Safety of magnetic resonance imaging in patients with permanent pacemakers: a collaborative clinical approach

被引:32
作者
Boilson, Barry Anthony [1 ]
Wokhlu, Anita [1 ]
Acker, Nancy G. [1 ]
Felmlee, Joel P. [2 ,3 ]
Watson, Robert E., Jr. [2 ]
Julsrud, Paul R. [2 ,4 ]
Friedman, Paul A. [1 ]
Cha, Yong-Mei [1 ]
Rea, Robert F. [1 ]
Hayes, David L. [1 ]
Shen, Win-Kuang [1 ]
机构
[1] Mayo Clin, Div Cardiovasc Dis, Rochester, MN 55905 USA
[2] Mayo Clin, Dept Radiol, Rochester, MN 55905 USA
[3] Mayo Clin, Dept Physiol & Biomed Engn, Rochester, MN 55905 USA
[4] Mayo Clin, Dept Hlth Sci Res, Rochester, MN 55905 USA
关键词
Arrhythmias; Device interactions; Magnetic resonance imaging; Permanent pacemakers; Power-on resetting; IMPLANTABLE-CARDIOVERTER-DEFIBRILLATORS; CARDIAC-PACEMAKERS; INDUSTRY VIEWPOINT; 1.5; TESLA; IN-VITRO; MRI-SAFE; ICDS; ISSUES; PERFORMANCE; 1.5-TESLA;
D O I
10.1007/s10840-011-9615-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study aimed to characterize the interactions of pacemakers with magnetic resonance imaging (MRI) and to identify device characteristics that could predict adverse interactions. The safety of MRI in patients with indwelling pacemaker systems remains uncertain. Previous studies demonstrated safety in most patients, but unpredictable, potentially concerning changes in pacemaker behavior have occurred. We prospectively studied patients with pacemaker devices in situ who were not pacemaker dependent and in whom MRI was essential for adequate diagnosis and treatment. All patients were monitored by electrocardiography and pulse oximetry during scanning; devices were interrogated and cardiac enzymes were measured before and after scanning. Of 32 patients studied (46 MRI examinations), 28 patients had a dual-chamber system and one had a biventricular device. Regions scanned were the head and spine. Devices were reprogrammed to asynchronous pacing or sense-only mode in all except six patients before MRI. During six scanning episodes (five patients), "power-on" resetting of the device was noted. Magnet-mode pacing was noted during four episodes (three patients). Occasional premature ventricular contractions were noted in one patient. No significant changes in battery voltage, sensed P wave and R wave, pacing thresholds, lead impedance, or cardiac enzymes were noted immediately after MRI or at 1-month follow-up. Overall, no significant changes were seen in pacemaker device function, and no adverse clinical events were observed. A minority of patients with older devices had unpredictable changes in device behavior, which stresses the need for close monitoring during and careful device interrogation after scanning.
引用
收藏
页码:59 / 67
页数:9
相关论文
共 30 条
  • [1] Pacing in magnetic resonance imaging environment:: Clinical and technical considerations on compatibility
    Duru, F
    Luechinger, R
    Scheidegger, MB
    Lüscher, TF
    Boesiger, P
    Candinas, R
    [J]. EUROPEAN HEART JOURNAL, 2001, 22 (02) : 113 - 124
  • [2] MR imaging in patients with cardiac pacemakers
    Duru, F
    Luechinger, R
    Candinas, R
    [J]. RADIOLOGY, 2001, 219 (03) : 856 - 857
  • [3] Food and drug administration perspective - Magnetic resonance imaging of pacemaker and implantable cardioverter-defibrillator patients
    Faris, Owen P.
    Shein, Mitchell
    [J]. CIRCULATION, 2006, 114 (12) : 1232 - 1233
  • [4] Unexpected asystole during 3T magnetic resonance imaging of a pacemaker-dependent patient with a 'modern' pacemaker
    Gimbel, J. Rod
    [J]. EUROPACE, 2009, 11 (09): : 1241 - 1242
  • [5] Safe sensible, sagacious: responsible scanning of pacemaker patients
    Gimbel, JR
    Wilkoff, BL
    Kanal, E
    Rozner, MA
    [J]. EUROPEAN HEART JOURNAL, 2005, 26 (16) : 1683 - 1684
  • [6] Goldsher D, 2006, ISR MED ASSOC J, V8, P91
  • [7] Can pacemakers, neurostimulators, leads, or guide wires be MRI safe? Technological concerns and possible resolutions.
    Helfer, JL
    Gray, RW
    Macdonald, SG
    Bibens, WT
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2006, 15 (02) : 114 - 120
  • [8] Do we need pacemakers resistant to magnetic resonance imaging?
    Irnich, W
    Irnich, B
    Bartsch, C
    Stertmann, WA
    Gufler, H
    Weiler, G
    [J]. EUROPACE, 2005, 7 (04): : 353 - 365
  • [9] Safety of Implantable Pacemakers and Cardioverter Defibrillators in the Magnetic Field of a Novel Remote Magnetic Navigation System
    Jilek, Clemens
    Tzeis, Stylianos
    Reents, Tilko
    Estner, Heidi-Luise
    Fichtner, Stephanie
    Ammar, Sonia
    Wu, Jinjin
    Hessling, Gabriele
    Deisenhofer, Isabel
    Kolb, Christof
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2010, 21 (10) : 1136 - 1141
  • [10] Current clinical issues for MRI scanning of pacemaker and defibrillator patients
    Kalin, R
    Stanton, MS
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (04): : 326 - 328