Multipolar radiofrequency ablation using internally cooled electrodes in ex vivo bovine liver: Correlation between volume of coagulation and amount of applied energy

被引:19
作者
Clasen, Stephan [1 ]
Rempp, Hansjoerg [1 ]
Schmidt, Diethard [1 ]
Schraml, Christina [1 ]
Hoffmann, Ruediger [1 ]
Claussen, Claus D. [1 ]
Pereira, Philippe L. [2 ]
机构
[1] Univ Hosp, Dept Diagnost & Intervent Radiol, D-72076 Tubingen, Germany
[2] SLK Clin Heilbronn, Clin Radiol Minimal Invas Therapies & Nucl Med, D-74078 Heilbronn, Germany
关键词
Liver; Radiofrequency ablation; Multipolar; Energy; HEPATIC-TUMORS; BIPOLAR; SIZE;
D O I
10.1016/j.ejrad.2010.10.031
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate the relationship between applied energy and volume of coagulation induced by multipolar radiofrequency (RF) ablation. Methods and materials: Multipolar RF ablations (n = 80) were performed in ex vivo bovine liver. Three bipolar applicators with two electrodes located on each applicator shaft were placed in a triangular array. The power-output (75-225 W) and the distance between the different applicators (2, 3, 4, 5 cm) were systematically varied. The volume of confluent white coagulation and the amount of applied energy were assessed. Based on our experimental data the relationship between the volume of coagulation and applied energy was assessed by nonlinear regression analysis. The variability explained by the model was determined by the parameter r(2). Results: The volume of coagulation increases with higher amounts of applied energy. The maximum amount of energy was applied at a power-output of 75 W and an applicator distance of 5 cm. The corresponding maximum volume of coagulation was 324 cm3 and required an application of 453 kJ. The relationship between amount of applied energy (E) and volume (V) of coagulation can be described by the function, V = 4.39E0.7 (r(2) = 0.88). By approximation the volume of coagulation can be calculated by the linear function V = 0.61E + 40.7 (r(2) = 0.87). Conclusion: Ex vivo the relationship between volume of coagulation and amount of applied energy can be described by mathematical modeling. The amount of applied energy correlates to the volume of coagulation and may be a useful parameter to monitor multipolar RF ablation. (C) 2010 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:111 / 113
页数:3
相关论文
共 14 条
  • [1] In vivo efficiency of four commercial monopolar radiofrequency ablation systems -: A comparative experimental study in pig liver
    Brieger, J
    Pereira, PL
    Trübenbach, J
    Schenk, M
    Kröber, SM
    Schmidt, D
    Aubé, C
    Claussen, CD
    Schick, F
    [J]. INVESTIGATIVE RADIOLOGY, 2003, 38 (10) : 609 - 616
  • [2] Multipolar radiofrequency ablation with internally cooled electrodes:: Experimental study in ex vivo bovine liver with mathernatic modeling
    Clasen, S
    Schmidt, D
    Boss, A
    Dietz, K
    Kröber, SM
    Claussen, CD
    Pereira, PL
    [J]. RADIOLOGY, 2006, 238 (03) : 881 - 890
  • [3] Bipolar Radiofrequency ablation using internally cooled electrodes in ex vivo bovine liver -: Prediction of coagulation volume from applied energy
    Clasen, Stephan
    Schmidt, Diethard
    Dietz, Klaus
    Boss, Andreas
    Kroeber, Stefan M.
    Schraml, Christina
    Fritz, Jan
    Claussen, Claus D.
    Pereira, Philippe L.
    [J]. INVESTIGATIVE RADIOLOGY, 2007, 42 (01) : 29 - 36
  • [4] Radiofrequency ablation of liver tumours: systematic review
    Decadt, B
    Siriwardena, AK
    [J]. LANCET ONCOLOGY, 2004, 5 (09) : 550 - 560
  • [5] Multipolar radiofrequency ablation of hepatic tumors: Initial experience
    Frericks, BB
    Ritz, JP
    Roggan, A
    Wolf, KJ
    Albrecht, T
    [J]. RADIOLOGY, 2005, 237 (03) : 1056 - 1062
  • [6] Image-Guided Tumor Ablation: Standardization of terminology and reporting criteria
    Goldberg, SN
    Grassi, CJ
    Cardella, JF
    Charboneau, JW
    Dodd, GD
    Dupuy, DE
    Gervais, D
    Gillams, AR
    Kane, RA
    Lee, FT
    Livraghi, T
    McGahan, J
    Phillips, DA
    Rhim, H
    Silverman, SG
    [J]. RADIOLOGY, 2005, 235 (03) : 728 - 739
  • [7] Hepatic bipolar radio-frequency ablation between separated multiprong electrodes
    Haemmerich, D
    Staelin, ST
    Tungjitkusolmun, S
    Lee, FT
    Mahvi, DM
    Webster, JG
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (10) : 1145 - 1152
  • [8] Multiple-electrode radiofrequency ablation creates confluent areas of necrosis: In vivo porcine liver results
    Laeseke, Paul F.
    Sampson, Lisa A.
    Haemmerich, Dieter
    Brace, Christopher L.
    Fine, Jason P.
    Frey, Tina M.
    Winter, Thomas C., III
    Lee, Fred T., Jr.
    [J]. RADIOLOGY, 2006, 241 (01) : 116 - 124
  • [9] Lee JM, 2005, EUR J RADIOL, V54, P408, DOI 10.1016/j.ejrad.2004.06.004
  • [10] Effect of vessel size on creation of hepatic radiofrequency lesions in Pigs: Assessment of the "heat sink" effect
    Lu, DSK
    Raman, SS
    Vodopich, DJ
    Wang, M
    Sayre, J
    Lassman, C
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 178 (01) : 47 - 51