal; anesthesia; -; pediatric; ventilation; mechanical; ventilation - airway pressure; airway; resistance; respiratory; mechanics; SPONTANEOUSLY BREATHING PATIENTS; INTRATRACHEAL PRESSURE; TRACHEAL PRESSURE; ADDITIONAL WORK; RESISTANCE; COMPENSATION; OBSTRUCTION; CONNECTORS; CHILDREN; SUPPORT;
D O I
10.1111/pan.12595
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BackgroundDuring mechanical ventilation, the airway pressure (P-aw) is usually monitored. However, P-aw comprises the endotracheal tube (ETT)-related pressure drop (P-ETT) and thus does not reflect the pressure in the patients' lungs. Therefore, monitoring of mechanical ventilation should be based on the tracheal pressure (P-trach). We systematically investigated potential factors influencing P-ETT in pediatric ETTs. MethodsIn this study, the flow-dependent pressure drop across pediatric ETTs from four manufacturers [2.0-4.5mm inner diameter (ID)] was estimated in a physical model of the upper airways. Additionally, P-ETT was examined with the ETTs shortened to 75% of their original length and at different curvatures. In nine healthy mechanically ventilated children (aged between 9days and 29months), P-trach was compared to P-aw. ResultsP(ETT) was nonlinearly flow dependent. Low IDs corresponded to high P-ETT. Differences between ETTs from different manufacturers were identified. Shortening of the ETTs' length by 25% reduced P-ETT on average by 14% of the value at original length. Ventilation frequency and tube curvature did not influence P-ETT to a relevant extent. In the pediatric patients, the root mean square deviation between P-aw and P-trach was 2.3cm H2O. ConclusionP(aw) and P-trach differ considerably (by P-ETT) during mechanical ventilation of pediatric patients. The ETTs' ID, tube length, and manufacturer type are significant factors for P-ETT and should be taken into account when P-aw is valuated. For this purpose, P-trach can be continuously calculated with good precision by means of the Rohrer approximation.
引用
收藏
页码:413 / 420
页数:8
相关论文
共 23 条
[11]   TRACHEAL TUBES AND CONNECTORS USED IN NEONATES - DIMENSIONS AND RESISTANCE TO BREATHING [J].
HATCH, DJ .
BRITISH JOURNAL OF ANAESTHESIA, 1978, 50 (09) :959-964
[12]   Good short-term agreement between measured and calculated tracheal pressure [J].
Lichtwarck-Aschoff, M ;
Helmer, A ;
Kawati, R ;
Lattuada, M ;
Sjöstrand, UH ;
Zügel, N ;
Guttmann, J ;
Hedenstierna, G .
BRITISH JOURNAL OF ANAESTHESIA, 2003, 91 (02) :239-248
[13]  
Lichtwarck-Aschoff M, 2003, PEDIATR RES, V53, P533, DOI 10.1203/01.PDR.0000053202.65876.51
[14]  
Loring E, 1979, LUNG, V156, P33
[15]   Resistance of pediatric and neonatal endotracheal tubes: Influence of flow rate, size, and shape [J].
Manczur, T ;
Greenough, A ;
Nicholson, GP ;
Rafferty, GF .
CRITICAL CARE MEDICINE, 2000, 28 (05) :1595-1598
[16]   Breathing pattern associated with respiratory comfort during automatic tube compensation and pressure support ventilation in normal subjects [J].
Mols, G ;
Rohr, E ;
Benzing, A ;
Haberthür, C ;
Geiger, K ;
Guttmann, J .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2000, 44 (03) :223-230
[17]   Ventilator Y-piece pressure compared with intratracheal airway pressure in healthy intubated children [J].
Nasiroglu O. ;
Weldon B.C. ;
Berman L.S. ;
Haque I.U. .
Journal of Clinical Monitoring and Computing, 2006, 20 (2) :95-100
[18]   PARTIAL OBSTRUCTION OF ENDOTRACHEAL-TUBES IN CHILDREN - INCIDENCE, ETIOLOGY, SIGNIFICANCE [J].
REDDING, GJ ;
FAN, L ;
COTTON, EK ;
BROOKS, JG .
CRITICAL CARE MEDICINE, 1979, 7 (05) :227-231
[19]   The streaming-resistance in human airways and the influence of irregular branching of the bronchial system on the breathing-distribution in differing lung-precincts. [J].
Rohrer, F .
PFLUGERS ARCHIV FUR DIE GESAMTE PHYSIOLOGIE DES MENSCHEN UND DER TIERE, 1915, 162 (5/6) :225-299
[20]   Detection of partial endotracheal tube obstruction by forced pressure oscillations [J].
Schumann, Stefan ;
Lichtwarck-Aschoff, Michael ;
Haberthuer, Christoph ;
Stahl, Claudius A. ;
Moeller, Knut ;
Guttmann, Josef .
RESPIRATORY PHYSIOLOGY & NEUROBIOLOGY, 2007, 155 (03) :227-233