Gallium trace on-line preconcentration/separation and determination using a polyurethane foam mini-column and flame atomic absorption spectrometry. Application in aluminum alloys, natural waters and urine

被引:41
作者
Antheraidis, AN [1 ]
Zachariadis, GA [1 ]
Stratis, JA [1 ]
机构
[1] Aristotle Univ Thessaloniki, Dept Chem, Analyt Chem Lab, Thessaloniki 54124, Greece
关键词
gallium; atomic absorption spectrometry; on-line preconcentration; polyurethane foam;
D O I
10.1016/S0039-9140(03)00156-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sensitive and selective flow injection time-based method for on-line preconcentration/separation and determination of gallium by flame atomic absorption spectrometry at trace levels was developed. The on-line formed gallium chloride complex is sorbed onto a polyether-type polyurethane foam mini-column, followed by on-line quantitative elution with isobutyl methyl ketone and direct introduction into the flame pneumatic nebulizer of the atomic absorption spectrometer. All chemical and flow variables of the system as,well as the possible interferences were studied. The manner of strong HCl solutions propulsion was investigated and established using a combination of two displacement bottles. For 90 s preconcentration time, a sample frequency of 28 h(-1), an enhancement factor of 40, a detection limit of 6 mug 1(-1) and a precision expressed as relative standard deviation (s,) of 3.3% (at 1.00 mg 1(-1)) were achieved. The calibration curve is linear over the concentration range 0.02-3.00 mg 1(-1). The accuracy of the developed method was sufficient and evaluated by the analysis of a silicon-aluminum alloy standard reference material. Finally, it was successfully applied to gallium determination in commercial aluminum alloys, natural waters and urine. (C) 2003 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:929 / 936
页数:8
相关论文
共 27 条
[21]  
MARCZENKO Z, 2000, SEPERATION PRECONCEN
[22]  
ORIANS KJ, 1993, ANAL CHIM ACTA, V282, P73
[23]  
Palagyi S., 1992, PRECONCENTRATION TEC
[24]   Preconcentration of lead complexed with O,O-diethyl-dithiophosphate by column solid-phase extraction using different sorbents in a flow injection system coupled to a flame atomic absorption spectrometer [J].
Quináia, SP ;
da Silva, JBB ;
Rollemberg, MDE ;
Curtius, AJ .
TALANTA, 2001, 54 (04) :687-696
[25]  
STRATIS JA, 1989, ANAL LETT, V22, P1021
[26]   DETERMINATION OF TIN, BISMUTH, ANTIMONY, INDIUM, GALLIUM AND ARSENIC BY SOLVENT-EXTRACTION CUM ATOMIC-ABSORPTION SPECTROPHOTOMETRY [J].
VENKAJI, K ;
NAIDU, PP ;
RAO, TJP .
TALANTA, 1994, 41 (08) :1281-1290
[27]   Preconcentration of trace elements in sea water with poly (acrylaminophosphonic-dithiocarbamate) chelating fiber for their determination by inductively coupled plasma mass spectrometry [J].
Wen, B ;
Shan, XQ ;
Liu, RX ;
Tang, HX .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 363 (03) :251-255