Calibration of pyroelectric infrared detectors using a light bulb as a black body radiation source

被引:0
作者
Pakluea, S. [1 ,2 ,3 ]
Jitvisate, M. [3 ]
Rimjaem, S. [1 ,3 ]
机构
[1] Chiang Mai Univ, Dept Phys & Mat Sci, Fac Sci, Plasma & Beam Phys Res Facil, Chiang Mai 50200, Thailand
[2] Chiang Mai Univ, Fac Sci, Dept Phys & Mat Sci, Master Degree Program Phys, Chiang Mai 50200, Thailand
[3] Commiss Higher Educ, Thailand Ctr Excellence Phys, Bangkok 10400, Thailand
来源
SIAM PHYSICS CONGRESS 2019 (SPC2019): PHYSICS BEYOND DISRUPTION SOCIETY | 2019年 / 1380卷
关键词
PLANCKS-CONSTANT DETERMINATION;
D O I
10.1088/1742-6596/1380/1/012152
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The pyroelectric detector is used for measuring infrared radiation. It has uniform response covering wide wavelength from 0.001 mu m to 1000 mu m and can measure the high-power level of radiation at room temperature operation. In this research, we calibrated the pyroelectric detector by using a tungsten light bulb as an infrared source. Properties of blackbody radiation from the light bulb are investigated. The interested properties include the spectral distribution and the relation of radiation intensity to heating temperature. The most radiation emitted from the light bulb lies in infrared wavelength. The pyroelectric detectors were calibrated against a pyroelectric Joulemeter with known responsivity. The results of this study will be applied to the measure of the radiation produced from the accelerator-base radiation source at the PBP-CMU Electron Linac Laboratory.