Bioreactance Is Not Interchangeable with Thermodilution for Measuring Cardiac Output during Adult Liver Transplantation

被引:18
作者
Han, Sangbin [1 ]
Lee, Jong Hwan [1 ]
Kim, Gaabsoo [1 ]
Ko, Justin Sangwook [1 ]
Choi, Soo Joo [1 ]
Kwon, Ji Hae [1 ]
Heo, Burn Young [1 ]
Gwak, Mi Sook [1 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Anesthesiol & Pain Med, Seoul, South Korea
关键词
PULMONARY-ARTERY CATHETER; SURGERY; MONITORS; SYSTEM; FLUID;
D O I
10.1371/journal.pone.0127981
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background Thermodilution technique using a pulmonary artery catheter is widely used for the assessment of cardiac output (CO) in patients undergoing liver transplantation. However, the unclearness of the risk-benefit ratio of this method has led to an interest in less invasive modalities. Thus, we evaluated whether noninvasive bioreactance CO monitoring is interchangeable with thermodilution technique. Methods Nineteen recipients undergoing adult-to-adult living donor liver transplantation were enrolled in this prospective observational study. COs were recorded automatically by the two devices and compared simultaneously at 3-minute intervals. The Bland-Altman plot was used to evaluate the agreement between bioreactance and thermodilution. Clinically acceptable agreement was defined as a percentage error of limits of agreement <30%. The four quadrant plot was used to evaluate concordance between bioreactance and thermodilution. Clinically acceptable concordance was defined as a concordance rate >92%. Results A total of 2640 datasets were collected. The mean CO difference between the two techniques was 0.9 l/min, and the 95% limits of agreement were -3.5 l/min and 5.4 l/min with a percentage error of 53.9%. The percentage errors in the dissection, anhepatic, and reperfusion phase were 50.6%, 56.1%, and 53.5%, respectively. The concordance rate between the two techniques was 54.8%. Conclusion Bioreactance and thermodilution failed to show acceptable interchangeability in terms of both estimating CO and tracking CO changes in patients undergoing liver transplantation. Thus, the use of bioreactance as an alternative CO monitoring to thermodilution, in spite of its noninvasiveness, would be hard to recommend in these surgical patients.
引用
收藏
页数:11
相关论文
共 29 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Continuous versus intermittent thermodilution cardiac output measurement during orthotopic liver transplantation [J].
Bottiger, BW ;
Sinner, B ;
Motsch, J ;
Bach, A ;
Bauer, H ;
Martin, E .
ANAESTHESIA, 1997, 52 (03) :207-214
[3]   Correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off-pump coronary artery bypass surgery [J].
Cheung, Hoiyin ;
Dong, Quan ;
Dong, Rong ;
Yu, Buwei .
JOURNAL OF ANESTHESIA, 2015, 29 (03) :416-420
[4]   A comparison of noninvasive bioreactance with oesophageal Doppler estimation of stroke volume during open abdominal surgery An observational study [J].
Conway, Daniel H. ;
Hussain, Osman A. ;
Gall, Iain .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2013, 30 (08) :501-508
[5]   The effect of peripheral resistance on impedance cardiography measurements in the anesthetized dog [J].
Critchley, LAH ;
Peng, ZY ;
Fok, BS ;
James, AE .
ANESTHESIA AND ANALGESIA, 2005, 100 (06) :1708-1712
[6]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[7]   A Critical Review of the Ability of Continuous Cardiac Output Monitors to Measure Trends in Cardiac Output [J].
Critchley, Lester A. ;
Lee, Anna ;
Ho, Anthony M. -H. .
ANESTHESIA AND ANALGESIA, 2010, 111 (05) :1180-1192
[8]   6/2/06 Perioperative Assessment of the Cardiovascular System in ESLD and Transplantation [J].
De Wolf, Andre M. .
INTERNATIONAL ANESTHESIOLOGY CLINICS, 2006, 44 (04) :59-78
[9]   Pulmonary artery catheter: Rest in peace? Not just quite yet... [J].
De Wolf, Andre M. .
LIVER TRANSPLANTATION, 2008, 14 (07) :917-918
[10]  
Feltracco P, 2012, MINERVA ANESTESIOL, V78, P1372