Bedside Blood Glucose Monitoring in Critically Ill Patients: Comparison Between Arterial and Capillary Glucose

被引:2
|
作者
Li, Xu [1 ]
Ma, Yanquan [2 ]
Chen, Tianlu [1 ]
Tang, Jie [1 ]
Ma, Xiaochun [1 ]
机构
[1] China Med Univ, Affiliated Hosp 1, Dept Intens Care Unit, North Nanjing St 115, Shenyang 110001, Liaoning, Peoples R China
[2] Gen Hosp Panjin Oilfield, Dept Intens Care Unit, Panjin, Liaoning, Peoples R China
来源
AMERICAN JOURNAL OF THE MEDICAL SCIENCES | 2017年 / 354卷 / 05期
关键词
Terms: Critically ill patient; Blood glucose monitoring; Glucometer; Blood gas analyzer; INTENSIVE-CARE-UNIT; INSULIN THERAPY; HYPOGLYCEMIA; GLUCOMETERS; ANEMIA; ADULTS; ERROR; RISK;
D O I
10.1016/j.amjms.2017.05.005
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Critically ill patients are at high risk of hypoglycemia and are particularly vulnerable to unrecognized hypoglycemia. Close blood glucose monitoring is therefore crucial. There are several options to conduct frequent blood glucose measurement and a number of conditions in intensive care unit patients may affect the accuracy of blood glucose measurement. The aim of the study was to compare the accuracy of capillary glucose by bedside glucometer with arterial samples by bedside glucometer and arterial samples by blood gas analyzer in critically ill patients through a prospective case-control study. Materials and Methods: Arterial and capillary samples from 60 patients were taken simultaneously and were tested immediately at the bedside. Results of the paired measurements were compared and expressed as a correlation coefficient. Resuls: Capillary glucose in the study group and control group were 9.73 +/- 2.28 mmol/L and 8.9 +/- 1.86 mmol/L, respectively; mean arterial glucose measured by glucometer in the study group and control group were 9.25 +/- 2.05 mmol/L and 8.4 +/- 1.89 mmol/L, respectively; and mean arterial glucose measured by blood gas analyzer in the study group and control group were 8.41 +/- 1.9 mmol/L and 8.24 +/- 1.5 mmol/L, respectively. Correlation between capillary values and arterial values measured by glucometer was less in the study group (r = 0.936, P < 0.001 and r = 0.973, P < 0.001). Correlation between capillary values measured by glucometer and arterial values measured by blood gas analyzer was also less in the study group (r = 0.897, P = 0.001 and r = 0.964, P < 0.001). Conclusions: Capillary blood glucose monitoring is reliable only in a selected group of critically ill patients.
引用
收藏
页码:458 / 461
页数:4
相关论文
共 50 条
  • [1] Accuracy of bedside capillary blood glucose measurements in critically ill patients
    Critchell, C. Dana
    Savarese, Vincent
    Callahan, Amy
    Aboud, Christine
    Jabbour, Serge
    Marik, Paul
    INTENSIVE CARE MEDICINE, 2007, 33 (12) : 2079 - 2084
  • [2] Accuracy of bedside capillary blood glucose measurements in critically ill patients
    C. Dana Critchell
    Vincent Savarese
    Amy Callahan
    Christine Aboud
    Serge Jabbour
    Paul Marik
    Intensive Care Medicine, 2007, 33 : 2079 - 2084
  • [3] Accuracy of bedside capillary blood glucose measurements in critically ill patients.
    Critchell, Caroline D.
    Callahan, Amy
    Aboud, Christine
    Jabbour, Serge
    Marik, Paul
    CRITICAL CARE MEDICINE, 2006, 34 (12) : A68 - A68
  • [4] Comparison between arterial and capillary blood glucose monitoring in patients with shock
    Juneja, Deven
    Pandey, Rameshwar
    Singh, Omender
    EUROPEAN JOURNAL OF INTERNAL MEDICINE, 2011, 22 (03) : 241 - 244
  • [5] COMPARISON BETWEEN ARTERIAL AND CAPILLARY BLOOD GLUCOSE MONITORING IN PATIENTS WITH SHOCK
    Juneja, D.
    Pandey, R.
    Javeri, Y.
    Singh, O.
    Gupta, M.
    Singh, G.
    Gupta, C.
    Kumar, P.
    INTENSIVE CARE MEDICINE, 2010, 36 : S170 - S170
  • [6] Bedside capillary blood glucose measurements in critically ill patients: Influence of catecholamine therapy
    Hassen, M. Fekih
    Ayed, S.
    Gharbi, R.
    Ali, H. Ben Sik
    Marghli, S.
    Elatrous, S.
    DIABETES RESEARCH AND CLINICAL PRACTICE, 2010, 87 (01) : 87 - 91
  • [7] A comparison of arterial blood glucose and peripheral blood glucose levels in critically ill patients: measurements using the arterial blood gas analyzer and the rapid glucose meter
    Deng, Ting
    Liu, Mengjiao
    Pan, Lu
    Jiang, Kai
    Li, Yi
    ANNALS OF PALLIATIVE MEDICINE, 2021, 10 (03) : 3179 - 3184
  • [8] Agreement of blood glucose measured with glucose meter in arterial, central venous, and capillary samples in adult critically ill patients
    Rodriguez-Delgado, Esther
    Garcia del Moral, Raimundo
    Cobos-Vargas, Angel
    Martin-Lopez, Javier
    Colmenero, Manuel
    NURSING IN CRITICAL CARE, 2022, 27 (05) : 711 - 717
  • [9] Comparison of bedside and laboratory blood glucose estimations in critically ill children with shock
    Ramachandran, Bala
    Sethuraman, Ramaswami
    Ravikumar, Karnam G.
    Kissoon, Niranjan
    PEDIATRIC CRITICAL CARE MEDICINE, 2011, 12 (06) : E297 - E301
  • [10] Relative accuracy of arterial and capillary glucose meter measurements in critically ill patients
    Lonjaret, L.
    Claverie, V.
    Berard, E.
    Riu-Poulenc, B.
    Geeraerts, T.
    Genestal, M.
    Fourcade, O.
    DIABETES & METABOLISM, 2012, 38 (03) : 230 - 235