Charge collection efficiency characterization of a HgI2 Frisch collar spectrometer with collimated high energy gamma rays

被引:13
作者
Kargar, A. [1 ]
Ariesanti, E. [2 ]
James, S. [2 ]
McGregor, D. S. [2 ]
机构
[1] Radiat Monitoring Devices Inc, Watertown, MA 02472 USA
[2] Kansas State Univ, Mech & Nucl Engn Dept, SMART Lab, Manhattan, KS 66506 USA
关键词
HgI2; Frisch collar detector; Gamma-ray;
D O I
10.1016/j.nima.2010.08.057
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In this study, a 2.1 x 2.1 x 4.1 mm(3) HgI2 Frisch collar device was characterized through probing the device with a highly collimated 662 keV gamma rays (Cs-137 check source) along the length and width of the device. In a systematic series of experiments, the detector was probed along its central line under different operating voltages of 1600, 1300, 1000, 800, 600 and 500 V. The experimental results were confirmed through a simulation of the charge collection for a device with the same size and operating conditions. It is shown that the HgI2 Frisch collar device has a uniform response to gamma rays over two-thirds of the detector volume. The HgI2 crystals and the Frisch collar detectors were grown and fabricated within the S.M.A.R.T Laboratory at Kansas State University. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:186 / 192
页数:7
相关论文
共 25 条
[1]  
Ariesanti E., NUCL INSTRUM A UNPUB
[2]  
Ariesanti E., MERCURIC IODID UNPUB
[3]   CHARGE-TRANSPORT IN ARRAYS OF SEMICONDUCTOR GAMMA-RAY DETECTORS [J].
BARRETT, HH ;
ESKIN, JD ;
BARBER, HB .
PHYSICAL REVIEW LETTERS, 1995, 75 (01) :156-159
[4]   Large area/volume CZT nuclear detectors [J].
Bolotnikov, AE ;
Camarda, GS ;
Carini, GA ;
Wright, GW ;
Li, L ;
Burger, A ;
Groza, M ;
James, RB .
Physica Status Solidi C - Conferences and Critical Reviews, Vol 2, No 5, 2005, 2 (05) :1495-1503
[5]   Charge collection characteristics of Frisch collar CdZnTe gamma-ray spectrometers [J].
Harrison, Mark J. ;
Kargar, Alireza ;
McGregor, Douglas S. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 579 (01) :134-137
[6]   CdZnTe Frisch collar detectors for γ-ray spectroscopy [J].
Kargar, A ;
Jones, AM ;
McNeil, WJ ;
Harrison, MJ ;
McGregor, DS .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 558 (02) :497-503
[7]   Effect of Crystal Length on CdZnTe Frisch Collar Device Performance [J].
Kargar, A. ;
Brooks, A. C. ;
Harrison, M. J. ;
Chen, H. ;
Awadalla, S. ;
Bindley, G. ;
McGregor, D. S. .
2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, :2017-+
[8]  
Kargar A., 2009, SPIE P, V7449, P744908
[9]  
Kargar A., COMP SPECTROSC UNPUB
[10]  
Kargar A., CHARGE CO A IN PRESS