A SiPM based real time dosimeter for radiotherapic beams

被引:8
|
作者
Berra, A. [1 ,2 ]
Conti, V. [3 ]
Lietti, D. [1 ,2 ]
Milan, L. [1 ,2 ]
Novati, C. [1 ,2 ]
Ostinelli, A. [3 ]
Prest, M. [1 ,2 ]
Romano, C. [1 ,2 ]
Vallazza, E. [4 ]
机构
[1] Univ Insubria, I-22100 Como, Italy
[2] Ist Nazl Fis Nucl, Sez Milano Bicocca, I-22100 Como, Italy
[3] Osped St Anna, Serv Fis Sanitaria, Como, Italy
[4] Ist Nazl Fis Nucl, Sez Trieste, Trieste, Italy
关键词
Radiation dosimetry; Silicon PhotoMultipliers; Scintillator; SCINTILLATING-FIBER DOSIMETER; RADIATION-THERAPY; PHOTON BEAMS; IMRT; VERIFICATION; DELIVERY; GAMMA; WATER;
D O I
10.1016/j.nima.2014.11.001
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This paper describes the development of a scintillator dosimeter prototype for radiotherapic applications based on plastic scintillating fibers readout by Silicon PhotoMultipliers. The dosimeter, whose probes are water equivalent, could be used for quality control measurements, beam characterization and in vivo dosimetry, allowing a real time measurement of the dose spatial distribution. This paper describes the preliminary percentual depth dose scan performed with clinical 6 and 18 MV photon beams, comparing the results with a reference curve. The measurements were performed using a Varian Clinac iX linear accelerator at the Radiotherapy Department of the St Anna Hospital in Como (IT). The prototype has given promising results, allowing real time measurements of relative dose without applying any correction factors. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:72 / 80
页数:9
相关论文
共 50 条
  • [1] Single and multichannel scintillating fiber dosimeter for radiotherapic beams with SiPM readout
    Berra, A.
    Ferri, A.
    Novati, C.
    Ostinelli, A.
    Paternoster, G.
    Piemonte, C.
    Prest, M.
    Vallazza, E.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 838 : 1 - 11
  • [2] Real-time dosimeter based on LiF:Mg,Cu,P and SiPM
    Tang, Kaiyong
    Zhang, Siyuan
    RADIATION MEASUREMENTS, 2021, 145
  • [3] A new real-time personal dosimeter with position monitoring based on a scintillator
    Moon, Sun Young
    Yoon, Myonggeun
    Kim, Sun-Young
    Shin, Dongho
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2021, 78 (11) : 1133 - 1140
  • [4] A new real-time personal dosimeter with position monitoring based on a scintillator
    Sun Young Moon
    Myonggeun Yoon
    Sun-Young Kim
    Dongho Shin
    Journal of the Korean Physical Society, 2021, 78 : 1133 - 1140
  • [5] Development of implantable real time micro dosimeter system
    Yamada, Shogo
    Nakata, Eiko
    Momose, Genki
    Kikuchi, Youhei
    Ogawa, Yoshihiro
    Mori, Issei
    Oishi, Mikio
    Ishii, Keizo
    FUTURE MEDICAL ENGINEERING BASED ON BIONANOTECHNOLOGY, PROCEEDINGS, 2006, : 271 - +
  • [6] Real-Time Fibered Optically Stimulated Luminescence Dosimeter Based on SrS:Ce,Sm Phosphor
    Benoit, D.
    Garcia, P.
    Matias-Vaille, S.
    Ravotti, F.
    Vaille, J-R.
    Glaser, M.
    Brichard, B.
    Fernandez, A. F.
    Chatry, C.
    Dusseau, L.
    RADECS 2007: PROCEEDINGS OF THE 9TH EUROPEAN CONFERENCE ON RADIATION AND ITS EFFECTS ON COMPONENTS AND SYSTEMS, 2007, : 356 - +
  • [7] Real-Time Fibered Optically Stimulated Luminescence Dosimeter Based on SrS:Ce,Sm Phosphor
    Benoit, D.
    Garcia, P.
    Vaille, S. Matias
    Ravotti, F.
    Vaille, J. -R
    Glaser, M.
    Brichard, B.
    Fernandez, A. F.
    Chatry, C.
    Dusseau, L.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (04) : 2154 - 2160
  • [8] A Filterless Fluorescence Detector Based on a Time-Gated SiPM
    Buonanno, Luca
    Putelli, Chiara
    Di Vita, Davide
    Fiorini, Carlo
    Carminati, Marco
    2021 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (IEEE BIOCAS 2021), 2021,
  • [9] Time Resolution of Scintillation Detectors Based on SiPM in Comparison to Photomultipliers
    Szczesniak, T.
    Moszynski, M.
    Grodzicka, M.
    Wolski, D.
    Swiderski, L.
    Szawlowski, M.
    Kapusta, M.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 1728 - 1735
  • [10] Characterization of Yb-doped silica optical fiber as real-time dosimeter
    Veronese, Ivan
    Chiodini, Norberto
    Cialdi, Simone
    D'Ippolito, Eduardo
    Fasoli, Mauro
    Gallo, Salvatore
    Mones, Eleonora
    Vedda, Anna
    Loi, Gianfranco
    HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS XIX, 2017, 10392