The effect of echo contrast agent on Doppler velocity measurements

被引:15
作者
Yokoyama, N
Schwarz, KQ
Chen, XC
Steinmetz, SD
Becher, H
Schimpky, C
Schlief, R
机构
[1] Univ Rochester, Med Ctr, Cardiol Unit, Rochester, NY 14642 USA
[2] Ctr Biomed Ultrasound, Rochester, NY USA
[3] Univ Bonn, Med Klin, D-5300 Bonn, Germany
[4] Schering AG, Div Clin Res, D-1000 Berlin, Germany
关键词
contrast agents; echocardiography; Doppler ultrasound;
D O I
10.1016/S0301-5629(02)00770-6
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The purpose of this investigation was to determine the effect of echo contrast agents on spectral Doppler velocity measurements. SH U 508A was administered by IV injection in 15 patients. The transmitral flow velocity was measured at the E- and A-wave peaks before the start and at the peak of the contrast effect. The Doppler velocity was determined from the Doppler video spectral display and from power spectral analysis of the audio Doppler signal. The Doppler signal intensity was also measured. The Doppler signal intensity increased 17.4 +/- 3.5 dB (p < 0.0001) following echo contrast injection. This was associated with a significant increase in the spectral peak velocity as determined from either the video display or audio analysis. (p < 0.0001). The velocity corresponding to the audio power peak frequency (the modal velocity) did not change significantly (P = NS) and was independent of Doppler signal strength. (C) 2003 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:765 / 770
页数:6
相关论文
共 17 条
  • [1] Doppler frequency shift and time-domain velocity enhancement induced by an ultrasonographic contrast agent
    Basseau, F
    Grenier, N
    Trillaud, H
    Douws, C
    Saint-Amon, A
    [J]. JOURNAL OF ULTRASOUND IN MEDICINE, 2000, 19 (03) : 171 - 176
  • [2] Flow dynamics of stenotic aortic valves assessed by signal processing of Doppler spectrograms
    Bermejo, J
    Antoranz, JC
    García-Fernández, MA
    Moreno, MM
    Delcán, JL
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2000, 85 (05) : 611 - 617
  • [3] FORSBERG F, 1994, J ULTRAS MED, V13, P357
  • [4] Gutberlet M, 1998, J ULTRAS MED, V17, P97
  • [5] Administration of an intravenous perfluorocarbon contrast agent improves echocardiographic determination of left ventricular volumes and ejection fraction: Comparison with cine magnetic resonance imaging
    Hundley, WG
    Kizilbash, AM
    Afridi, I
    Franco, F
    Peshock, RM
    Grayburn, PA
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (05) : 1426 - 1432
  • [6] Comparison of native and contrast-enhanced harmonic echocardiography for visualization of left ventricular endocardial border
    Kasprzak, JD
    Paelinck, B
    Ten Cate, FJ
    Vletter, WB
    de Jong, N
    Poldermans, D
    Elhendy, A
    Bouakaz, A
    Roelandt, JRTC
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1999, 83 (02) : 211 - 217
  • [7] Improvement of pulmonary venous flow Doppler signal after intravenous injection of levovist
    Lambertz, H
    Schuhmacher, U
    Tries, HP
    Stein, T
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 1997, 10 (09) : 891 - 898
  • [8] SHU 508 a (Levovist)-enhanced Doppler echocardiography improves the assessment of valvular heart disease
    Lepper, W
    Franke, A
    von Bibra, H
    Becher, H
    Holz, K
    Näbauer, M
    Vered, Z
    Hoffmann, R
    Flachskampf, FA
    Hanrath, P
    [J]. ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2001, 18 (05): : 363 - 372
  • [9] Effect of US contrast agents on spectral velocities: In vitro evaluation
    Melany, ML
    Grant, EG
    Farooki, S
    McElroy, D
    Kimme-Smith, C
    [J]. RADIOLOGY, 1999, 211 (02) : 427 - 431
  • [10] CLINICAL-APPLICATION OF TRANSPULMONARY CONTRAST-ENHANCED DOPPLER TECHNIQUE IN THE ASSESSMENT OF SEVERITY OF AORTIC-STENOSIS
    NAKATANI, S
    IMANISHI, T
    TERASAWA, A
    BEPPU, S
    NAGATA, S
    MIYATAKE, K
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 20 (04) : 973 - 978