The VNIIFTRI magnetic temperature scale is compared with the provisional low-temperature scale (PLTS-2000), based on the melting pressure of He-3, in the 0.37-0.56 K range. The magnetic temperature scale is reproduced from the readings of a rhodium-iron alloy resistance thermometer. The disagreement between the scales does not exceed 0.23 mK.
引用
收藏
页码:180 / 185
页数:6
相关论文
共 6 条
[1]   The VNIIFTRI magnetic temperature scale in the 0.3-3 K range [J].
Astrov, D. N. ;
Ermakov, N. B. ;
Sviridenko, V. I. .
MEASUREMENT TECHNIQUES, 2007, 50 (08) :870-879
[2]   Thermometer working standard for temperatures below 0.5 K [J].
Astrov, DN ;
Korostin, SV ;
Polunin, SP .
MEASUREMENT TECHNIQUES USSR, 1995, 38 (12) :1356-1359
[3]   Study of standard thermometers and their reproduction of the λ point in liquid helium [J].
Astrov, DN ;
Korostin, SV .
MEASUREMENT TECHNIQUES, 1999, 42 (07) :689-692
[4]   Purification of 3He from a 4He impurity using adsorption [J].
Dmitriev, VV ;
Zavjalov, VV ;
Zmeev, DY ;
Levitin, LV .
INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2004, 47 (04) :567-569
[5]  
RUSBY RL, 2007, LOW TEMP PHYS, P156
[6]   HIGHLY SENSITIVE CAPACITIVE PRESSURE GAUGE [J].
STRATY, GC ;
ADAMS, ED .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1969, 40 (11) :1393-&