The effect of echo contrast agent on Doppler velocity measurements

被引:15
作者
Yokoyama, N
Schwarz, KQ
Chen, XC
Steinmetz, SD
Becher, H
Schimpky, C
Schlief, R
机构
[1] Univ Rochester, Med Ctr, Cardiol Unit, Rochester, NY 14642 USA
[2] Ctr Biomed Ultrasound, Rochester, NY USA
[3] Univ Bonn, Med Klin, D-5300 Bonn, Germany
[4] Schering AG, Div Clin Res, D-1000 Berlin, Germany
关键词
contrast agents; echocardiography; Doppler ultrasound;
D O I
10.1016/S0301-5629(02)00770-6
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The purpose of this investigation was to determine the effect of echo contrast agents on spectral Doppler velocity measurements. SH U 508A was administered by IV injection in 15 patients. The transmitral flow velocity was measured at the E- and A-wave peaks before the start and at the peak of the contrast effect. The Doppler velocity was determined from the Doppler video spectral display and from power spectral analysis of the audio Doppler signal. The Doppler signal intensity was also measured. The Doppler signal intensity increased 17.4 +/- 3.5 dB (p < 0.0001) following echo contrast injection. This was associated with a significant increase in the spectral peak velocity as determined from either the video display or audio analysis. (p < 0.0001). The velocity corresponding to the audio power peak frequency (the modal velocity) did not change significantly (P = NS) and was independent of Doppler signal strength. (C) 2003 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:765 / 770
页数:6
相关论文
共 17 条
  • [11] Schlief R, 1990, Echocardiography, V7, P61, DOI 10.1111/j.1540-8175.1990.tb00349.x
  • [12] VOLUMETRIC ARTERIAL FLOW QUANTIFICATION USING ECHO CONTRAST - AN IN-VITRO COMPARISON OF 3 ULTRASONIC INTENSITY METHODS - RADIO-FREQUENCY, VIDEO AND DOPPLER
    SCHWARZ, KQ
    BEZANTE, GP
    CHEN, XC
    MOTTLEY, JG
    SCHLIEF, R
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (06) : 447 - 460
  • [13] WHEN CAN DOPPLER BE USED IN-PLACE OF INTEGRATED BACKSCATTER AS A MEASURE OF SCATTERED ULTRASOUND INTENSITY
    SCHWARZ, KQ
    BEZANTE, GP
    CHEN, XC
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1995, 21 (02) : 231 - 242
  • [14] QUANTITATIVE ECHO CONTRAST CONCENTRATION MEASUREMENT BY DOPPLER SONOGRAPHY
    SCHWARZ, KQ
    BEZANTE, GP
    CHEN, XC
    SCHLIEF, R
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (04) : 289 - 297
  • [15] DOPPLER ENHANCEMENT WITH SH-U-508A IN MULTIPLE VASCULAR REGIONS
    SCHWARZ, KQ
    BECHER, H
    SCHIMPFKY, C
    VORWERK, D
    BOGDAHN, U
    SCHLIEF, R
    [J]. RADIOLOGY, 1994, 193 (01) : 195 - 201
  • [16] Enhancement of Mitral Regurgitation and Normal Left Atrial Color Doppler Flow Signals With Peripheral Venous Injection of a Saccharide-Based Contrast Agent
    von Bibra, Helen
    Becher, Harald
    Firschke, Christian
    Schlief, Reinhard
    Emslander, Hans Peter
    Schoemig, Albert
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (02) : 521 - 528
  • [17] CLINICAL-EVALUATION OF LEFT-HEART DOPPLER CONTRAST ENHANCEMENT BY A SACCHARIDE-BASED TRANSPULMONARY CONTRAST AGENT
    VONBIBRA, H
    SUTHERLAND, G
    BECHER, H
    NEUDERT, J
    NIHOYANNOPOULOS, P
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 25 (02) : 500 - 508