Assessing noise sources at synchrotron infrared ports

被引:6
作者
Lerch, Ph [1 ]
Dumas, P. [2 ]
Schilcher, T. [1 ]
Nadji, A. [2 ]
Luedeke, A. [1 ]
Hubert, N. [2 ]
Cassinari, L. [2 ]
Boege, M. [1 ]
Denard, J. -C. [2 ]
Stingelin, L. [1 ]
Nadolski, L. [2 ]
Garvey, T. [1 ]
Albert, S. [3 ]
Gough, Ch [1 ]
Quack, M. [3 ]
Wambach, J. [1 ]
Dehler, M. [1 ]
Filhol, J. -M. [2 ]
机构
[1] Paul Scherrer Inst, Swiss Light Source, CH-5232 Villigen, Switzerland
[2] Synchrotron SOLEIL, F-91192 Gif Sur Yvette, France
[3] Swiss Fed Inst Technol, Phys Chem Lab, CH-8093 Zurich, Switzerland
关键词
noise; FTIR; spectroscopy; FACILITY SOLEIL; LIGHT-SOURCE; PERFORMANCE; RADIATION; BEAMLINE; SPECTROSCOPY;
D O I
10.1107/S0909049511041884
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Today, the vast majority of electron storage rings delivering synchrotron radiation for general user operation offer a dedicated infrared port. There is growing interest expressed by various scientific communities to exploit the mid-IR emission in microspectroscopy, as well as the far infrared (also called THz) range for spectroscopy. Compared with a thermal (laboratory-based source), IR synchrotron radiation sources offer enhanced brilliance of about two to three orders of magnitude in the mid-IR energy range, and enhanced flux and brilliance in the far-IR energy range. Synchrotron radiation also has a unique combination of a broad wavelength band together with a well defined time structure. Thermal sources (globar, mercury filament) have excellent stability. Because the sampling rate of a typical IR Fourier-transform spectroscopy experiment is in the kHz range (depending on the bandwidth of the detector), instabilities of various origins present in synchrotron radiation sources play a crucial role. Noise recordings at two different IR ports located at the Swiss Light Source and SOLEIL (France), under conditions relevant to real experiments, are discussed. The lowest electron beam fluctuations detectable in IR spectra have been quantified and are shown to be much smaller than what is routinely recorded by beam-position monitors.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 18 条
[1]   Synchrotron-based highest resolution Fourier transform infrared spectroscopy of naphthalene (C10H8) and indole (C8H7N) and its application to astrophysical problems [J].
Albert, Sieghard ;
Albert, Karen Keppler ;
Lerch, Philippe ;
Quack, Martin .
FARADAY DISCUSSIONS, 2011, 150 :71-99
[2]  
[Anonymous], 2011, HDB HIGH RESOLUTION
[3]   Reducing the sensitivity of Fourier transform infrared spectroscopy to line-frequency source variations [J].
Bosch, RA ;
Julian, RL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (03) :1420-1422
[4]  
Byrd J. M., 1999, Proceedings of the 1999 Particle Accelerator Conference (Cat. No.99CH36366), P1806, DOI 10.1109/PAC.1999.794266
[5]   MIDINFRARED BEAMLINE AT THE NATIONAL SYNCHROTRON LIGHT-SOURCE PORT U2B [J].
CARR, GL ;
HANFLAND, M ;
WILLIAMS, GP .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02) :1643-1645
[6]   PERFORMANCE OF AN INFRARED MICROSPECTROMETER AT THE NSLS [J].
CARR, GL ;
REFFNER, JA ;
WILLIAMS, GP .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (02) :1490-1492
[7]   Resolution limits for infrared microspectroscopy explored with synchrotron radiation [J].
Carr, GL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2001, 72 (03) :1613-1619
[8]   Ultra-broadband infrared pump-probe spectroscopy using synchrotron radiation and a tuneable pump [J].
Carroll, Lee ;
Friedli, Peter ;
Lerch, Philippe ;
Schneider, Joerg ;
Treyer, Daniel ;
Hunziker, Stephan ;
Stutz, Stefan ;
Sigg, Hans .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (06)
[9]  
Dehler M, 1999, P DIPAC, P168
[10]   Synchrotron infrared microscopy at the French Synchrotron Facility SOLEIL [J].
Dumas, P. ;
Polack, F. ;
Lagarde, B. ;
Chubar, O. ;
Giorgetta, J. L. ;
Lefrancois, S. .
INFRARED PHYSICS & TECHNOLOGY, 2006, 49 (1-2) :152-160