Validity and reliability of the T-Scan®III for measuring force under laboratory conditions

被引:49
作者
Cerna, M. [1 ]
Ferreira, R. [2 ]
Zaror, C. [2 ]
Navarro, P. [3 ]
Sandoval, P. [2 ]
机构
[1] Univ La Frontera, Fac Sci & Engn, Dept Phys Sci, Temuco, Chile
[2] Univ La Frontera, Fac Dent, Dept Pediat Dent & Orthodont, Destristry Program, Temuco, Chile
[3] Univ La Frontera, Fac Dent, Temuco, Chile
关键词
validity; reliability; T-Scan((R))III; bite force; bite registration; dental occlusion;
D O I
10.1111/joor.12284
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Although measuring bite force is an important indicator of the health of the masticatory system, few commercially available transducers have been validated for routine clinical use. T-Scan((R))III Occlusal Analysis System allows to record the bite force distribution, indicating its relative intensity and occlusal timing. Nevertheless, even fewer studies have evaluated the validity and reliability of the latest generation of the T-Scan((R)) occlusal analysis system. To determine the validity and reliability of the T-Scan((R))III system when measuring total absolute bite force under laboratory conditions. Known forces were applied to 18 T-Scan((R))III sensors, which were classified into two groups differentiated by their production series. Both Lin's concordance correlation coefficient (CCC) and the intra-class correlation coefficient (ICC) were used to assess the system's reliability and validity. Considering all the sensors studied, a substantial level (Lin's CCC 0969) and a very good level of reliability (CCI 0994) were obtained. When evaluating the validity of the system, a poor (Lin's CCC 0530) and moderate (ICC 0693) agreement were also obtained. The main factor that negatively influenced the validity of the T-Scan((R))III under these study conditions was the significant difference in the behaviour of the two sensor groups. The T-Scan((R))III showed a high degree of reliability when used to perform consecutive measurements. However, the system showed an insufficient degree of validity for measuring absolute force when estimating total occlusal force under laboratory conditions.
引用
收藏
页码:544 / 551
页数:8
相关论文
共 18 条
[1]  
Alfaro P, 2012, REV ADM, V69, P108
[2]  
Baldini Alberto, 2012, Ann Stomatol (Roma), V3, P51
[3]  
Cabral CW, 2006, BRAZ J ORAL SCI, V5, P1239
[4]   Analysis of two methods for occlusal contact registration with the T-Scan system [J].
Cartagena, AG ;
Sequeros, OG ;
Garcia, VCG .
JOURNAL OF ORAL REHABILITATION, 1997, 24 (06) :426-432
[5]   In vitro study on the sensibility and reproducibility of the new T-Scan (R) III HD system [J].
da Silva Martins, Mario Jorge ;
Caramelo, Francisco Jose ;
Ramalho da Fonseca, Julio Andre ;
Gomes Nicolau, Pedro Miguel .
REVISTA PORTUGUESA DE ESTOMATOLOGIA MEDICINA DENTARIA E CIRURGIA MAXILOFACIAL, 2014, 55 (01) :14-22
[6]   Computerized analysis of occlusal contacts in bruxism patients treated with occlusal splint therapy [J].
Gumus, Hasan Onder ;
Kilinc, Halil Ibrahim ;
Tuna, Suleyman Hakan ;
Ozcan, Nihal .
JOURNAL OF ADVANCED PROSTHODONTICS, 2013, 5 (03) :256-261
[7]  
Kerstein Robert B, 2008, Dent Implantol Update, V19, P41
[8]   Time-dependent analysis and representation of force distribution and occlusion contact in the masticatory cycle [J].
Koos, Bernd ;
Hoeller, Johanna ;
Schille, Christine ;
Godt, Arnim .
JOURNAL OF OROFACIAL ORTHOPEDICS-FORTSCHRITTE DER KIEFERORTHOPADIE, 2012, 73 (03) :204-214
[9]   Precision of an Instrumentation-based Method of Analyzing Occlusion and its Resulting Distribution of Forces in the Dental Arch [J].
Koos, Bernd ;
Godt, Arnim ;
Schille, Christine ;
Goez, Gernot .
JOURNAL OF OROFACIAL ORTHOPEDICS-FORTSCHRITTE DER KIEFERORTHOPADIE, 2010, 71 (06) :403-410
[10]   A CONCORDANCE CORRELATION-COEFFICIENT TO EVALUATE REPRODUCIBILITY [J].
LIN, LI .