Image quality optimization using a narrow vertical detector dental cone-beam CT

被引:8
作者
Brasil, Danieli Moura [1 ]
Pauwels, Ruben [2 ,3 ,4 ,5 ]
Coucke, Wim
Haiter-Neto, Francisco [1 ]
Jacobs, Reinhilde [2 ,3 ,6 ]
机构
[1] Univ Campinas UNICAMP, Dept Oral Diag, Div Oral Radiol, Piracicaba Dent Sch, Piracicaba, Brazil
[2] Katholieke Univ Leuven, Dept Imaging & Pathol, Fac Med, Leuven, Belgium
[3] Univ Hosp Leuven, Dept Oral & Maxillofacial Surg, Leuven, Belgium
[4] Katholieke Univ Leuven, Dept Mech Engn, Leuven, Belgium
[5] Chulalongkorn Univ, Fac Dent, Dept Radiol, Bangkok, Thailand
[6] Karolinska Inst, Dept Dent Med, Stockholm, Sweden
关键词
image quality; cone-beam CT; phantoms; imaging; optimization; computed-assisted image analysis; COMPUTED-TOMOGRAPHY; REDUCTION; SETTINGS; SIZE;
D O I
10.1259/dmfr.20180357
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To determine the optimized kV setting for a narrow detector cone-beam CT (CBCT) unit. Methods: Clinical (CL) and quantitative (QUANT) evaluations of image quality were performed using an anthropomorphic phantom. Technical (TECH) evaluation was performed with a polymethyl methacrylate phantom. Images were obtained using a PaX-i3D Green CBCT (Vatech, Hwaseong, Korea) device, with a large 21 x 19 and a medium 12 x 9cm field of view (FOY), and high-dose (HD-ranging from 85 to 110 kV) and low-dose (LD-ranging from 75 to 95 kV) protocols, totaling four groups (21 x 19cm HD, 21 x 19cm LD, 12 x 9cm HD, 12 x 9cm LD). The radiation dose within each group was fixed by adapting the mA according to a predetermined dose-area product. For CL evaluation, three observers assessed images based on overall quality, sharpness, contrast, artefacts, and noise. For QUANT evaluation, mean gray value shift, % increase of standard deviation (SD), % of beam hardening and contrast-to-noise ratio (CNR) were calculated. For TECH evaluation, segmentation accuracy, CNR, metal artefact SD, metal object area, and sharpness were measured. Representative parameters were chosen for CL, QUANT, and TECH evaluations to determine the optimal kV based on biplot graphs. kV values of the same protocol were compared by the bootstrapping approach. The ones that had statistical differences with the best kV were considered as worse quality. Results: Overall, kV values within the same group showed similar quality (p > 0.05), except for 110 kV in 21 x 19cm HD and 85 kV in 12 x 9cm HD of CL score; also 85, 90 kV in 21 x 19cm HD and 75, 80 kV in 21 x 19cm LD of QUANT score which were worse (p < 0.05). Conclusion: At a constant dose, low and high kV protocols yield acceptable image quality for a narrow-detector CBCT unit.
引用
收藏
页数:10
相关论文
共 28 条
[1]   Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom [J].
Bamba, J. ;
Araki, K. ;
Endo, A. ;
Okano, T. .
DENTOMAXILLOFACIAL RADIOLOGY, 2013, 42 (08)
[2]   Relationship between physical factors and subjective image quality of cone-beam computed tomography images according to diagnostic task [J].
Choi, Jin-Woo ;
Lee, Sam-Sun ;
Choi, Soon-Chul ;
Heo, Min-Suk ;
Huh, Kyung-Heo ;
Yi, Won-Jin ;
Kang, Se-Ryong ;
Han, Dong-Hun ;
Kim, Eun-Kyung .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2015, 119 (03) :357-365
[3]   Objective and subjective image evaluation of maxillary alveolar bone based on cone beam computed tomography exposure parameters [J].
de Moura, Pollyana Marques ;
Hallac, Rami R. ;
Seaward, James R. ;
Kane, Alex A. ;
Aguiar, Marcelo ;
Raggio, Ronir ;
Gutfilen, Bianca .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2016, 121 (05) :557-565
[4]   Image quality assessment of clinically-applied CBCT protocols using a QAT phantom [J].
Elkhateeb, Sara M. ;
Torgersen, Gerald R. ;
Arnout, Eman A. .
DENTOMAXILLOFACIAL RADIOLOGY, 2016, 45 (05)
[5]   As Low Dose as Sufficient Quality: Optimization of Cone-beam Computed Tomographic Scanning Protocol for Tooth Autotransplantation Planning and Follow-up in Children [J].
EzEldeen, Mostafa ;
Stratis, Andreas ;
Coucke, Wim ;
Codari, Marina ;
Politis, Constantinus ;
Jacobs, Reinhilde .
JOURNAL OF ENDODONTICS, 2017, 43 (02) :210-217
[6]  
Farman A.G., 2009, Semi-nars in Orthodontics, V15, P2, DOI DOI 10.1053/J.S0D0.2008.09.001
[7]   Principal component analysis: a review and recent developments [J].
Jolliffe, Ian T. ;
Cadima, Jorge .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2016, 374 (2065)
[8]   elastix: A Toolbox for Intensity-Based Medical Image Registration [J].
Klein, Stefan ;
Staring, Marius ;
Murphy, Keelin ;
Viergever, Max A. ;
Pluim, Josien P. W. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (01) :196-205
[9]   Image quality produced by different cone-beam computed tomography settings [J].
Kwong, Jeffrey C. ;
Palomo, J. Martin ;
Landers, Michael A. ;
Figueroa, Alex ;
Hans, Mark G. .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2008, 133 (02) :317-327
[10]   MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174