A flexible enzyme-electrode sensor with cylindrical working electrode modified with a 3D nanostructure for implantable continuous glucose monitoring

被引:50
作者
Pu, Zhihua [1 ]
Tu, Jiaan [1 ]
Han, Ruixue [2 ]
Zhang, Xingguo [1 ]
Wu, Jianwei [1 ]
Fang, Chao [1 ]
Wu, Hao [1 ]
Zhang, Xiaoli [1 ]
Yu, Haixia [2 ]
Li, Dachao [1 ]
机构
[1] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
[2] Tianjin Univ, Tianjin Key Lab Biomed Detecting Tech & Instrumen, Tianjin 300072, Peoples R China
基金
国家重点研发计划; 中国国家自然科学基金;
关键词
MICROELECTRODES; OPPORTUNITIES; BIOSENSORS; PATTERNS; TIME;
D O I
10.1039/c8lc00908b
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel cylindrical flexible enzyme-electrode sensor was fabricated with a bigger working electrode (WE) surface than the traditional pin-like one for implantable continuous glucose monitoring. On the WE surface, a 3D nanostructure consisting of graphene and platinum nanoparticles was constructed to enhance the sensitivity; in conjunction with the bigger WE, this nanostructure enabled hypoglycemia detection, which is still a big challenge in clinics. The cylindrical sensor was fabricated by rotated inkjet printing which enabled direct patterning of microstructures on a curved surface, thus overcoming the restriction of the traditional planar micromachining by photolithography. Specifically, the cylindrical substrate (polyetheretherketone, PEEK) was modified by (3-aminopropyl) trimethoxysilane and (3-mercaptopropyl) trimethoxysilane to facilitate surface wettability, which discourages the coalescence of adjacent droplets, and to facilitate the adhesion of metals to PEEK in order to construct robust electrodes. A synchronous heating method was used to evaporate the solvent of the droplets quickly to prevent them from running along the cylindrical surface, which affects the formation of the printed electrode significantly. In vitro experimental results showed that the proposed sensor was able to detect the glucose concentration ranging from 0 to 570 mg dL(-1) which demonstrated its capability for physiological glucose detection. In vivo experiments were conducted with rats, and the measurement results recorded using the implanted cylindrical sensor showed great compliance to those recorded using a commercial glucometer which exhibited the viability of the proposed sensor for implantable continuous glucose monitoring, even under the hypoglycemic conditions.
引用
收藏
页码:3570 / 3577
页数:8
相关论文
共 33 条
  • [1] Low-Cost Inkjet Printing Technology for the Rapid Prototyping of Transducers
    Ando, Bruno
    Baglio, Salvatore
    Bulsara, Adi R.
    Emery, Teresa
    Marletta, Vincenzo
    Pistorio, Antonio
    [J]. SENSORS, 2017, 17 (04)
  • [2] Diabetes mellitus, blood glucose and the risk of atrial fibrillation: A systematic review and meta-analysis of cohort studies
    Aune, Dagfinn
    Feng, Tingting
    Schlesinger, Sabrina
    Janszky, Imre
    Norat, Teresa
    Riboli, Elio
    [J]. JOURNAL OF DIABETES AND ITS COMPLICATIONS, 2018, 32 (05) : 501 - 511
  • [3] Current and Emerging Technology for Continuous Glucose Monitoring
    Chen, Cheng
    Zhao, Xue-Ling
    Li, Zhan-Hong
    Zhu, Zhi-Gang
    Qian, Shao-Hong
    Flewitt, Andrew J.
    [J]. SENSORS, 2017, 17 (01)
  • [4] The fabrication of submicron patterns on curved substrates using a polydimethylsiloxane film mould
    Choi, WM
    Park, OO
    [J]. NANOTECHNOLOGY, 2004, 15 (12) : 1767 - 1770
  • [5] Evaluating the clinical accuracy of two continuous glucose sensors using continuous glucose-error grid analysis
    Clarke, WL
    Anderson, S
    Farhy, L
    Breton, M
    Gonder-Frederick, L
    Cox, D
    Kovatchev, B
    [J]. DIABETES CARE, 2005, 28 (10) : 2412 - 2417
  • [6] Interstitial Fluid Glucose Is Not Just a Shifted-in-Time but a Distorted Mirror of Blood Glucose: Insight from an In Silico Study
    Cobelli, Claudio
    Schiavon, Michele
    Man, Chiara Dalla
    Basu, Ananda
    Basu, Rita
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2016, 18 (08) : 505 - 511
  • [7] Inkjet and Aerosol Jet Printing of Electrochemical Devices for Energy Conversion and Storage
    Deiner, L. Jay
    Reitz, Thomas L.
    [J]. ADVANCED ENGINEERING MATERIALS, 2017, 19 (07)
  • [8] QoS-Constrained Transceiver Design and Power Splitting for Downlink Multiuser MIMO SWIPT Systems
    Dong, Anming
    Zhang, Haixia
    Wu, Dalei
    Yuan, Dongfeng
    [J]. 2016 IEEE INTERNATIONAL CONFERENCE ON COMMUNICATIONS (ICC), 2016,
  • [9] Inkjet printing wearable electronic devices
    Gao, Meng
    Li, Lihong
    Song, Yanlin
    [J]. JOURNAL OF MATERIALS CHEMISTRY C, 2017, 5 (12) : 2971 - 2993