Validation of a Novel 3-Dimensional Sonographic Method for Assessing Gastric Accommodation in Healthy Adults

被引:5
作者
Buisman, Wijnand J. [1 ]
van Herwaarden-Lindeboom, Maud Y. A. [1 ]
Mauritz, Femke A. [1 ]
El Ouamari, Mourad [2 ]
Hausken, Trygve [3 ,4 ]
Olafsdottir, Edda J. [5 ]
van der Zee, David C. [1 ]
Gilja, Odd Helge [3 ,4 ]
机构
[1] Univ Med Ctr Utrecht, Dept Pediat Surg, Room KE 04-140-5,POB 85090, NL-3548 EA Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Radiol, Utrecht, Netherlands
[3] Univ Bergen, Dept Clin Med, Bergen, Norway
[4] Haukeland Hosp, Natl Ctr Ultrasound Gastroenterol, Bergen, Norway
[5] Haukeland Hosp, Dept Pediat, Bergen, Norway
关键词
gastric accommodation; gastric volume; gastrointestinal ultrasound; stomach; 3-dimensional sonography; IN-VITRO EVALUATION; INTRAGASTRIC DISTRIBUTION; FUNCTIONAL DYSPEPSIA; ULTRASONOGRAPHY; BAROSTAT; VOLUME; MEAL; ULTRASOUND; AGREEMENT; MOTILITY;
D O I
10.7863/ultra.15.04086
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objectives-A novel automated 3-dimensional (3D) sonographic method has been developed for measuring gastric volumes. This study aimed to validate and assess the reliability of this novel 3D sonographic method compared to the reference standard in 3D gastric sonography: freehand magneto-based 3D sonography. Methods-A prospective study with 8 balloons (in vitro) and 16 stomachs of healthy volunteers (in vivo) was performed. After a 500-mL liquid meal, 1 preprandial and 3 postprandial volume scans of the stomachs were performed by the novel 3D sonographic method and the current reference-standard 3D sonographic method. Results-The in vitro study showed a mean volume difference between the novel method and the true balloon volume of -1.3 mL; limits of agreement (LoA) were small (-39.3 to 12.3 mL), with an intraclass correlation coefficient (ICC) of 0.998. The in vivo study showed a mean gastric volume of 321 mL between the novel method and the freehand magneto-based method, with a mean volume difference of -4.4 mL; LoA were -40.1 to 31.2 mL, and the ICC was 0.991. The intraobserver and interobserver variability rates were low, at 0.8 mL (LoA, -24.0 to 25.6 mL), with an ICC of 0.995, and 0.5 mL (LoA, of -26.8 to 27.8 mL), with an ICC of 0.999, respectively. Conclusions-The novel 3D sonographic method with automated acquisition showed good agreement with the current reference-standard gastric 3D sonographic method, with low intraobserver and interobserver variability. This novel 3D sonographic method is a valid and reliable technique for determining gastric accommodation.
引用
收藏
页码:1411 / 1418
页数:8
相关论文
共 22 条
  • [1] Defining Left Ventricular Apex-to-Base Twist Mechanics Computed From High-Resolution 3D Echocardiography Validation Against Sonomicrometry
    Ashraf, Muhammad
    Myronenko, Andriy
    Nguyen, Thuan
    Inage, Akio
    Smith, Wayne
    Lowe, Robert I.
    Thiele, Karl
    Kroeker, Carol A. Gibbons
    Tyberg, John V.
    Smallhorn, Jeffrey F.
    Sahn, David J.
    Song, Xubo
    [J]. JACC-CARDIOVASCULAR IMAGING, 2010, 3 (03) : 227 - 234
  • [2] Three- and 4-dimensional ultrasound in obstetrics and gynecology - Proceedings of the American Institute of Ultrasound in Medicine Consensus Conference
    Benacerraf, BR
    Benson, CB
    Abuhamad, AZ
    Copel, JA
    Abramowicz, JS
    DeVore, GR
    Doubilet, PM
    Lee, W
    Lev-Toaff, AS
    Merz, E
    Nelson, TR
    O'Neill, MJ
    Parsons, AK
    Platt, LD
    Pretorius, DH
    Timor-Tritsch, IE
    [J]. JOURNAL OF ULTRASOUND IN MEDICINE, 2005, 24 (12) : 1587 - 1597
  • [3] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] Assessment of gastric accommodation: overview and evaluation of current methods
    De Schepper, HU
    Cremonini, F
    Chitkara, D
    Camilleri, M
    [J]. NEUROGASTROENTEROLOGY AND MOTILITY, 2004, 16 (03) : 275 - 285
  • [6] Gastric motility: Comparison of assessment with real-time MR imaging or barostat measurement - Initial experience
    de Zwart, IM
    Mearadji, B
    Lamb, HJ
    Eilers, PHC
    Masclee, AAM
    de Roos, A
    Kunz, P
    [J]. RADIOLOGY, 2002, 224 (02) : 592 - 597
  • [7] Gastric accommodation and motility are influenced by the barostat device: assessment with magnetic resonance imaging
    de Zwart, Ingrid M.
    Haans, Jeoffrey J. L.
    Verbeek, Paul
    Eilers, Paul H. C.
    de Roos, Albert
    Masclee, Ad A. M.
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-GASTROINTESTINAL AND LIVER PHYSIOLOGY, 2007, 292 (01): : G208 - G214
  • [8] Validation of 3D ultrasound vessel wall volume: An imaging phenotype of carotid atherosclerosis
    Egger, Micaela
    Spence, J. David
    Fenster, Aaron
    Parraga, Grace
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2007, 33 (06) : 905 - 914
  • [9] Gastric accommodation assessed by ultrasonography
    Gilja, Odd Helge
    Lunding, Johan
    Hausken, Trygve
    Gregersen, Hans
    [J]. WORLD JOURNAL OF GASTROENTEROLOGY, 2006, 12 (18) : 2825 - 2829
  • [10] Three-dimensional ultrasonography of the gastric antrum in patients with functional dyspepsia
    Gilja, OH
    Hausken, T
    Odegaard, S
    Berstad, A
    [J]. SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 1996, 31 (09) : 847 - 855