Closed Sinusoid Trajectory for C-Arm CT Imaging

被引:1
|
作者
Yang, Haiquan [1 ]
Li, Meihua [1 ]
Koizumi, Kazuhito [1 ]
Kudo, Hiroyuki [2 ]
机构
[1] Uni Hite Syst Corp, Dept Res & Dev, Kanagawa, Japan
[2] Univ Tsukuba, Grad Sch Syst & Informat Engn, Tsukuba, Ibaraki 305, Japan
来源
2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6 | 2006年
关键词
computed tomography; cone-beam reconstruction; closed sinusoid; C-arm;
D O I
10.1109/NSSMIC.2006.353750
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, we investigate cone-beam reconstruction for C-arm CT imaging using a closed sinusoid trajectory, which was firstly studied in [17] and can be considered a generalization of a saddle trajectory [7]. Inspired by the work in [10] and [13], we propose an exact shift-invariant filtered backprojection (FBP) reconstruction algorithm when the scanning trajectory is a closed sinusoid, or a perturbed closed sinusoid. We determine the filtering lines on the detector plane and give the steps to implement the algorithm for the planar detector geometry in a C-arm system. We also give two view independent reconstruction formulae for this geometry, both of which do not involve the view differencing operation and improve the reconstructed images by reducing discretization and sampling errors. The proposed algorithm is tested by simulation studies.
引用
收藏
页码:3480 / 3484
页数:5
相关论文
共 50 条
  • [41] C-arm rotation encoding with accelerometers
    Grzeda, Victor
    Fichtinger, Gabor
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2010, 5 (04) : 385 - 391
  • [42] Exposure From the Large C-Arm Versus the Mini C-Arm Using Hand/Wrist and Elbow Phantoms
    Singer, Gordon
    Herron, Brent
    Herron, David
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2011, 36A (04): : 628 - 631
  • [43] Artifacts Reduction in Low-Contrast Neurological Imaging with C-arm System
    Xia, Dan
    Chang, Yu-Bing
    Siddiqui, Adnan H.
    Zhang, Zheng
    Manak, Joe
    Sidky, Emil Y.
    Pan, Xiaochuan
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [44] Volume-of-interest, (VOI) imaging in C-arm flat-detector CT for high image quality at reduced dose
    Kolditz, Daniel
    Kyriakou, Yiannis
    Kalender, Willi A.
    MEDICAL PHYSICS, 2010, 37 (06) : 2719 - 2730
  • [45] Imaging Guidance with C-arm CT: Prospective Evaluation of Its Impact on Patient Radiation Exposure during Transhepatic Arterial Chemoembolization
    Kothary, Nishita
    Abdelmaksoud, Mohamed H. K.
    Tognolini, Alessia
    Fahrig, Rebecca
    Rosenberg, Jarrett
    Hovsepian, David M.
    Ganguly, Arundhuti
    Louie, John D.
    Kuo, William T.
    Hwang, Gloria L.
    Holzer, Alison
    Sze, Daniel Y.
    Hofmann, Lawrence V.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2011, 22 (11) : 1535 - 1544
  • [46] Low Dose, Low Noise and High Resolution Volume of Interest (VOI) Imaging in C-Arm Flat-Detector CT
    Kolditz, Daniel
    Kyriakou, Yiannis
    Kalender, Willi A.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [47] Patient-bounded extrapolation using low-dose priors for volume-of-interest imaging in C-arm CT
    Xia, Y.
    Bauer, S.
    Maier, A.
    Berger, M.
    Hornegger, J.
    MEDICAL PHYSICS, 2015, 42 (04) : 1787 - 1796
  • [48] C-arm CT在肝肿瘤TACE治疗中的应用
    张庆辉
    刘鹏程
    中国医学创新, 2012, 9 (19) : 156 - 159
  • [49] Reproducibility of Parenchymal Blood Volume Measurements Using an Angiographic C-arm CT System
    Mueller, Kerstin
    Fahrig, Rebecca
    Manhart, Michael
    Deuerling-Zheng, Yu
    Rosenberg, Jarrett
    Moore, Teri
    Ganguly, Arundhuti
    Kothary, Nishita
    ACADEMIC RADIOLOGY, 2016, 23 (11) : 1441 - 1445
  • [50] Scatter correction using a primary modulator on a clinical angiography C-arm CT system
    Bier, Bastian
    Berger, Martin
    Maier, Andreas
    Kachelriess, Marc
    Ritschl, Ludwig
    Mueller, Kerstin
    Choi, Jang-Hwan
    Fahrig, Rebecca
    MEDICAL PHYSICS, 2017, 44 (09) : E125 - E137