Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation

被引:32
|
作者
Roncevic, Mirjana Bjelos [1 ]
Busic, Mladen [1 ]
Cima, Ivan [1 ]
Elabjer, Biljana Kuzmanovic [1 ]
Bosnar, Damir [1 ]
Miletic, Daliborka [1 ]
机构
[1] Univ Hosp Sveti Duh, Univ Eye Clin, Zagreb 10000, Croatia
关键词
Biometry; Cataract; Optical low-coherence reflectometry; Applanation ultrasound; Intraocular lens; ANTERIOR-CHAMBER DEPTH; CATARACT-SURGERY; AXIAL LENGTH; IMMERSION ULTRASOUND; OCULAR BIOMETRY; INTERFEROMETRY; ACCURACY; REPRODUCIBILITY; IOLMASTER; CONTACT;
D O I
10.1007/s00417-010-1509-4
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The aim of the study was to determine whether the innovative non-contact optical low-coherence reflectometry method utilized by the Lenstar LS 900A (R) agrees sufficiently with applanation ultrasound A-scan technique in routine biometric measurement and intraocular lens power calculation to replace it. Twenty-two patients hospitalized at our eye clinic undergoing cataract surgery were assigned to have five consecutive measurements of axial length by two examiners in a single session using applanation ultrasound and the Lenstar. The applanation ultrasound intraocular lens power calculation was based on automated keratometry and applanation ultrasound axial length measurements. The Lenstar intraocular lens power calculation was based on its measurement of keratometry and axial length. Bland-Altman analysis was used to assess interobserver repeatability of applanation ultrasound and the Lenstar as well as agreement between the Lenstar and applanation ultrasound for axial length measurement and intraocular lens power calculation. Thirty-two eyes of 22 patients were analyzed. In 95% of the observations, predicted refractive error corresponded to -0.26 +/- 0.62 D and 0.01 +/- 0.20 D obtained with applanation ultrasound and the Lenstar, respectively. Based on excellent repeatability of the Lenstar and acceptable repeatability of applanation ultrasound, two techniques may be used interchangeably. The predicted refractive error of +/- 0.20 D in 95% of the observations has never been achieved. Optical low-coherence reflectometry might become a new standard method for biometric measurement needed for intraocular lens-power calculation in patients with cataract.
引用
收藏
页码:69 / 75
页数:7
相关论文
共 50 条
  • [1] Comparison of optical low-coherence reflectometry and applanation ultrasound biometry on intraocular lens power calculation
    Mirjana Bjeloš Rončević
    Mladen Bušić
    Ivan Čima
    Biljana Kuzmanović Elabjer
    Damir Bosnar
    Daliborka Miletić
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2011, 249 : 69 - 75
  • [2] Predictive Accuracy of Intraocular Lens Power Calculation: Comparison of Optical Low-Coherence Reflectometry and Immersion Ultrasound Biometry
    Turhan, Semra Akkaya
    Toker, Ebru
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2015, 41 (04): : 245 - 251
  • [3] Comparison of the Zeiss IOLMaster and applanation A-scan ultrasound: biometry for intraocular lens calculation
    Rose, LT
    Moshegov, CN
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2003, 31 (02) : 121 - 124
  • [4] Comparison of a new optical biometry with an optical low-coherence reflectometry for ocular biometry
    Guler, Emre
    Kulak, Ali Ender
    Totan, Yuksel
    Yuvarlak, Armagan
    Hepsen, Ibrahim Feyzi
    CONTACT LENS & ANTERIOR EYE, 2016, 39 (05) : 336 - 341
  • [5] Comparison of two optical biometers in intraocular lens power calculation
    Hui, Sheng
    Yi, Lu
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2014, 62 (09) : 931 - 934
  • [6] Intraocular lens power calculation: Clinical comparison of 2 optical biometry devices
    Rabsilber, Tanja M.
    Jepsen, Charlotte
    Auffarth, Gerd U.
    Holzer, Mike P.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (02) : 230 - 234
  • [7] Comparison of preoperative and postoperative measurements of optical low-coherence reflectometry biometry and assessment of its refractive predictability
    Akalin, Irfan
    Tufek, Melek
    Turkyilmaz, Mustafa
    Ozturk, Faruk
    INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (06) : 1337 - 1343
  • [8] Ultrasound biometry and intraocular lens power calculation: comparative study of the contact and immersion techniques
    Re Martins, Felipe Cortizas
    Miyaji, Marcos Enrique
    Lima, Vagner Loduca
    Carvalho Lima Rehder, Jose Ricardo
    REVISTA BRASILEIRA DE OFTALMOLOGIA, 2009, 68 (04) : 212 - 215
  • [9] Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis
    W. Haigis
    B. Lege
    N. Miller
    B. Schneider
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2000, 238 : 765 - 773
  • [10] Comparison of Intraocular Lens Power Estimation by Optical Biometry and Ultrasound Biometry in Cataract Surgery
    Mohanty, Gayatree
    Rajguru, Himansu
    Agarwal, Khushi
    JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2022, 16 (06) : NC15 - NC18