Knotless Anchor Fixation for Transosseous Meniscal Root Repair Using Suture Tape Is Inferior Compared With Button or Screw Fixation: A Biomechanical Study

被引:6
作者
Robinson, James R. [1 ,2 ]
Hernandez, Bruno Agostinho [1 ,2 ]
Taylor, Clare [1 ,2 ]
Gill, Harinderjit S. [1 ,3 ]
机构
[1] Univ Bath, Dept Mech Engn, Healthcare Engn Res Unit, Bath, Avon, England
[2] Avon Orthopaed Ctr, Bristol, Avon, England
[3] Univ Bath, Ctr Therapeut Innovat, Bath, Avon, England
关键词
tibial fixation; knotless anchors; mechanical testing; failure load; meniscal root repair; POSTERIOR ROOT; TENSILE FORCES; CONSEQUENCES; FIBERWIRE; AVULSION; STRENGTH; OUTCOMES; TEAR;
D O I
10.1177/2325967120912185
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: A 2 mm-wide ultrahigh-molecular-weight polyethylene (UHMWPE) tape improves the contact pressure at root repair sites compared with high-strength suture and provides a stronger repair construct. UHMWPE tape is commonly used in rotator cuff repair, and fixation is often achieved with knotless suture anchors. The optimal method for tape fixation for meniscal root repair has not been established. Hypothesis: The use of suture anchors for the tibial fixation of 2-mm UHMWPE tape transosseous root repairs will lead to better biomechanical performance compared with other fixation methods. Methods: The medial meniscal posterior root attachment in 25 porcine knees was divided, and a standardized transtibial root repair was performed using 2-mm UHMWPE tape. The testing was performed by cyclic loading followed by load to failure. Tibial fixation was randomized to 5 tibial fixation types: (1) cortical fixation button, (2) pound-in suture anchor with screw-down interference suture locking, (3) tap-in suture anchor with inner locking plug, (4) postscrew, and (5) postscrew and washer. Results: There was no difference in displacement during cyclic loading between tibial fixation groups except for a highly significant difference in the maximum load at failure. Repairs in both suture anchor fixation groups all failed by tape slippage at relatively low loads (median, 145 and 116 N, respectively). Repairs tied over a cortical button, postscrew, or screw and washer failed by tape breakage at loads of 431, 405, and 528 N. Conclusion: For meniscal root repairs with 2-mm UHMWPE tape, use of suture anchors offers weaker fixation compared with tying over a button or postscrew/washer. While suture anchor fixation may be adequate for nonweightbearing postoperative protocols, it may not allow for more accelerated weightbearing.
引用
收藏
页数:8
相关论文
共 32 条
  • [1] Risk factors for knee instability after anterior cruciate ligament reconstruction
    Ahn, Ji Hyun
    Lee, Sung Hyun
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2016, 24 (09) : 2936 - 2942
  • [2] Biomechanical consequences of a tear of the posterior root of the medial meniscus
    Allaire, Robert
    Muriuki, Muturi
    Gilbertson, Lars
    Harner, Christopher D.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (09) : 1922 - 1931
  • [3] An evaluation of meniscal collagenous structure using optical projection tomography
    Andrews, Stephen H. J.
    Ronsky, Janet L.
    Rattner, Jerome B.
    Shrive, Nigel G.
    Jamniczky, Heather A.
    [J]. BMC MEDICAL IMAGING, 2013, 13
  • [4] Biomechanical Comparison of Arthroscopic Repair Constructs for Meniscal Root Tears
    Anz, Adam W.
    Branch, Eric A.
    Saliman, Justin D.
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2014, 42 (11) : 2699 - 2706
  • [5] Two-Year Outcomes After Arthroscopic Rotator Cuff Repair in Recreational Athletes Older Than 70 Years
    Bhatia, Sanjeev
    Greenspoon, Joshua A.
    Horan, Marilee P.
    Warth, Ryan J.
    Millett, Peter J.
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2015, 43 (07) : 1737 - 1742
  • [6] A Biomechanical Comparison of the Pullout Strength of No. 2 FiberWire Suture and 2-mm FiberWire Tape in Bovine Rotator Cuff Tendons
    Bisson, Leslie J.
    Manohar, Leslie M.
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2010, 26 (11) : 1463 - 1468
  • [7] Biomechanical Validation of Load-Sharing Rip-Stop Fixation for the Repair of Tissue-Deficient Rotator Cuff Tears
    Burkhart, Stephen S.
    Denard, Patrick J.
    Konicek, John
    Hanypsiak, Bryan T.
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2014, 42 (02) : 457 - 462
  • [8] Biomechanical Evaluation of a Transtibial Pull-out Meniscal Root Repair Challenging the Bungee Effect
    Cerminara, Anthony J.
    LaPrade, Christopher M.
    Smith, Sean D.
    Ellman, Michael B.
    Wijdicks, Coen A.
    LaPrade, Robert F.
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2014, 42 (12) : 2988 - 2995
  • [9] The failure mode of two reabsorbable fixation systems: Swivelock with Fibertape versus Bio-Corkscrew with Fiberwire in bovine rotator cuff
    De Carli, Angelo
    Lanzetti, Riccardo Maria
    Monaco, Edoardo
    Labianca, Luca
    Mossa, Luigi
    Feretti, Andrea
    [J]. JOURNAL OF ORTHOPAEDIC SCIENCE, 2012, 17 (06) : 789 - 795
  • [10] Biomechanical evaluation of different suture materials for arthroscopic transtibial pull-out repair of posterior meniscus root tears
    Feucht, Matthias J.
    Grande, Eduardo
    Brunhuber, Johannes
    Rosenstiel, Nikolaus
    Burgkart, Rainer
    Imhoff, Andreas B.
    Braun, Sepp
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2015, 23 (01) : 132 - 139