Metal Artifact Reduction by Dual-Energy Computed Tomography Using Energetic Extrapolation A Systematically Optimized Protocol

被引:137
|
作者
Meinel, Felix G. [1 ]
Bischoff, Bernhard [1 ]
Zhang, Qiaowei [2 ]
Bamberg, Fabian [1 ]
Reiser, Maximilian F. [1 ]
Johnson, Thorsten R. C. [1 ]
机构
[1] Univ Munich, Dept Clin Radiol, D-81377 Munich, Germany
[2] Zhejiang Univ, Sch Med, Sir Run Run Shaw Hosp, Dept Radiol, Hangzhou 310003, Zhejiang, Peoples R China
关键词
metal implants; metal artifacts; dual-energy CT; dual-source CT; energetic extrapolation; CONE-BEAM CT; BONE REMOVAL; RENAL MASSES; IN-VITRO; ANGIOGRAPHY; PROJECTIONS; MULTISLICE; EXPERIENCE; IMPLANTS;
D O I
10.1097/RLI.0b013e31824c86a3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: Energetic extrapolation is a promising strategy to reduce metal artifacts in dual-source computed tomography (DSCT). We performed this study to systematically optimize image acquisition parameters for this approach in a hip phantom and assess its value in a clinical study. Materials and Methods: Titanium and steel hip prostheses were placed in a standard hip phantom and a water tank and scanned on a DSCT scanner. Tube spectra, tube current ratio, collimation, pitch, and rotation time were optimized in a stepwise process. Artifacts were quantified by measuring the standard deviation of the computed tomography density in a doughnut-shaped region of interest placed around the prosthesis. A total of 22 adult individuals with metallic implants referred for computed tomography for a musculoskeletal indication were scanned using the optimized protocol. Degree of artifacts and diagnostic image quality were rated visually (0-10) and maximum streak intensity was measured. Results: Sn140/100 kVp proved superior to Sn140/80 kVp. There was a benefit for increasing tube current ratio from 1:1 to 3:1, but not beyond, in favor of the Sn140 kVp spectrum. Artifacts were less severe for a collimation of 32 x 0.6 mm as compared with 40 x 0.6 mm. A pitch of 0.5 at a rotation time of 0.5 seconds per rotation was preferable to other combinations with comparable scanning times. In the clinical study, increasing the extrapolated photon energy from 64 to 120 keV decreased the severity of artifacts from 8.0 to 2.0 (P < 0.001) and decreased streak intensity from 871 to 153 HU (P < 0.001). The median diagnostic image quality rating improved from 2.5 to 8.0 (P < 0.001). The median energy level visually perceived as optimal for diagnostic evaluation was 113 keV (range, 100-130 keV). Conclusions: Sn140/100 kVp with a tube current ratio of 3:1, a collimation of 32 x 0.6 mm, and extrapolated energies of 105 to 120 keV are optimal parameters for a dedicated DSCT protocol that effectively reduces metal artifacts by energetic extrapolation. The protocol effectively reduces metal artifacts in all types of metal implants. The optimized reconstructions yielded relevant additional findings.
引用
收藏
页码:406 / 414
页数:9
相关论文
共 50 条
  • [1] Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation
    Bamberg, Fabian
    Dierks, Alexander
    Nikolaou, Konstantin
    Reiser, Maximilian F.
    Becker, Christoph R.
    Johnson, Thorsten R. C.
    EUROPEAN RADIOLOGY, 2011, 21 (07) : 1424 - 1429
  • [2] Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation
    Fabian Bamberg
    Alexander Dierks
    Konstantin Nikolaou
    Maximilian F. Reiser
    Christoph R. Becker
    Thorsten R. C. Johnson
    European Radiology, 2011, 21 : 1424 - 1429
  • [3] Applications of Dual-Energy Computed Tomography for Artifact Reduction in the Head, Neck, and Spine
    Liao, Eric
    Srinivasan, Ashok
    NEUROIMAGING CLINICS OF NORTH AMERICA, 2017, 27 (03) : 489 - +
  • [4] Metal artifact reduction by monoenergetic extrapolation of dual-energy CT in patients with metallic implants
    Lee, Ka Yin Gregory
    Cheng, Hei Man Joyce
    Chu, Chi Yeung
    Tam, Chi Wai Annie
    Kan, Wai Kuen
    JOURNAL OF ORTHOPAEDIC SURGERY, 2019, 27 (02)
  • [5] Single-Source Dual-Energy Computed Tomography Use of Monoenergetic Extrapolation for a Reduction of Metal Artifacts
    Mangold, Stefanie
    Gatidis, Sergios
    Luz, Oliver
    Koenig, Benjamin
    Schabel, Christoph
    Bongers, Malte N.
    Flohr, Thomas G.
    Claussen, Claus D.
    Thomas, Christoph
    INVESTIGATIVE RADIOLOGY, 2014, 49 (12) : 788 - 793
  • [6] Dual-Energy Computed Tomography Dose Reduction, Series Reduction, and Contrast Load Reduction in Dual-Energy Computed Tomography
    Parakh, Anushri
    Macri, Francesco
    Sahani, Dushyant
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2018, 56 (04) : 601 - +
  • [7] How Well Does Dual-Energy Computed Tomography With Metal Artifact Reduction Software Improve Image Quality and Quantify Computed Tomography Number and Iodine Concentration?
    Ohira, Shingo
    Kanayama, Naoyuki
    Wada, Kentaro
    Karino, Tsukasa
    Nitta, Yuya
    Ueda, Yoshihiro
    Miyazaki, Masayoshi
    Koizumi, Masahiko
    Teshima, Teruki
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (04) : 655 - 660
  • [8] Quantifying near metal visibility using dual energy computed tomography and iterative metal artifact reduction in a fracture phantom
    Hakvoort, E. T.
    Wellenberg, R. H. H.
    Streekstra, G. J.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 69 : 9 - 18
  • [9] Metal Artifact Reduction Computed Tomography of Arthroplasty Implants Effects of Combined Modeled Iterative Reconstruction and Dual-Energy Virtual Monoenergetic Extrapolation at Higher Photon Energies
    Khodarahmi, Iman
    Haroun, Reham R.
    Lee, Moses
    Fung, George S. K.
    Fuld, Matthew K.
    Schon, Lew C.
    Fishman, Elliot K.
    Fritz, Jan
    INVESTIGATIVE RADIOLOGY, 2018, 53 (12) : 728 - 735
  • [10] Metal artifact reduction by dual-layer computed tomography using virtual monoenergetic images
    Neuhaus, Victor
    Hokamp, Nils Grosse
    Abdullayev, Nuran
    Rau, Robert
    Mpotsaris, Anastasios
    Maintz, David
    Borggrefe, Jan
    EUROPEAN JOURNAL OF RADIOLOGY, 2017, 93 : 143 - 148