Background. The incidence of severe sternal wound complications in high-risk cardiac patients presents a significant need for more stabile sternal fixation techniques after median sternotomy procedures. Rigid metal plates, a potential alternative to wire fixation, are thought to promote faster sternal healing by reducing motion at the wound site. The goal of this study was to compare the stability provided by commercially available sternal plates with standard wires using an in vitro model. Methods. Lateral distraction tests were conducted on bisected polyurethane sternal models fixed with either a standard 7 wire configuration (n = 5) or a 3 plate configuration (n = 3). To assure controlled loading, the sternal models were attached to a computer-controlled test machine by a novel tethering system that distributes the total force (180N) equally to eight locations on the sternum. Stability was defined as the ability to restrict sternal separation at seven locations along the midline quantified using digital image analysis. Results. Our results indicate that rigid plate fixation significantly reduced lateral motion relative to wire fixation. The lower sternal region most noticeably benefited from plate fixation as the splaying observed for wire fixation was reduced. Conclusions. Under these loading conditions, plating increased stability at the midline compared to wires; this increased stability may facilitate the recovery of high-risk patients undergoing cardiac operation. To enhance in vitro testing methods, future studies should incorporate additional in vivo loading conditions applied to the sternum. Alternate plating configurations should also be examined to further increase stability.
引用
收藏
页码:962 / 968
页数:7
相关论文
共 27 条
  • [1] *AM HEART ASS, 2002, HEART DIS STROK STAT
  • [2] A biomechanical study of median sternotomy closure techniques
    Casha, AR
    Yang, L
    Kay, PH
    Saleh, M
    Cooper, GJ
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1999, 15 (03) : 365 - 369
  • [3] BIOMECHANICAL STUDY OF STERNAL CLOSURE TECHNIQUES
    CHENG, W
    CAMERON, DE
    WARDEN, KE
    FONGER, JD
    GOTT, VL
    [J]. ANNALS OF THORACIC SURGERY, 1993, 55 (03) : 737 - 740
  • [4] Sternal wound reconstruction with transverse plate fixation
    Cicilioni, OJ
    Stieg, FH
    Papanicolaou, G
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2005, 115 (05) : 1297 - 1303
  • [5] The effect of mechanical stability on local vascularization and tissue differentiation in callus healing
    Claes, L
    Eckert-Hübner, K
    Augat, P
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (05) : 1099 - 1105
  • [6] Cohen DJ, 2002, ANN THORAC SURG, V73, P563
  • [7] POSTERIOR STABILIZATION OF CERVICAL-SPINE FRACTURES AND SUBLUXATIONS USING PLATES AND SCREWS
    COOPER, PR
    COHEN, A
    ROSIELLO, A
    KOSLOW, M
    [J]. NEUROSURGERY, 1988, 23 (03) : 300 - 306
  • [8] Lower sternal reinforcement improves the stability of sternal closure
    Dasika, UK
    Trumble, DR
    Magovern, JA
    [J]. ANNALS OF THORACIC SURGERY, 2003, 75 (05) : 1618 - 1621
  • [9] Treatment of sternal nonunion with the Dall-Miles cable system
    Eich, BS
    Heinz, TR
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2000, 106 (05) : 1075 - 1078
  • [10] EXPERIMENTAL AND CLINICAL-RESULTS OF NEW ADVANCES IN THE TREATMENT OF FACIAL TRAUMA
    EWERS, R
    HARLE, F
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 1985, 75 (01) : 25 - 31