Respiratory artefact causing malpositioning of liver dome lesion in right lower lung

被引:48
作者
Sarikaya, I [1 ]
Yeung, HWD [1 ]
Erdi, Y [1 ]
Larson, SM [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Nucl Med Sect, New York, NY 10021 USA
关键词
PET; CT; respiratory artefact;
D O I
10.1097/01.rlu.0000093095.28642.2b
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The new combined positron emission (PET)/computed tomographic (CT) scanners have many advantages over PET scanners alone. However, physicians must be aware of the potential artefacts observed in PET/CT scanners. A body PET/CT was performed on an 81-year-old man with colorectal cancer. The CT-based, attenuation-corrected PET image showed a right lower lung lesion. However, there was no lung lesion on the transmission CT image. Nonattenuation-corrected PET, and rod source-based, attenuation-corrected PET images demonstrated focal uptake in the dome of the liver. Dedicated CT with intravenous contrast confirmed that the lesion was in the liver dome and not in the right lower lung. The liver lesion was misplaced to the right lower lung in the CT-based, attenuation-corrected PET image because of a respiratory artefact. To overcome this respiration artefact the authors suggest a routine review of the nonattenuation-corrected PET images, particularly when evaluating liver dome and lower lung lesions.
引用
收藏
页码:943 / 944
页数:2
相关论文
共 8 条
  • [1] Bengel FM, 1997, EUR J NUCL MED, V24, P1091
  • [2] Dual-modality PET/CT imaging: the effect of respiratory motion on combined image quality in clinical oncology
    Beyer, T
    Antoch, G
    Blodgett, T
    Freudenberg, LF
    Akhurst, T
    Mueller, S
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (04) : 588 - 596
  • [3] PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients
    Burger, C
    Goerres, G
    Schoenes, S
    Buck, A
    Lonn, AHR
    von Schulthess, GK
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (07) : 922 - 927
  • [4] Positron emission tomography imaging in oncology
    Delbeke, D
    Martin, WH
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2001, 39 (05) : 883 - +
  • [5] PET-CT image co-registration in the thorax: influence of respiration
    Goerres, GW
    Kamel, E
    Heidelberg, TNH
    Schwitter, MR
    Burger, C
    von Schulthess, GK
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (03) : 351 - 360
  • [6] Nakamoto Y, 2002, J NUCL MED, V43, P1137
  • [7] Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    Osman, MM
    Cohade, C
    Nakamoto, Y
    Wahl, RL
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (04) : 603 - 606
  • [8] Osman MM, 2003, J NUCL MED, V44, P240