Accuracy assessment of the axial images obtained from cone beam computed tomography

被引:12
作者
Panzarella, F. K. [1 ]
Junqueira, J. L. C. [1 ]
Oliveira, L. B. [1 ]
de Araujo, N. S. [1 ]
Costa, C. [1 ]
机构
[1] Fac Sao Leopoldo Mandic, Sch Dent, Campinas, SP, Brazil
关键词
cone beam computed tomography; accuracy; image quality; FLAT-PANEL DETECTOR; LINEAR MEASUREMENTS; DENTAL USE; SPATIAL-RESOLUTION; VOLUMETRIC CT; MULTISLICE; MACHINE; QUALITY; NEWTOM; FIELD;
D O I
10.1259/dmfr/88722046
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: The aim of this study was to evaluate accuracy of linear measurements assessed from axial tomograms and the influence of the use of different protocols in two cone beam CT (CBCT) units. Methods: A cylinder object consisting of NylonH (R) (Day Brazil, Sao Paulo, Brazil) with radio-paque markers was radiographically examined applying different protocols from NewTom 3G (TM) (Quantitative Radiology s.r.l, Verona, Veneto, Italy) and i-CAT (TM) (Imaging Sciences International, Hatfield, PA) units. Horizontal (A-B) and vertical (C-D) distances were assessed from axial tomograms and measured using a digital calliper that provided the gold standard for actual values. Results: There were differences when considering acquisition protocols to each CBCT unit. Concerning all analysed protocols from i-CAT (TM) and Newtom 3G (TM), both A-B and C-D distances presented underestimated values. Measurements of the axial images obtained from NewTom 3G (TM) (6 inch 0.16 mm and 9 inch 0.25 mm) were similar to the ones obtained from i-CAT (TM) (13 cm 20 s 0.3 mm, 13 cm 20 s 0.4 mm and 13 cm 40 s 0.25 mm). Conclusion: The use of different protocols from CBCT machines influences linear measurements assessed from axial images. Linear distances were underestimated in both equipments. Our findings suggest that the best protocol for the i-CAT (TM) is 13 cm 20 s 0.3 mm and for the NewTom 3G (TM), the use of 6 inch or 9 inch is recommended. Dentomaxillofacial Radiology (2011) 40, 369-378. doi: 10.1259/dmfr/88722046
引用
收藏
页码:369 / 378
页数:10
相关论文
共 34 条
[1]  
[Anonymous], 2003, ORTHOD CRANIFOC R S1, DOI DOI 10.1007/S13181-011-0207-X
[2]   Development of a compact computed tomographic apparatus for dental use [J].
Arai, Y ;
Tammisalo, E ;
Iwai, K ;
Hashimoto, K ;
Shinoda, K .
DENTOMAXILLOFACIAL RADIOLOGY, 1999, 28 (04) :245-248
[3]   Characteristics of a newly developed dentomaxillofacial X-ray cone beam CT scanner (CB MercuRay™):: system configuration and physical properties [J].
Araki, K ;
Maki, K ;
Seki, K ;
Sakamaki, K ;
Harata, Y ;
Sakaino, R ;
Okano, T ;
Seo, K .
DENTOMAXILLOFACIAL RADIOLOGY, 2004, 33 (01) :51-59
[4]   Using a flat-panel detector in high resolution cone beam CT for dental imaging [J].
Baba, R ;
Ueda, K ;
Okabe, M .
DENTOMAXILLOFACIAL RADIOLOGY, 2004, 33 (05) :285-290
[5]   Comparison of flat-panel detector and image-intensifier detector for cone-beam CT [J].
Baba, R ;
Konno, Y ;
Ueda, K ;
Ikeda, S .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2002, 26 (03) :153-158
[6]   Image distortion and spatial resolution of a commercially available cone-beam computed tomography machine [J].
Ballrick, John W. ;
Palomo, J. Martin ;
Ruch, Edward ;
Amberman, B. Douglas ;
Hans, Mark G. .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2008, 134 (04) :573-582
[7]   Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and for maxillofacial imaging [J].
Bartling, S. H. ;
Majdani, O. ;
Gupta, R. ;
Rodt, T. ;
Dullin, C. ;
Fitzgerald, P. F. ;
Becker, H. .
DENTOMAXILLOFACIAL RADIOLOGY, 2007, 36 (06) :317-327
[8]  
DAMSTRA J, 2010, BR J ORAL MAXIL 0706
[9]   Development of imaging selection criteria and procedures should precede cephalometric assessment with cone-beam computed tomography [J].
Farman, Allan G. ;
Scarfe, William C. .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2006, 130 (02) :257-265
[10]   Comparison of image validity between cone beam computed tomography for dental use and multidetector row helical computed tomography [J].
Hashimoto, K. ;
Kawashima, S. ;
Kameoka, S. ;
Akiyama, Y. ;
Honjoya, T. ;
Ejima, K. ;
Sawada, K. .
DENTOMAXILLOFACIAL RADIOLOGY, 2007, 36 (08) :465-471