Analysis of polar organic micropollutants in water with ion chromatography electrospray mass spectrometry

被引:101
作者
Bauer, KH
Knepper, TP
Maes, A
Schatz, V
Voihsel, M
机构
[1] ESWE, Inst Water Res & Water Technol, D-65201 Wiesbaden, Germany
[2] Perkin Elmer Sciex, D-63225 Langen, Germany
[3] Riedwerke, D-64521 Gross Gerau, Germany
关键词
water analysis; mass spectrometry; polar organic micropollutants; glycophosphate; aminoethylphosphonic acid; ethylenediamintetraacetate;
D O I
10.1016/S0021-9673(99)00048-5
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The coupling of ion chromatography (IC) with electrospray mass spectrometry (ES-MS) opens new ways for the determination of polar organic micropollutants in water samples. The technique of conductivity suppression has been found to reduce the background signal in the range of about two-orders of magnitude leading to a significant increase in sensitivity. In addition, the formation of salt adducts has been avoided. The usefulness of this method was proven on several polar and environmentally relevant micropollutants such as the herbicide glyphosate and its metabolite aminomethylphosphonic acid (AMPA), the chelating agent ethylenediamine tetraacetate (EDTA) and diacetonketogulonic acid (DAG). This present study has shown that IC-ES-MS is a simple, sensitive and quick method for the determination of these polar organic traces in water samples after separation on an anion-exchange column without any derivatization. In this work, several possibilities of applications of IC-ES-MS (with varying conditions) are presented. Analysis of glyphosate, AMPA, DAG and EDTA in ground and surface waters has been achieved by IC-ES-MS without additional sample preparation at a concentration level of 1 mu g/l. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:117 / 128
页数:12
相关论文
共 16 条
[1]  
Bauer K.H., 1997, VOM WASSER, V88, P351
[2]  
BAUER KH, 1993, VOM WASSER, V81, P167
[3]  
BRAUCH HJ, 1996, ARW GEWAESSERZUSTAND, P34
[4]   Electrospray ion chromatography tandem mass spectrometry of oxyhalides at sub-ppb levels [J].
Charles, L ;
Pépin, D .
ANALYTICAL CHEMISTRY, 1998, 70 (02) :353-359
[5]   AUTOMATED ONLINE LIQUID-SOLID EXTRACTION FOLLOWED BY LIQUID CHROMATOGRAPHY-HIGH-FLOW PNEUMATICALLY ASSISTED ELECTROSPRAY MASS-SPECTROMETRY FOR THE DETERMINATION OF ACIDIC HERBICIDES IN ENVIRONMENTAL WATERS [J].
CHIRON, S ;
PAPILLOUD, S ;
HAERDI, W ;
BARCELO, D .
ANALYTICAL CHEMISTRY, 1995, 67 (09) :1637-1643
[6]   ION CHROMATOGRAPHY MASS-SPECTROMETRY FOR THE DETERMINATION OF ORGANIC AMMONIUM AND SULFATE COMPOUNDS [J].
CONBOY, JJ ;
HENION, JD ;
MARTIN, MW ;
ZWEIGENBAUM, JA .
ANALYTICAL CHEMISTRY, 1990, 62 (08) :800-807
[7]  
*EDIN, 38413P8 EDIN EDTA DT
[8]  
KARRENBROCK F, 1995, ENTWICKLUNG BELASTUN, P67
[9]  
LINDNER K, 1996, ERFASSUNG IDENTIFIZI
[10]  
*METR, MAN AN SEP COL DUAL, V2