Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation

被引:32
|
作者
Roncevic, Mirjana Bjelos [1 ]
Busic, Mladen [1 ]
Cima, Ivan [1 ]
Elabjer, Biljana Kuzmanovic [1 ]
Bosnar, Damir [1 ]
Miletic, Daliborka [1 ]
机构
[1] Univ Hosp Sveti Duh, Univ Eye Clin, Zagreb 10000, Croatia
关键词
Biometry; Cataract; Optical low-coherence reflectometry; Applanation ultrasound; Intraocular lens; ANTERIOR-CHAMBER DEPTH; CATARACT-SURGERY; AXIAL LENGTH; IMMERSION ULTRASOUND; OCULAR BIOMETRY; INTERFEROMETRY; ACCURACY; REPRODUCIBILITY; IOLMASTER; CONTACT;
D O I
10.1007/s00417-010-1509-4
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The aim of the study was to determine whether the innovative non-contact optical low-coherence reflectometry method utilized by the Lenstar LS 900A (R) agrees sufficiently with applanation ultrasound A-scan technique in routine biometric measurement and intraocular lens power calculation to replace it. Twenty-two patients hospitalized at our eye clinic undergoing cataract surgery were assigned to have five consecutive measurements of axial length by two examiners in a single session using applanation ultrasound and the Lenstar. The applanation ultrasound intraocular lens power calculation was based on automated keratometry and applanation ultrasound axial length measurements. The Lenstar intraocular lens power calculation was based on its measurement of keratometry and axial length. Bland-Altman analysis was used to assess interobserver repeatability of applanation ultrasound and the Lenstar as well as agreement between the Lenstar and applanation ultrasound for axial length measurement and intraocular lens power calculation. Thirty-two eyes of 22 patients were analyzed. In 95% of the observations, predicted refractive error corresponded to -0.26 +/- 0.62 D and 0.01 +/- 0.20 D obtained with applanation ultrasound and the Lenstar, respectively. Based on excellent repeatability of the Lenstar and acceptable repeatability of applanation ultrasound, two techniques may be used interchangeably. The predicted refractive error of +/- 0.20 D in 95% of the observations has never been achieved. Optical low-coherence reflectometry might become a new standard method for biometric measurement needed for intraocular lens-power calculation in patients with cataract.
引用
收藏
页码:69 / 75
页数:7
相关论文
共 50 条
  • [21] Challenges in pediatric cataract surgery: comparison of intraocular lens power calculation formulas using optical biometry
    Yilmaz, Ibrahim Edhem
    Kimyon, Sabit
    Mete, Alper
    INTERNATIONAL OPHTHALMOLOGY, 2022, 42 (10) : 3071 - 3077
  • [22] Biommetry and intraocular lens power calculation results with a new optical biometry device: Comparison with the gold standard
    Kaswin, Godefroy
    Rousseau, Antoine
    Mgarrech, Mohamed
    Barreau, Emmanuel
    Labetoulle, Marc
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (04) : 593 - 600
  • [23] Comparison of intraocular lens power prediction using immersion ultrasound and optical biometry with and without formula optimization
    Nemeth, Gabor
    Nagy, Attila
    Berta, Andras
    Modis, Laszlo, Jr.
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 250 (09) : 1321 - 1325
  • [24] Comparison of the ultrasonographic method with 2 partial coherence interferometry methods for intraocular lens power calculation
    Salouti, Ramin
    Nowroozzadeh, Mohammad H.
    Zamani, Mohammad
    Ghoreyshi, Maryam
    Salouti, Rosa
    OPTOMETRY-JOURNAL OF THE AMERICAN OPTOMETRIC ASSOCIATION, 2011, 82 (03) : 140 - 147
  • [25] Comparison of Immersion Ultrasound, Partial Coherence Interferometry, and Low Coherence Reflectometry for Ocular Biometry in Cataract Patients
    Montes-Mico, Robert
    Carones, Francesco
    Buttacchio, Antonietta
    Ferrer-Blasco, Teresa
    Madrid-Costa, David
    JOURNAL OF REFRACTIVE SURGERY, 2011, 27 (09) : 664 - 670
  • [26] Comparing Ultrasound Biometry with Partial Coherence Interferometry for Intraocular Lens Power Calculations: A Randomized Study
    Raymond, Simon
    Favilla, Ian
    Santamaria, Linda
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2009, 50 (06) : 2547 - 2552
  • [27] Meta-analysis of optical low-coherence reflectometry versus partial coherence interferometry biometry
    Huang, Jinhai
    McAlinden, Colm
    Huang, Yingying
    Wen, Daizong
    Savini, Giacomo
    Tu, Ruixue
    Wang, Qinmei
    SCIENTIFIC REPORTS, 2017, 7 : 1 - 10
  • [28] Comparison of biometry and intraocular lens power calculation performed by a new optical biometry device and a reference biometer
    Ventura, Bruna V.
    Ventura, Marcelo C.
    Wang, Li
    Koch, Douglas D.
    Weikert, Mitchell P.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (01) : 74 - 79
  • [29] Optical low-coherence reflectometry enables preoperative detection of zonular weakness in pseudoexfoliation syndrome
    Bosnar, Damir
    Elabjer, Biljana Kuzmanovic
    Busic, Mladen
    Roncevic, Mirjana Bjelos
    Miletic, Daliborka
    Barac, Josip
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 250 (01) : 87 - 93
  • [30] Comparison of immersion ultrasound and low coherence reflectometry for ocular biometry in cataract patients
    Yan Li
    Hong-Xun Li
    Yang-Chen Liu
    Ya-Tu Guo
    Jian-Min Gao
    Bin Wu
    Nan Zhang
    Dong Liu
    Xiao-Yong Yuan
    International Journal of Ophthalmology, 2018, (06) : 966 - 969