Introduction The most commonly used tool for implant positioning are conventional instruments (CI) followed by computer-assisted surgery (CAS). A number of studies have investigated the cutting error of the tibial component when CAS is used, but most of them were focused on the cutting angles. The accuracy of CAS to determine the depth of the cut has not received much attention, even though implications are similar or worse, than with an angle mismatch. Materials and methods This was an ethics board approved, prospective study of 23 consecutive varus TKAs by a single surgeon. Implant positioning was performed using CAS; however, the depth of the tibial cut was determined with both CAS and CI. Targeted alignment was the mechanical axis and 3 degrees of posterior slope. The planned and the achieved cut, as determined by CAS needed to match. The achieved cut was then measured using a caliper and compared to the depth of the cut as per CAS. Medial and lateral cuts were analyzed separately. Analysis of variance and Bland-Altman plots were used for the comparison. Results Mean medial navigated cut was 6.3 (+/- 2.2) mm, mean measured medial cut was 6.6 (+/- 2.3) mm. Mean lateral navigated cut was 8.9 (+/- 1.8) mm, mean measured lateral cut was 8.8 (+/- 1.5) mm. There was a statistical significance for both the medial (p < 0.001) and the lateral (p = 0.004) navigated and measured cuts. Conclusions The results of this study suggest that the tibial cut depth, measured by the navigation, does not match the actual bony cuts performed, even if a perfect cut was achieved in both sagittal and coronal plane. Surgeons should be aware of the measurement error in the navigation system and potentially add an additional step for verifying the achieved depth of the cut.
引用
收藏
页码:801 / 806
页数:6
相关论文
共 30 条
[1]   Improving outcomes in total knee arthroplasty-do navigation or customized implants have a role? [J].
Beal, Matthew D. ;
Delagramaticas, Dimitri ;
Fitz, David .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2016, 11
[2]   Tibial component failure mechanisms in total knee arthroplasty [J].
Berend, ME ;
Ritter, MA ;
Meding, JB ;
Faris, PM ;
Keating, EM ;
Redelman, R ;
Faris, GW ;
Davis, KE .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) :26-34
[3]   Influence of anxiety and pain catastrophizing on the course of pain within the first year after uncomplicated total knee replacement: a prospective study [J].
Bierke, Sebastian ;
Petersen, Wolf .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2017, 137 (12) :1735-1742
[4]   Accuracy of a hand-held surgical navigation system for tibial resection in total knee arthroplasty [J].
Bugbee, William D. ;
Kermanshahi, Arash Y. ;
Munro, Michelle M. ;
McCauley, Julie C. ;
Copp, Steven N. .
KNEE, 2014, 21 (06) :1225-1228
[5]   Computer-assisted Total Knee Arthroplasty Is Currently of No Proven Clinical Benefit: A Systematic Review [J].
Burnett, R. Stephen J. ;
Barrack, Robert L. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (01) :264-276
[6]   Patient age of less than 55 years is not an independent predictor of functional improvement or satisfaction after total knee arthroplasty [J].
Clement, N. D. ;
Walker, L. C. ;
Bardgett, M. ;
Weir, D. ;
Holland, J. ;
Gerrand, C. ;
Deehan, D. J. .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2018, 138 (12) :1755-1763
[7]   I can't get no satisfaction after my total knee replacement RHYMES AND REASONS [J].
Dunbar, M. J. ;
Richardson, G. ;
Robertsson, O. .
BONE & JOINT JOURNAL, 2013, 95B (11) :148-152
[8]   Cutting and implanting errors in minimally invasive total knee arthroplasty using a navigation system [J].
Hasegawa, Masahiro ;
Yoshida, Kakunoshin ;
Wakabayashi, Hiroki ;
Sudo, Akihiro .
INTERNATIONAL ORTHOPAEDICS, 2013, 37 (01) :27-30
[9]   Calipered Kinematically Aligned Total Knee Arthroplasty: An Accurate Technique That Improves Patient Outcomes and Implant Survival [J].
Howell, Stephen M. .
ORTHOPEDICS, 2019, 42 (03) :126-+
[10]   Porous Tantalum Metaphyseal Cones for Severe Tibial Bone Loss in Revision Knee Arthroplasty A Five to Nine-Year Follow-up [J].
Kamath, Atul F. ;
Lewallen, David G. ;
Hanssen, Arlen D. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2015, 97A (03) :216-223