Conventional instruments are more accurate for measuring the depth of the tibial cut than computer-assisted surgery in total knee arthroplasty: a prospective study

被引:14
作者
Klasan, Antonio [1 ]
Putnis, Sven Edward [1 ]
Grasso, Samuel [1 ]
Neri, Thomas [1 ]
Coolican, Myles Raphael [1 ]
机构
[1] Sydney Orthopaed Res Inst, 445 Victoria Ave, Chatswood, NSW 2067, Australia
关键词
Knee; Arthroplasty; Tibial component; Bone cut; Navigation; PATIENT-SPECIFIC INSTRUMENTATION; NAVIGATION SYSTEM; CUTTING ERRORS; REPLACEMENT; ALIGNMENT; RESECTION; PAIN;
D O I
10.1007/s00402-020-03403-9
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction The most commonly used tool for implant positioning are conventional instruments (CI) followed by computer-assisted surgery (CAS). A number of studies have investigated the cutting error of the tibial component when CAS is used, but most of them were focused on the cutting angles. The accuracy of CAS to determine the depth of the cut has not received much attention, even though implications are similar or worse, than with an angle mismatch. Materials and methods This was an ethics board approved, prospective study of 23 consecutive varus TKAs by a single surgeon. Implant positioning was performed using CAS; however, the depth of the tibial cut was determined with both CAS and CI. Targeted alignment was the mechanical axis and 3 degrees of posterior slope. The planned and the achieved cut, as determined by CAS needed to match. The achieved cut was then measured using a caliper and compared to the depth of the cut as per CAS. Medial and lateral cuts were analyzed separately. Analysis of variance and Bland-Altman plots were used for the comparison. Results Mean medial navigated cut was 6.3 (+/- 2.2) mm, mean measured medial cut was 6.6 (+/- 2.3) mm. Mean lateral navigated cut was 8.9 (+/- 1.8) mm, mean measured lateral cut was 8.8 (+/- 1.5) mm. There was a statistical significance for both the medial (p < 0.001) and the lateral (p = 0.004) navigated and measured cuts. Conclusions The results of this study suggest that the tibial cut depth, measured by the navigation, does not match the actual bony cuts performed, even if a perfect cut was achieved in both sagittal and coronal plane. Surgeons should be aware of the measurement error in the navigation system and potentially add an additional step for verifying the achieved depth of the cut.
引用
收藏
页码:801 / 806
页数:6
相关论文
共 50 条
[41]   Computer-assisted surgery and patient-specific instrumentation improve the accuracy of tibial baseplate rotation in total knee arthroplasty compared to conventional instrumentation: a systematic review and meta-analysis [J].
Tandogan, Reha N. ;
Kort, Nanne P. ;
Ercin, Ersin ;
van Rooij, Floris ;
Nover, Luca ;
Saffarini, Mo ;
Hirschmann, Michael T. ;
Becker, Roland ;
Dejour, David .
KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2022, 30 (08) :2654-2665
[42]   No Benefit of Robotic-Assisted over Computer-Assisted Surgery for Achieving Neutral Coronal Alignment in Total Knee Arthroplasty [J].
Klasan, Antonio ;
Anelli-Monti, Victoria ;
Huber, Stephanie ;
Zacherl, Maximillian ;
Hofstaetter, Jochen ;
Kammerlander, Christian ;
Sadoghi, Patrick .
JOURNAL OF KNEE SURGERY, 2024, 37 (01) :8-13
[43]   Computer-Assisted Kinematic and Mechanical Axis Total Knee Arthroplasty: A Prospective Randomized Controlled Trial of Bilateral Simultaneous Surgery [J].
McEwen, Peter J. ;
Dlaska, Constantine E. ;
Jovanovic, Ivana A. ;
Doma, Kenji ;
Brandon, Benjamin J. .
JOURNAL OF ARTHROPLASTY, 2020, 35 (02) :443-450
[44]   A retrospective study of articular surface mounting computer-assisted total knee arthroplasty in alignment, knee function, and survival [J].
Ng, Eugene Pak-Lin ;
Wong, Hing-Cheong .
JOURNAL OF ORTHOPAEDICS TRAUMA AND REHABILITATION, 2023, 30 (01)
[45]   Re-learning curve for conventional total knee arthroplasty following 30 consecutive computer-assisted total knee arthroplasties [J].
Nicholson, Luke T. ;
Trofa, David ;
Smith, Eric .
COMPUTER AIDED SURGERY, 2013, 18 (3-4) :63-67
[46]   The predictive factors of secondary patellar resurfacing in computer-assisted total knee arthroplasty. A prospective cohort study [J].
Franck, Florent ;
Ouanezar, Herve ;
Jacquel, Alexandre ;
Pibarot, Vincent ;
Wegrzyn, Julien .
INTERNATIONAL ORTHOPAEDICS, 2018, 42 (05) :1051-1060
[47]   Postoperative radiologic outcome comparison between conventional and computer-assisted navigation total knee arthroplasty in extra-articular tibia vara [J].
Cho, Yongun ;
Shin, Hun-Kyu ;
Kim, Eugene ;
Park, Jai Hyung ;
Park, Se-Jin ;
Jang, Taedong ;
Jeong, Hwa-Jae .
JOURNAL OF ORTHOPAEDIC SURGERY, 2020, 28 (01)
[48]   Functional outcome after computer-assisted versus conventional total knee arthroplasty: a randomized controlled study [J].
Jörg Lützner ;
Klaus-Peter Günther ;
Stephan Kirschner .
Knee Surgery, Sports Traumatology, Arthroscopy, 2010, 18 :1339-1344
[49]   Revisiting Short-term Outcomes of Conventional and Computer-Assisted Total Knee Arthroplasty: A Population-based Study [J].
Ahmed, Abdalrahman G. ;
Tian, Yao ;
Hasan, Mohamed ;
Harris, Alexandra ;
Ghomrawi, Hassan M. K. .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS GLOBAL RESEARCH AND REVIEWS, 2022, 6 (06)
[50]   The influence of computer-assisted surgery on rotational, coronal and sagittal alignment in revision total knee arthroplasty [J].
Meijer, Marrigje F. ;
Stevens, Martin ;
Boerboom, Alexander L. ;
Bulstra, Sjoerd K. ;
Reininga, Inge H. F. .
BMC MUSCULOSKELETAL DISORDERS, 2014, 15