共 30 条
Conventional instruments are more accurate for measuring the depth of the tibial cut than computer-assisted surgery in total knee arthroplasty: a prospective study
被引:16
作者:
Klasan, Antonio
[1
]
Putnis, Sven Edward
[1
]
Grasso, Samuel
[1
]
Neri, Thomas
[1
]
Coolican, Myles Raphael
[1
]
机构:
[1] Sydney Orthopaed Res Inst, 445 Victoria Ave, Chatswood, NSW 2067, Australia
关键词:
Knee;
Arthroplasty;
Tibial component;
Bone cut;
Navigation;
PATIENT-SPECIFIC INSTRUMENTATION;
NAVIGATION SYSTEM;
CUTTING ERRORS;
REPLACEMENT;
ALIGNMENT;
RESECTION;
PAIN;
D O I:
10.1007/s00402-020-03403-9
中图分类号:
R826.8 [整形外科学];
R782.2 [口腔颌面部整形外科学];
R726.2 [小儿整形外科学];
R62 [整形外科学(修复外科学)];
学科分类号:
摘要:
Introduction The most commonly used tool for implant positioning are conventional instruments (CI) followed by computer-assisted surgery (CAS). A number of studies have investigated the cutting error of the tibial component when CAS is used, but most of them were focused on the cutting angles. The accuracy of CAS to determine the depth of the cut has not received much attention, even though implications are similar or worse, than with an angle mismatch. Materials and methods This was an ethics board approved, prospective study of 23 consecutive varus TKAs by a single surgeon. Implant positioning was performed using CAS; however, the depth of the tibial cut was determined with both CAS and CI. Targeted alignment was the mechanical axis and 3 degrees of posterior slope. The planned and the achieved cut, as determined by CAS needed to match. The achieved cut was then measured using a caliper and compared to the depth of the cut as per CAS. Medial and lateral cuts were analyzed separately. Analysis of variance and Bland-Altman plots were used for the comparison. Results Mean medial navigated cut was 6.3 (+/- 2.2) mm, mean measured medial cut was 6.6 (+/- 2.3) mm. Mean lateral navigated cut was 8.9 (+/- 1.8) mm, mean measured lateral cut was 8.8 (+/- 1.5) mm. There was a statistical significance for both the medial (p < 0.001) and the lateral (p = 0.004) navigated and measured cuts. Conclusions The results of this study suggest that the tibial cut depth, measured by the navigation, does not match the actual bony cuts performed, even if a perfect cut was achieved in both sagittal and coronal plane. Surgeons should be aware of the measurement error in the navigation system and potentially add an additional step for verifying the achieved depth of the cut.
引用
收藏
页码:801 / 806
页数:6
相关论文