Efficacy and safety of the dual-layer flow-diverting stent (FRED) for the treatment of intracranial aneurysms

被引:32
作者
Guimaraens, Leopoldo [1 ,2 ]
Vivas, Elio [1 ,2 ]
Saldana, Jesus [1 ,2 ]
Llibre, Juan Carlos [1 ,2 ]
Gil, Alberto [1 ,2 ]
Balaguer, Ernest [1 ]
Rodriguez-Campello, Ana [3 ]
Cuadrado-Godia, Elisa [3 ]
Ois, Angel [3 ]
机构
[1] Hosp Gen Cataluna, Intervent Neuroradiol, Sant Cugat Del Valles 08195, Spain
[2] Hosp del Mar, Intervent Neuroradiol, Barcelona, Spain
[3] Hosp del Mar, Neurol, Barcelona, Spain
关键词
aneurysm; flow Diverter; stent; REDIRECTION ENDOLUMINAL DEVICE; MULTICENTER EXPERIENCE; GRADING SCALE;
D O I
10.1136/neurintsurg-2019-015371
中图分类号
R445 [影像诊断学];
学科分类号
100207 ;
摘要
Purpose To describe the efficacy and complications of treating cerebral aneurysms with the Flow Re-direction Endoluminal Device (FRED) and to identify predictors for aneurysm occlusion. Methods A prospective observational registry including all consecutive aneurysms treated with FRED between December 2015 and July 2018 was designed in one therapeutic neuroangiography department. The primary endpoint for treatment efficacy was complete or near-complete occlusion (O'Kelly-Marotta (OKM) C-D), assessed by three-dimensional digital subtraction angiography. Major (all symptomatics) and minor complications were described and those with modified Rankin Scale scores 3-6 were considered clinically relevant. Univariate and multivariate analyses were performed to identify predictors of efficacy. Results A total of 185 aneurysms were analyzed in 150 patients (mean age 54.3 +/- 11.5 years). Mean follow-up was 18.99 +/- 11.32 months (range 0-43). Efficacy was evaluated in 156 (84.32%) cases: 132 (84.6%) had OKM C-D occlusion, 31/47 (66%) within the first year and 101/109 (92.7%) later on. Major complications were observed in 12 (6.5%) cases: three strokes (one transient ischemic accident, two minor strokes), six intra-stent thrombosis, and three with bleeding, but only one (0.5%) was clinically relevant. Minor complications (all asymptomatic) were observed in 10 (5.4%) cases: three shortening/repositioning of stent; two arterial dissection, two arterial occlusion, and three intra-stent stenosis. Independent predictors of occlusion were immediate OKM grade B-C-D (OR 4.01, 95% CI 1.51 to 10.62), single aneurysm (OR 3.29, 95% CI 1.05 to 10.32), and small size aneurysm (OR 4.74, 95% CI 1.57 to 14.30). Conclusion The FRED stent fully complied with efficacy and safety requirements for treatment of intracranial aneurysms. Three predictors of aneurysm occlusion were identified.
引用
收藏
页码:521 / 525
页数:5
相关论文
共 19 条
[1]   Endothelialization of over- and undersized flow-diverter stents at covered vessel side branches: An in vivo and in silico study [J].
Berg, Philipp ;
Iosif, Christina ;
Ponsonnard, Sebastien ;
Yardin, Catherine ;
Janiga, Gabor ;
Mounayer, Charbel .
JOURNAL OF BIOMECHANICS, 2016, 49 (01) :4-12
[2]   Endovascular treatment of cerebral aneurysms using flow-diverter devices: A systematic review [J].
Briganti, Francesco ;
Leone, Giuseppe ;
Marseglia, Mariano ;
Mariniello, Giuseppe ;
Caranci, Ferdinando ;
Brunetti, Arturo ;
Maiuri, Francesco .
NEURORADIOLOGY JOURNAL, 2015, 28 (04) :365-375
[3]   Spontaneous Delayed Migration/Shortening of the Pipeline Embolization Device: Report of 5 Cases [J].
Chalouhi, N. ;
Tjoumakaris, S. I. ;
Gonzalez, L. F. ;
Hasan, D. ;
Pema, P. J. ;
Gould, G. ;
Rosenwasser, R. H. ;
Jabbour, P. M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (12) :2326-2330
[4]   Hemodynamics of Flow Diverters [J].
Dholakia, Ronak ;
Sadasivan, Chander ;
Fiorella, David J. ;
Woo, Henry H. ;
Lieber, Baruch B. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2017, 139 (02)
[5]   TreatmentTreatment of Intra- and Extracranial Aneurysms Using the Flow-Redirection Endoluminal Device: Multicenter Experience and Follow-Up Results [J].
Drescher, F. ;
Weber, W. ;
Berlis, A. ;
Rohde, S. ;
Carolus, A. ;
Fischer, S. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2017, 38 (01) :105-112
[6]   Observer Variability of an Angiographic Grading Scale Used for the Assessment of Intracranial Aneurysms Treated with Flow-Diverting Stents [J].
Joshi, M. D. ;
O'Kelly, C. J. ;
Krings, T. ;
Fiorella, D. ;
Marotta, T. R. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (08) :1589-1592
[7]   European Multicenter Study for the Evaluation of a Dual-Layer Flow-Diverting Stent for Treatment of Wide-Neck Intracranial Aneurysms: The European Flow-Redirection Intraluminal Device Study [J].
Killer-Oberpfalzer, M. ;
Kocer, N. ;
Griessenauer, C. J. ;
Janssen, H. ;
Engelhorn, T. ;
Holtmannspotter, M. ;
Buhk, J. H. ;
Finkenzeller, T. ;
Fesl, G. ;
Trenkler, J. ;
Reith, W. ;
Berlis, A. ;
Hausegger, K. ;
Augustin, M. ;
Islak, C. ;
Minnich, B. ;
Moehlenbruch, M. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2018, 39 (05) :841-847
[8]   Flow Re-direction Endoluminal Device in treatment of cerebral aneurysms: initial experience with short-term follow-up results [J].
Kocer, Naci ;
Islak, Civan ;
Kizilkilic, Osman ;
Kocak, Burak ;
Saglam, Muzaffer ;
Tureci, Ercan .
JOURNAL OF NEUROSURGERY, 2014, 120 (05) :1158-1171
[9]   A comparison between the new Low-profile Visualized Intraluminal Support (LVIS Blue) stent and the Flow Redirection Endoluminal Device (FRED) in bench-top and cadaver studies [J].
Matsuda, Yoshikazu ;
Chung, Joonho ;
Keigher, Kiffon ;
Lopes, Demetrius .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2018, 10 (03) :274-278
[10]   Multicenter Experience with FRED Jr Flow Re-Direction Endoluminal Device for Intracranial Aneurysms in Small Arteries [J].
Moehlenbruch, M. A. ;
Kizilkilic, O. ;
Killer-Oberpfalzer, M. ;
Baltacioglu, F. ;
Islak, C. ;
Bendszus, M. ;
Cekirge, S. ;
Saatci, I. ;
Kocer, N. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2017, 38 (10) :1959-1965