Reliability and validity of the Biodex system 3 pro isokinetic dynamometer velocity, torque and position measurements

被引:553
作者
Drouin, JM
Valovich-mcLeod, TC
Shultz, SJ
Gansneder, BM
Perrin, DH
机构
[1] Georgia So Univ, Dept Publ Hlth, Statesboro, GA 30460 USA
[2] Arizona Sch Hlth Sci, Dept Sports Hlth Care, Mesa, AZ 85206 USA
[3] Univ N Carolina, Dept Exercise & Sports Sci, Greensboro, NC 27402 USA
[4] Univ Virginia, Dept Educ Leadership & Fdn, Charlottesville, VA 22904 USA
[5] Univ N Carolina, Sch Hlth & Human Performance, Greensboro, NC 27402 USA
关键词
muscle function; muscle testing; reliability; validity;
D O I
10.1007/s00421-003-0933-0
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
This study quantitatively assessed the mechanical reliability and validity of position, torque and velocity measurements of the Biodex System 3 isokinetic dynamometer. Trial-to-trial and day-to-day reliability were assessed during three trials on two separate days. To assess instrument validity, measurement of each variable using the Biodex System 3 dynamometer was compared to a criterion measure of position, torque and velocity. Position was assessed at 5degrees increments across the available range of motion of the dynamometer. Torque measures were assessed isometrically by hanging six different calibrated weights from the lever arm. Velocity was assessed (30degrees/s to 500degrees/s) across a 70degrees arc of motion by manually accelerating the weighted lever arm. With the exception of a systematic decrease in velocity at speeds of 300degrees/s and higher, the Biodex System 3 performed with acceptable mechanical reliability and validity on all variables tested.
引用
收藏
页码:22 / 29
页数:8
相关论文
共 15 条
  • [1] [Anonymous], FDN CLIN RES APPL PR
  • [2] [Anonymous], 1992, Isokinetics Exercise Science
  • [3] BEMBEN M G, 1988, Journal of Orthopaedic and Sports Physical Therapy, V10, P12
  • [4] MANAGEMENT OF INFECTIONS IN ADULT INTENSIVE-CARE UNIT PATIENTS .1.
    BROWN, RB
    TERES, D
    [J]. INFECTIOUS DISEASES IN CLINICAL PRACTICE, 1993, 2 (03) : 163 - 168
  • [5] Denegar CR, 1993, J SPORT REHABIL, V2, P35, DOI DOI 10.1123/JSR.2.1.35
  • [6] FARRELL M, 1986, MED SCI SPORT EXER, V18, P44
  • [7] FEIRING D C, 1990, Journal of Orthopaedic and Sports Physical Therapy, V11, P298
  • [8] The utility of isokinetic dynamometry in the assessment of human muscle function
    Gleeson, NP
    Mercer, TH
    [J]. SPORTS MEDICINE, 1996, 21 (01) : 18 - 34
  • [9] Prerequisites and limitations to isokinetic measurements in humans - Investigations on a servomotor-controlled dynamometer
    Handel, M
    Dickhuth, HH
    Mayer, F
    Gulch, RW
    [J]. EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY, 1996, 73 (3-4): : 225 - 230
  • [10] Iossifidou AN, 2000, J ORTHOP SPORT PHYS, V30, P317, DOI 10.2519/jospt.2000.30.6.317