Comparison of Prothrombin Time Derived From CoaguChek XS and Laboratory Test According to Fibrinogen Level

被引:8
作者
Kim, Sue Jung [1 ]
Lee, Eun Young [1 ]
Park, Rojin [2 ]
Kim, Juwon [3 ]
Song, Jaewoo [1 ]
机构
[1] Yonsei Univ, Dept Lab Med, Coll Med, Seoul 120752, South Korea
[2] Soon Chun Hyang Univ Hosp, Dept Lab Med, Seoul, South Korea
[3] Yonsei Univ, Wonju Coll Med, Dept Lab Med, Wonju, South Korea
关键词
prothrombin time; INR; fibrinogen; point-of-care; CoaguChek XS; QUALITY ASSESSMENT; CARE; MANAGEMENT; SPLIT;
D O I
10.1002/jcla.21722
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BackgroundCoaguChek XS is one of the most widely used point-of-care (POC) devices to evaluate prothrombin time for monitoring oral anticoagulant therapy. Unlike laboratory methods, it detects electrical signals produced by thrombin activity to derive the international normalized ratio (INR). Therefore, we hypothesized that laboratory methods and CoaguChek XS could produce different results according to fibrinogen level. MethodsWe compared INR values obtained from the CoaguChek XS and conventional laboratory method with 91 plasma samples covering a wide range of fibrinogen levels. ResultsThe samples were stratified into low, mid, and high fibrinogen groups by fibrinogen levels of <130 mg/dl, 130-450 mg/dl, and >450 mg/dl, respectively. The mean INR difference of the low fibrinogen group was significantly different from that of the mid or high fibrinogen group (P < 0.001). In the low fibrinogen group, CoaguChek XS INR showed a negative bias compared with the laboratory INR, while the mid and high fibrinogen groups had positive bias. ConclusionOur results suggest that patient selection according to fibrinogen status should precede the implementation of POC testing using CoaguChek XS. Also, periodic comparisons between CoaguChek XS and laboratory INR results should be continued during the use of CoaguChek XS.
引用
收藏
页码:28 / 31
页数:4
相关论文
共 20 条
[1]   Pharmacology and management of the vitamin K antagonists [J].
Ansell, Jack ;
Hirsh, Jack ;
Hylek, Elaine ;
Jacobson, Alan ;
Crowther, Mark ;
Palareti, Gualtiero .
CHEST, 2008, 133 (06) :160S-198S
[2]   Patient self-testing and self-management of oral anticoagulation [J].
Braun, Siegmund ;
Spannagl, Michael ;
Voeller, Heinz .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 393 (05) :1463-1471
[3]   International normalised ratio (INR) measured on the CoaguChek S and XS compared with the laboratory for determination of precision and accuracy [J].
Christensen, Thomas D. ;
Larsen, Torben B. ;
Jensen, Claus ;
Maegaard, Marianne ;
Sorensen, Benny .
THROMBOSIS AND HAEMOSTASIS, 2009, 101 (03) :563-569
[4]  
Garcia-Alamino JM., 2010, COCHRANE DB SYST REV, DOI 10.1002/14651858.CD003839.pub2.CD003839
[5]   Patient self-testing is a reliable and acceptable alternative to laboratory INR monitoring [J].
Gardiner, C ;
Williams, K ;
Mackie, IJ ;
Machin, SJ ;
Cohen, H .
BRITISH JOURNAL OF HAEMATOLOGY, 2005, 128 (02) :242-247
[6]   External quality assessment (EQA) for CoaguChek monitors [J].
Jespersen, Joergen ;
Poller, Leon ;
van den Besselaar, A. M. H. P. ;
van der Meer, Felix J. M. ;
Palareti, Gualtiero ;
Tripodi, Armando ;
Shiach, Caroline ;
Keown, Michelle ;
Ibrahim, Saied .
THROMBOSIS AND HAEMOSTASIS, 2010, 103 (05) :936-941
[7]   A national field study of quality assessment of CoaguChek point-of-care testing prothrombin time monitors [J].
Meijer, Piet ;
Kluft, Cornelis ;
Poller, Leon ;
van der Meer, Felix J. M. ;
Keown, Michelle ;
Ibrahim, Saied ;
van den Besselaar, Anton M. H. P. ;
Tripodi, Armando ;
Jespersen, Jorgen .
AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2006, 126 (05) :756-761
[8]   Results of the performance verification of the CoaguChek XS system [J].
Plesch, W. ;
Wolf, T. ;
Breitenbeck, N. ;
Dikkeschei, L. D. ;
Cervero, A. ;
Perez, P. L. ;
van den Besselaar, A. M. H. P. .
THROMBOSIS RESEARCH, 2008, 123 (02) :381-389
[9]   Coagulopathy in liver diseases [J].
Pluta, A. ;
Gutkowski, K. ;
Hartleb, M. .
ADVANCES IN MEDICAL SCIENCES, 2010, 55 (01) :16-21
[10]   European Concerted Action on Anticoagulation. Quality assessment of the CoaguChek mini and TAS PT-NC point-of-care whole-blood prothrombin time monitors [J].
Poller, L ;
Keown, M ;
Chauhan, N ;
van den Besselaar, AMHP ;
Tripodi, A ;
Shiach, C ;
Jespersen, J .
CLINICAL CHEMISTRY, 2004, 50 (03) :537-544